Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
Oct. 21, 2026DxU 840m Iris… and 1 more productClass IIThe rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage… Recall began Aug. 12, 2026
Oct. 14, 2026Battle Bandage…Class IICompression bandage may be degraded, which may cause unwanted foreign material to be introduced inside the wound, which may lead to wound exploration… Recall began Sept. 2, 2026
Oct. 7, 2026Intrapulmonary Percussive Ventilation… and 6 more productsClass IILabeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV… Recall began Aug. 18, 2026
Oct. 7, 2026HeartSine samaritan PAD (Public Access Defibrillator)…Class IIIProducts were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check. Recall began July 27, 2026
Sept. 23, 2026Happiest Baby… and 1 more productClass IIExtra small (ES) and Extra-large (XL) infant sleep sacks have not been reviewed for safety and effectiveness by the FDA. Recall began Aug. 7, 2026
Sept. 23, 2026Plum administration sets and 7 more productsClass IIIntravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored… Recall began May 12, 2026
Sept. 16, 2026HeartSine SAM 350P 350P US Device…Class IIDue to a damaged (over stretched) membrane tail that causes an intermittent fault. Recall began July 27, 2026
Aug. 19, 2026ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000Class IIA strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector… Recall began June 26, 2026
July 29, 2026SnapshotGLO Wound Imaging Device…Class IIThe software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully…
July 15, 20265008X Blood Tubing Set…Class IIPotential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the… Recall began May 11, 2026
July 15, 2026Intracept Access Instruments…Class IIIntraosseous nerve ablation system instruments may have sterile barrier breach of the outer tray seal, which may introduce a risk of sterility… Recall began April 22, 2026
July 8, 2026OTP5000 THORA/PARA TRAY… and 13 more productsClass IIDue to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled… Recall began May 21, 2026
July 1, 2026INZONE DETACHMENT SYSTEM…Class IIEmbolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint… Recall began April 6, 2026
June 24, 2026iLet Bionic Pancreas…Class IIThe reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure… Recall began Sept. 23, 2025
June 24, 2026Mito Red Light…Class IIThe near-infrared therapy device contains a lithium-ion battery that may malfunction, causing it to catch fire or burst into flames. Recall began Feb. 12, 2025
June 24, 2026PARACENTESIS KIT… and 34 more productsClass ILidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have… Recall began April 21, 2026
May 27, 2026Swan-Ganz Catheter… and 2 more productsClass IICatheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to… Recall began April 8, 2026
May 20, 2026Diowave Laser System…Class IISoftware update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms… Recall began Jan. 9, 2026
May 6, 2026Brand Name: Leksell Vantage Arc System REF: 1053958Class IILocking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures. Recall began March 19, 2026
April 22, 2026iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision SoftwareClass IIit was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI)… Recall began Aug. 13, 2025
April 8, 2026MiniMed 670G Insulin Pump (MMT-1760… and 13 more productsClass IIAll Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and… Recall began Feb. 13, 2026
April 1, 2026MiniMed" 780G…Class IIThree software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions… Recall began Jan. 30, 2026
April 1, 2026CentrosFLO Hemodialysis Catheters… and 4 more productsClass I16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies… Recall began Feb. 13, 2026
March 25, 2026Battery Charging Station…Class IIIThe Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used… Recall began Feb. 6, 2026
March 11, 2026Gelsoft Plus…Class IIGelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life.… Recall began Dec. 18, 2025
Feb. 25, 2026MOLLI 2 System System includes: MM1000 (Pack of 10)…Class IIDue to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during… Recall began Jan. 21, 2026
Feb. 25, 2026MONARCH Platform…Class IIDevice for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to… Recall began Jan. 21, 2026
Feb. 18, 2026da Vinci 5 Surgeon Console Viewer Display…Class IIRobotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field… Recall began Dec. 30, 2025
Feb. 11, 2026COULTER DxH Diluent…Class IIDue to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers. Recall began Nov. 10, 2025
Jan. 28, 2026SQ40S Blood Transfusion FilterClass IIDue to products being released for distribution prior to completion of the required quality control release process the sterility assurance cannot be… Recall began Nov. 3, 2025
Jan. 28, 2026The LDL-Cholesterol assay is comprised of two distinct phases…Class IIBeckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL… Recall began Dec. 8, 2025
Jan. 28, 2026Tego Connector: REF: 011-D1000…Class IINeedle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a… Recall began Dec. 19, 2025
Jan. 21, 2026CentriMag Blood Pump…Class IIAcute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the… Recall began Oct. 10, 2025
Jan. 14, 2026Codman Microsensor Basic Kit… and 1 more productClass IIPotential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits. Recall began Dec. 12, 2025
Jan. 14, 2026BD Pyxis ES Enterprise Server Catalog Numbers 1115-00…Class IIDue a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems. Recall began Dec. 3, 2025
Dec. 31, 2025BD Pyxis MedStation ES…Class IIAutomated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to… Recall began Nov. 20, 2025
Dec. 24, 2025Alinity ci-series System Control Module (SCM)…Class IIAbbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the… Recall began Nov. 13, 2025
Dec. 10, 2025Lumify Diagnostic Ultrasound System models : 795005…Class IIUltrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining. Recall began Oct. 31, 2025
Nov. 26, 2025MONARCH Bronchoscope…Class IIPotential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible… Recall began Oct. 18, 2025
Nov. 26, 2025PhD Hemostasis Valve…Class IIHemostasis valve has a manufacturing defect and when used the valve cap may remain depressed, leading to potential leakage, which may result in… Recall began Aug. 4, 2025
Nov. 19, 2025CEREPAK Detachable Coil System…Class IDetachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional… Recall began Oct. 2, 2025
Nov. 5, 2025Assay: IMMULITE 2000 Intact PTH…Class IIThe potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal… Recall began Sept. 22, 2025
Nov. 5, 2025SuRgical Planner (SRP)… and 6 more productsClass IIA software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any… Recall began Sept. 16, 2025
Oct. 22, 2025Zenith Alpha 2 Thoracic Endovascular Graft…Class IAffected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular… Recall began Sept. 18, 2025
Oct. 15, 2025Harris Teeter ASIAGO TEXAS TOAST… and 1 more productClass IUndeclared Milk. During production the firm discovered that the ingredients "cheese and butter" is not listed in the ingredient statement and… Recall began Sept. 11, 2025
Oct. 15, 2025CorNeat EverPatch… and 1 more productClass IIReviewed complaints have identified that conjunctival wound dehiscence may result in early exposure of the surgical patch. Recall began Aug. 18, 2025
Oct. 8, 2025GORE ACUSEAL Vascular Graft…Class IIReports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to… Recall began Aug. 26, 2025
Oct. 1, 2025FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative…Class IIFilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead… Recall began July 10, 2025
Sept. 24, 2025Vercise Deep Brain Stimulation Systems Surgical Implant Manual…Class IIWire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire… Recall began July 8, 2025
Sept. 10, 2025Interventional Fluoroscopic X-Ray System…Class IIPotential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series. Recall began Aug. 1, 2025
Sept. 10, 2025Superion IDS Kit Superion Indirect Decompression System 8MM…Class IIAs a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected… Recall began June 6, 2025
Sept. 3, 2025BD Pyxis Product Name / UDI-DI code / Catalog No…Class IIAntivirus software was not consistently installed on impacted devices during the implementation process. Recall began May 8, 2025
Sept. 3, 2025BD PhaSeal Injector Luer (N30C)…Class IIClosed system drug transfer devices were shipped to the U.S. market without a U.S. version of the instructions for use, which could be an issue if a… Recall began June 16, 2025
Aug. 27, 2025Burr Hole Cover Kit…Class IIResistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of… Recall began July 8, 2025
Aug. 27, 2025Pentax Medical Video Processor…Class IIDuring endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become… Recall began July 16, 2025
Aug. 20, 2025DreamStation Auto CPAP… and 2 more productsClass IDevices may possess a programming error resulting in an incorrect device configuration. Recall began June 30, 2025
Aug. 13, 2025UIBC (Unsaturated Iron Binding Capacity)…Class IIBeckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no… Recall began June 26, 2025
Aug. 13, 2025BD Alaris Pump Module Model… and 1 more productClass IInfusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy… Recall began July 8, 2025
Aug. 6, 2025Volcano Visions Digital IVUS Catheter: PV.014P (Platinum)…Class IIReports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or… Recall began June 13, 2025
Aug. 6, 2025HeartSine SAM 350P…Class IIDue to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks) Recall began June 30, 2025
July 30, 2025Stryker… and 1 more productClass IIDue to complaints their is the potential that irrigation solution may leaked into the handpiece and battery pack causing it to malfunction. Recall began June 18, 2025
July 30, 2025HSG Procedure Tray… and 2 more productsClass IIThe HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify… Recall began June 11, 2025
July 30, 2025Azurion 7 M20…Class IIPotential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose. Recall began June 4, 2025
July 16, 2025Brand Name: Alcon Laboratories… and 2 more productsClass IIDue to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area… Recall began April 28, 2025
July 2, 2025Product Name [REF]: Supply Server…Class IIDuring automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then… Recall began May 8, 2025
June 18, 2025Allon 2001…Class IIAn investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The… Recall began May 13, 2025
June 11, 2025Giant…Class IIUndeclared Allergen (walnuts), was not on outer packaging but was on the inner packaging of product. Recall began April 4, 2025
June 11, 2025AMCE GREEK SALAD AUTHENTIC FS…Class ISalmonella. Greek Salad products were made with recalled cucumbers from Bedner Grower's Company which are contaminated with Salmonella and implicated…
June 11, 2025GNC Women's 30+ Vitapak…Class IIIUndeclared Fish allergen (parvalbumin). Recall began April 25, 2025
June 4, 2025Atellica IM Homocysteine Calibrator (2 Pack)… and 1 more productClass IICertain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference… Recall began April 15, 2025
May 14, 2025Philips Tempus Pro Patient Monitor…Class IIPatient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required…
May 7, 20253.0 Liter Disposable Set (4 sets/case)…Class IIPotential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during… Recall began March 31, 2025
May 7, 2025Hyperthermia Pump Procedure Kit… and 2 more productsClass IIPotential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of… Recall began March 31, 2025
April 30, 2025Radiometer ABL90 Series…Class IIABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer… Recall began March 21, 2025
April 30, 2025Merit Medical… and 1 more productClass IIUpdate to IFU to provide additional information related to risk with device over-rotation and multiple deployment of fasteners. Recall began March 14, 2025
April 30, 2025BD PYXIS MEDBANK (Name…Class IIUsers trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a… Recall began March 11, 2025
April 30, 2025NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0…Class IINerve Monitoring System with certain software versions has potential for increased stimulus artifact, which may require troubleshooting, and may lead… Recall began March 10, 2025
April 23, 2025Maquet VH-3010 Power Supply…Class IIIncorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up. Recall began March 10, 2025
April 23, 2025HeartMate Mobile Power Unit: REF: 107754…Class ILeft Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or… Recall began March 13, 2025
April 23, 2025BD 4 Fr Single-Lumen PowerPICC Catheters…Class ICatheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher)… Recall began March 11, 2025
April 16, 2025BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower…Class IIDue to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect… Recall began March 11, 2025
March 26, 2025BD Pyxis: Anesthesia Station ES…Class IISync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount… Recall began Feb. 6, 2025
March 19, 2025BD Alaris Systems Manager… and 1 more productClass ISoftware issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).
March 12, 2025AURORA Surgiscope System…Class IIPossibility for the obturator to break (separate). Recall began Feb. 6, 2025
March 12, 2025Paradigm REF: MMT-712 and MMT-715 and 11 more productsClass IIAll Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin… Recall began Jan. 31, 2025
March 5, 2025BD Pyxis: MedFlex…Class IIAutomated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential… Recall began Jan. 10, 2025
Feb. 19, 2025Biological…Class IISpore test strip monitoring system, designed for routine monitoring of sterilization processes, may have a printing error on the return envelope that… Recall began Dec. 11, 2024
Feb. 12, 2025Temporary Titanium AbutmentsClass IICertain definitive dental implant restoration cases may have included a temporary titanium abutment (cylinder), rather than a definitive titanium… Recall began Dec. 6, 2024
Feb. 5, 2025Medtronic PIN…Class IIDue to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter. Recall began Dec. 4, 2024
Jan. 29, 2025Buttered Vegetables KEEP FROZEN…Class IIUndeclared Milk. During an FDA investigation the manufacturer was notified that the label does not list the allergen milk in the ingredient statement. Recall began Dec. 20, 2024
Jan. 1, 202520/30 INDEFLATOR… and 3 more productsClass IIInflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air… Recall began Nov. 6, 2024
Dec. 18, 2024BD Pyxis MedStation ES… and 2 more productsClass IIDue to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and… Recall began Nov. 14, 2024
Nov. 6, 20241) Flex Extenders…Class IIDue to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may… Recall began Sept. 27, 2024
Nov. 6, 2024Abbott Alinity m HPV AMP Kit… and 1 more productClass IIAbbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV… Recall began Sept. 12, 2024
Oct. 23, 2024Grabeez The Big Cheese SNACK MIX…Class IIUndeclared Almonds & Cashews. The firm was notified by their customers that the product contains a trail mix instead of the cheese mix. The trail mix…
Oct. 23, 2024S'mores Bites…Class IUndeclared Allergen, milk - a mix-up between the product description and the ingredient list. Milk Chocolate was listed in the product description… Recall began Sept. 13, 2024
Oct. 23, 2024Blue Diamond Digital Inflation Device… and 9 more productsClass IIInflation devices may have a small hole in the sterile barrier of the device and use of affected devices may result in exposure to pathogens/pyrogens… Recall began Aug. 19, 2024
Oct. 16, 2024Monarch Platform (Endoscopy)… and 1 more productClass IIEndoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which… Recall began Aug. 2, 2024
Oct. 9, 2024DAILIES TOTAL1…Class IIMaterial with a quality issue was used in contact lens production, so contact lenses may not meet quality and/or performance standards for the… Recall began Aug. 8, 2024
Oct. 9, 2024MiniMed 630G Insulin Pump… and 2 more productsClass IInsulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause… Recall began July 31, 2024
Oct. 2, 2024Karl Storz… and 7 more productsClass IIVarious medical device product IFUs contain reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA. Recall began July 23, 2024
Sept. 25, 2024Wegmans PECAN BLEND…Class IIUndeclared Almonds. The firm was notified by their customer that the product contains undeclared almonds. Recall began Aug. 23, 2024
Sept. 25, 2024ViewPoint… and 1 more productClass IIThe National Institute of Standards and Technology (NIST) has identified in the National Vulnerability Database (NVD) that NexGen Healthcare Mirth… Recall began Aug. 8, 2024
Sept. 25, 2024Alaris System PC Unit Model 8015… and 4 more productsClass IIDue to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or… Recall began Aug. 15, 2024
Sept. 18, 2024FILMARRAY 2.0 INSTRUMENT… and 1 more productClass IISome automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal… Recall began Aug. 7, 2024
Sept. 11, 2024Heparin… and 1 more productClass IIDue to packaging defects, sterility assurance and heparin activity may be compromised. Recall began July 1, 2024
Aug. 28, 2024Great Value…Class IIProduct contains inorganic arsenic above action level set in industry guidance (13.2ppb)
Aug. 28, 2024Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody…Class IITheir is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive… Recall began June 27, 2024
Aug. 28, 2024McGrath Mac Video Laryngoscope… and 1 more productClass IBattery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of… Recall began July 9, 2024
Aug. 28, 2024ClotTriever XL Catheter…Class IThrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent… Recall began July 19, 2024
Aug. 21, 2024WAVEWRITER ALPHA: Model: SC-1216…Class IISpinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device… Recall began July 17, 2024
Aug. 21, 2024Cepheid…Class IISpecimen collection device may leak after the patient sample swabs have been inserted into the test tubes, which could result in biohazard exposure… Recall began June 20, 2024
Aug. 7, 2024Arctic Sun Temperature Management Systems…Class IIIf temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden… Recall began June 27, 2024
July 31, 2024Abbott Infinity 5… and 3 more productsClass IITheir is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product… Recall began May 22, 2024
July 31, 2024Beckman Coulter…Class IILIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to… Recall began May 27, 2024
July 24, 2024BD Multitest 6-Color TBNK CE-IVD…Class IIMultitest with compromised raw material used to assess immune system status may result in flow cytometry plot with unusual staining pattern… Recall began May 8, 2024
July 17, 2024AURORA Surgiscope System…Class IIPotential sheath damage that could lead to tissue damage and/or track hemorrhage. Recall began June 7, 2024
July 17, 2024Kosmos on iOS impacting iOS…Class IIUltrasound system has a bug in affected iOS software, which after DICOM export, can cause inaccuracy in postprocessed measurements done in 3rd party… Recall began May 9, 2024
July 10, 2024Abbott Liberta RC…Class IIDeep brain stimulation system will first turn off after approximately 50 days after the system is activated and subsequently, this same action will… Recall began May 16, 2024
July 10, 2024BD Alaris System Manager REF: 9601…Class IIDue to a software issue the PC unit may not connect to the server which could impact wireless data transmission to and from the server. Recall began May 21, 2024
July 10, 2024BD Pyxis… and 1 more productClass IIMultiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical… Recall began May 13, 2024
June 19, 2024EsoFLIP…Class IIDue to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead… Recall began April 25, 2024
June 12, 2024Philips Tempus LS-Manual Defibrillator…Class IISimultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or… Recall began May 9, 2024
June 5, 2024BD Pyxis MedStation ES… and 5 more productsClass IIPotential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire. Recall began April 15, 2024
June 5, 2024Trevo XP ProVue Retriever 4 x 20… and 7 more productsClass IINeurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance… Recall began April 23, 2024
May 22, 2024HeartSine samaritan PAD (Public Access Defibrillator)…Class IIAutomated external defibrillators have a manufacturing issue that could prevent the device from delivering instructional voice prompts to the user… Recall began April 8, 2024
May 8, 2024Abbott Aveir Leadless Pacemaker…Class IITheir is a potential that electromagnetic interference may cause leadless pacemaker may result in mode change. Recall began Feb. 5, 2024
May 8, 2024Keller Funnel 2…Class IITheir is a potential that breast implants lack the lubricious coating that allows them to slide through the funnels during the implantation procedure. Recall began March 14, 2024
May 8, 2024Thoratec HeartMate 3… and 1 more productClass IPotential for a leak path at the seal interface between the left ventricle assist system inflow cannula and the titanium apical cuff. Recall began March 1, 2024
May 1, 2024Masimo Rad-G Refurbished… and 1 more productClass IITheir is a potential that Oximeter may automatically power off and on resulting in loss of monitoring. Recall began Feb. 15, 2024
May 1, 2024QUIDEL InflammaDry MMP-9 Test… and 1 more productClass IIThere are two windows on the test device, the control line window and the test results window. The line for the test results window were misaligned.… Recall began March 5, 2024
April 24, 2024Masimo Rad-G… and 3 more productsClass IITheir is a potential that Oximeter may automatically power off and on resulting in loss of monitoring. Recall began Feb. 15, 2024
April 24, 2024Tablo Hemodialysis System…Class IHemodialysis System, pre and post dialyzer peroxide-cured silicone tubing, with less than 336 hours usage, will be replaced because tubing may… Recall began March 6, 2024
April 3, 2024Epix Universal Clip Applier… and 1 more productClass IITheir is the potential that Universal Clip applier may not load clip after the trigger is actuated. Recall began Jan. 30, 2024
April 3, 2024Medivance Neonatal ArcticGel Pads…Class IINeonatal pads are experiencing reduced water flow, which may reduce or prevent water heating or cooling performance of the pads that are part of a… Recall began Feb. 16, 2024
March 13, 2024BD Pyxis: MedStation ES…Class IIAutomated dispensing cabinets, running affected software, and Windows 10 operating system may experience the application randomly hang, freeze and/or… Recall began Feb. 1, 2024
Feb. 28, 2024Philips Allura Xper Interventional Fluoroscopic X-ray Systems…Class IIDual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of… Recall began Jan. 2, 2024
Feb. 21, 2024Samsung Digital Diagnostic Mobile X-ray System…Class IIMobile x-ray systems moving arm frame welding issue which presents the potential of bodily harm due to fall of the arm. Recall began Dec. 28, 2023
Feb. 14, 2024Brasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023…Class IIThe device has a grip detail (right angle latch) out of specification causing the bur to not adequately latch in the handpiece. Recall began Jan. 11, 2024
Feb. 14, 2024Total immunoglobulin M (IgM total) IVD…Class IIPotential for a negative bias with quality control (QC) and patient sample results when using the Atellica CH Immunoglobulin M_2 (IgM_2) reagent. Recall began Dec. 15, 2023
Feb. 14, 2024Flow-i C20 Anesthesia System… and 4 more productsClass IIPotential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to… Recall began Dec. 11, 2023
Feb. 7, 2024VIVA for cats duck… and 3 more productsClass IFDA notified the firm that via our Adverse Event Reporting system that a positive test results for Listeria monocytogenes and Salmonella was found in…
Jan. 31, 2024TactiFlex Ablation Sensor Enabled Ablation Catheter…Class IIWhen the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and… Recall began Dec. 18, 2023
Jan. 24, 2024INDICAID COVID-19 Rapid Antigen At-Home Test…Class IICOVID-19 rapid antigen at home tests were released with an incorrect expiration date of 05 April 2024 when the correct expiration should be 21… Recall began Oct. 25, 2023
Jan. 17, 2024BD Recanalization System…Class IIRecanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute… Recall began Nov. 14, 2023
Jan. 17, 2024CUSTOM SPRYSTEP… and 2 more productsClass IIDue to potential air pockets between layers of material, their is the potential for complete struct breakage of knee-ankle-foot orthosis. Recall began Sept. 23, 2023
Jan. 17, 2024CD11b APC: ASR…Class IIIn vitro diagnostics product that impacts the quality of cell staining can cause low or dim fluorescence signal, which can lead to potential false… Recall began Oct. 25, 2023
Dec. 27, 2023Injection Needle… and 4 more productsClass IIThe efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery… Recall began Oct. 24, 2023
Dec. 20, 2023CereLink ICP Extension Cable (Extension Cable)…Class IFirm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated… Recall began Oct. 5, 2023
Dec. 13, 2023Whole cantaloupe… and 1 more productClass IThe affected product is being recalled from the marketplace due to possible Salmonella contamination. Recall began Nov. 6, 2023
Dec. 13, 2023BIOFIRE Blood Culture Identification 2 (BCID2) Panel…Class IIRoutine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control… Recall began Oct. 17, 2023
Dec. 13, 2023CARDIOHELP Emergency Drive…Class IProduct removal due to possible blocking or impairment of the CARDIOHELP Emergency Drive. This issue can lead to the patient being exposed to… Recall began Oct. 10, 2023
Dec. 6, 2023Acumen IQ Sensor… and 5 more productsClass IITheir is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer. Recall began Oct. 4, 2023
Nov. 29, 2023FilmArray TORCH REF HTFA-ASY-0104…Class IIDue to an increased risk of degradation of the power entry switch resulting from arcing inside of the power switch that may result in carbon build-up… Recall began Oct. 17, 2023
Nov. 15, 2023BD Alaris PCU REF 8015 and 2 more productsClass IAlaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated. Recall began Sept. 15, 2023
Nov. 1, 2023Whole cantaloupe…Class IPossible contamination with Salmonella spp. Recall began Sept. 22, 2023
Nov. 1, 2023SOZO Bilateral Arm L-Dex SoftwareClass IIBilateral L-Dex assessment software does not have the same level of sensitivity to help detect early signs of lymphedema as the unilateral arm L-Dex… Recall began Aug. 17, 2023
Nov. 1, 2023TruDi Navigation System…Class IIWhen using affected curette and software, there is a discrepancy between the actual curette tip location and the location displayed on navigation… Recall began Sept. 21, 2023
Nov. 1, 2023Astral 100 and Astral 150 ventilatorsClass IIf ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault… Recall began Sept. 13, 2023
Oct. 25, 2023Sunnyside Farms…Class IISafeway Fresh Foods of Vineland, NJ, is recalling its 32-oz. Diced Organic Butternut Squash product as it has the potential to be contaminated with… Recall began Sept. 15, 2023
Oct. 18, 2023CellQuicken Analyzer (Smart-Watch and Software)…Class IIUltrasound devices were not authorized, cleared, or approved by FDA, and thus their performance characteristics have not been adequately established.… Recall began Aug. 30, 2023
Oct. 11, 2023Heartmate 3 LVAS Implant Kit…Class IIInability to start and/or complete the coring procedure due to inadequate sharpness of the Apical Coring Knife, provided with Left Ventricular Assist… Recall began Aug. 7, 2023
Sept. 27, 2023Philips Tempus Pro Patient Monitor…Class IIFull screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from… Recall began Aug. 24, 2023
Sept. 27, 2023Five S 5 3x65…Class IIFlexible intubation endoscopes sterility assurance can not be confirmed. Recall began July 7, 2023
Sept. 13, 2023ISee Ortho-K Lens and 1 more productClass IIManufactured lenses are not covered by existing FDA approval Recall began June 26, 2023
Sept. 6, 2023Beckman Coulter…Class IILipemic interference failed to meet the performance specification listed within the IFU. Recall began July 12, 2023
Aug. 16, 2023Randox RX Series Urea Enzymatic Kinetic Assay Ref…Class IICarryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient… Recall began June 13, 2023
July 26, 2023Durepair Dura Regeneration Matrix: DUREPAIR DURA SUBSTITUTE 2 X 2…Class IIDura regeneration matrix may have out-of-specification endotoxin levels due to issues with in-process and finished goods endotoxin testing, which may… Recall began June 13, 2023
July 26, 2023Simplexa Direct Amplification Disc Kit…Class IIThere is a potential for the direct amplification disc to malfunction which may result in spillage of liquid from the disc. Recall began June 14, 2023
July 26, 2023Aesculap Surgical Needle…Class IIIncorrect product code marked on the product, etched with product code MD611, however; the correct product code is MD610. Recall began May 18, 2023
July 12, 2023BD Pyxis CII Safe…Class IIAutomated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage, which… Recall began May 22, 2023
July 12, 2023da Vinci Xi/X Tip-Up Fenestrated Grasper…Class IIFenestrated Graspers have a manufacturing issue has the potential to cause breakage on both upper and/or lower grips, which could lead to stainless… Recall began June 12, 2023
June 28, 2023Metal Tray…Class IIMetal sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for the… Recall began April 6, 2023
June 28, 2023KARL STORZ Radel Sterilization Trays…Class IIPlastic sterilization trays used to encase and protect medical devices during sterilization have a new 510(k) with instructions for use that is for… Recall began April 5, 2023
June 21, 2023Driver Instrument…Class IIPending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause… Recall began May 12, 2023
June 14, 2023StealthStation Cranial Software…Class IDuring non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing… Recall began April 11, 2023
May 31, 2023Copitas…Class IIProduct may exceed FDA guidance levels for leachable lead. Recall began May 1, 2023
May 24, 2023Beckman Coulter Bicarbonate… and 2 more productsClass IIThere is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control… Recall began April 5, 2023
April 26, 2023Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff…Class IA manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure… Recall began March 24, 2023
April 19, 2023Edward Lifesciences… and 1 more productClass IIThere is a potential that the distal end of the catheter may be bent beyond its intended curvature and when connected to patient monitoring device… Recall began Feb. 14, 2023
April 5, 2023Color Cuff Single Port…Class IIIncreased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing… Recall began March 1, 2023
March 29, 2023BD Pyxis Anesthesia ES System (PAS ES)…Class IIAutomated dispensing cabinet drawer firmware may have inadvertently been downgraded to an earlier version of drawer firmware during new device… Recall began Feb. 14, 2023
March 22, 2023Bistro to Go… and 408 more productsClass IOn 02/03/2023, FRESH IDEATION FOOD GROUP, LLC dba Fresh Creative Cuisine of Baltimore, MD is initiating a recall of various products sold from… Recall began Feb. 3, 2023
March 22, 2023Recreo Hair Growth Laser CombClass IIThe firm distributed the laser products without affixing the warning, aperture, certification, and identification labels. In addition, the firm… Recall began Sept. 1, 2022
March 15, 2023Simplexa COVID-19 Direct… and 1 more productClass IIDirect amplification Discs , used with COVID-19 and Flu A/B & RSV assays, may leak and cause contamination, cross-contamination, invalid results, and… Recall began Jan. 30, 2023
March 15, 2023Cardiosave Hybrid Model Nos… and 1 more productClass IAn unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may… Recall began Feb. 7, 2023
Feb. 15, 2023The Covera Vascular Covered Stent is a flexible…Class IIAn increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding. Recall began Dec. 22, 2022
Feb. 8, 2023AxiEM" Non-Invasive Patient TrackerClass IIDue to increasing complaint trend for the Non-Invasive Patient Tracker system where complaints analysis indicates that users are unable to… Recall began Dec. 15, 2022
Dec. 28, 2022Vivid S5 Ultrasound… and 7 more productsClass IIPotential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch… Recall began Nov. 10, 2022
Dec. 28, 2022The Natus Cranial Access Kit is designed to be a complete…Class IIDue to Cranial Access Kits containing recalled sterile surgical drapes. The sterile surgical drapes are being recalled due to the liner on the… Recall began Nov. 18, 2022
Dec. 14, 2022LineSider Spinal System 5.5/6.0mm… and 2 more productsClass IIPotential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the… Recall began Jan. 27, 2022
Dec. 7, 2022Accelerate PhenoTest BC kit REF 10101018 and 1 more productClass IIDue improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification… Recall began Oct. 17, 2022
Nov. 16, 2022Hammond's Salted Caramel Cookies…Class IPeanut Butter Chocolate Chip Cookies mispacked in Salted Caramel Cookies Box creating undeclared peanut allergen. Recall began Oct. 13, 2022
Nov. 16, 2022Insulin Pump/Model: MiniMed 630G/ MMT-1715… and 1 more productClass IIMedtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that… Recall began Sept. 20, 2022
Nov. 16, 2022Zio AT Clinical Reference Manual…Class IIAmbulatory ECG monitoring system labeling update: 1) Clinically actionable arrhythmia notifications delayed until patient registration is complete… Recall began Sept. 28, 2022
Nov. 16, 2022NaturaLyte Liquid Acid concentrate…Class IIThe "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl… Recall began Oct. 4, 2022
Nov. 9, 2022BD Trucount Tubes (Cat… and 2 more productsClass IIDue to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the… Recall began Oct. 3, 2022
Oct. 5, 2022Ambra PACS…Class IIA race condition between the storage system and services database has the potential to revert edits made to patient information upon first ingestion… Recall began June 23, 2022
Oct. 5, 2022t:slim X2 insulin pump… and 1 more productClass IIInsulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen… Recall began May 24, 2022
Sept. 28, 2022Puritan Bennett 560 Ventilator…Class IIMedtronic has confirmed six reports from customers outside the US where the ventilator became inoperable due to loss of gas supply caused by… Recall began Aug. 19, 2022
Sept. 28, 2022AquaC UNO H [120V… and 1 more productClass IIInternal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured Recall began June 21, 2022
Sept. 21, 2022QIAcube Connect MDx…Class IIDuring the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is… Recall began Nov. 21, 2021
Sept. 7, 2022Wegmans Lemon Dill Finishing Butter…Class IFirm's supplier of frozen dill notified firm of a recall due to potential contamination with Listeria monocytogenes.
Sept. 7, 2022T-SPOT.TB REF TB.300 US and 2 more productsClass IIDue to equipment failure, affected products were stored at temperatures outside the validated range of 2-8 degree C for a period longer than the… Recall began July 7, 2022
Aug. 31, 2022Vanta Clinician Programmer Application…Class IIPotential communication issue in which the Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 may be unable to connect with a Vanta… Recall began July 26, 2022
Aug. 31, 2022Vanta Clinician Programmer Application…Class IIPotential for Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 to display an Unexpected Device Error Code 1502 Message, and the user… Recall began July 26, 2022
Aug. 17, 2022cobas e801 Immunoassay AnalyzerClass IISoftware issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the… Recall began July 14, 2022
Aug. 10, 2022UroLift System REF UL400-4…Class IIThere is the potential that during implant deployment, the device may not properly deliver a implant. Recall began June 24, 2022
Aug. 3, 2022Da Vinci S/Si Large Clip Applier… and 1 more productClass IIDue to the potential for unexpected motion caused by the partial disengagement of the instrument from the system. Recall began July 1, 2022
July 27, 2022Trevo Trak 21 Microcatheter…Class IIMicrocatheter product label on the carton is missing the microcatheter product drawing/illustration. Recall began June 3, 2022
July 20, 2022Ventana HE 600 System…Class IIThere is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire Recall began June 1, 2022
July 20, 2022MiniMed 620G Insulin Pump (O.U.S… and 7 more productsClass IIDue to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a… Recall began May 23, 2022
July 13, 2022ABL800 Flex Analyzer Model Numbers 393-800 and 393-801Class IIThere is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+. Recall began May 18, 2022
July 13, 2022Palindrome Precision Chronic Catheter… and 8 more productsClass IThere is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which… Recall began June 8, 2022
July 6, 2022Volcano Visions PV… and 1 more productClass IIIDigital IVUS catheter instructions for use ( IFU) do not contain contrainidications Recall began May 13, 2022
June 8, 2022DYNEX Agility…Class IIControl samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that… Recall began July 5, 2020
June 8, 2022BD Anti-Kappa APC (Cat… and 1 more productClass IIAnti-Kappa APC & Anti-Kappa PE analyte reagents may exhibit uncharacteristic flow profiles. Recall began March 25, 2022
June 8, 2022The C304-HIS device features a guide wire to access the vein…Class IIThe firm's internal processes identified that certain lots may have an improperly sealed barrier that could potentially compromise sterility. Recall began April 20, 2022
June 1, 2022Vanta INS Implantable Neurostimulator…Class IIThe Implantable Neurostimulator Model may be susceptible to not communicating with the clinical programmer and/or patient programmer system. Recall began March 30, 2022
May 25, 2022Driscoll's Blackberries (Fresh) 12x6 oz…Class IIICalifornia Department of Pesticide Regulation detected the presence of Formetante HCI. This pesticide has a non-tolerance for blackberries. Recall began April 15, 2022
May 25, 2022Radiometer…Class IIArterial Blood Sampler sterility cannot be guaranteed throughout the shelf life of the product. Recall began April 8, 2022
May 25, 2022Dragonfly OpStar Imaging Catheter…Class IThere is a potential that the proximal marker may separate from the imaging catheter Recall began March 30, 2022
May 25, 2022Getinge 9100-Series Washer Disinfector models: 9120E…Class IIInadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to… Recall began April 1, 2022
May 4, 2022TearCare version 1.0 SmartHubsClass IIOphthalmic devices distributed prior to the recently-obtained 510(k) clearance do not fall within the relevant 510(k) exemption and so their… Recall began March 23, 2022
April 20, 2022ARIA Radiation Therapy Management (RTM) VERSIONS 13.6…Class IISoftware issue for treatment plan and image management application may result in mismatch values which could result in treatment in the wrong location Recall began March 21, 2022
April 20, 2022Access-9 Large Bore Hemostasis Valves… and 1 more productClass IIA design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the… Recall began March 1, 2022
April 13, 2022BenchMark ULTRA and DISCOVERY ULTRA InstrumentsClass IIPotential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1… Recall began Jan. 20, 2022
April 13, 2022SyncVision Systems…Class IIIf FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR… Recall began Jan. 19, 2022
April 6, 2022MiniMed 630G: MMT-1715…Class IINew or replacement insulin pumps are not pre-programmed with basal rates or other verified settings (bolus wizard settings, sensor settings, etc.)… Recall began Jan. 27, 2022
April 6, 2022RayStation 4-11B…Class IIIncorrect source to skin or surface distance (SSD) calculation may result in dose error when using RayStation system. Recall began Feb. 11, 2022
April 6, 2022FORE-SIGHT ELITE Absolute Tissue Oximeter… and 1 more productClass IIThe StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue… Recall began March 1, 2022
April 6, 2022Liko Universal Twinbar 670 QRH…Class IIAn incorrect center bolt was used in the assembly of the product which could result in a patient fall during use. Recall began Feb. 21, 2022
March 23, 2022Panther Fusion Extraction Reagent-X…Class IIFailed stability specifications for reagent kits that may result in false negative test results. Recall began Jan. 27, 2022
March 16, 2022Factor 75…Class IIUndeclared Milk Allergen
March 16, 2022CIRRUS HD-OCT…Class IIOptic nerve head angiography scan to be turned off due to its distribution without pre-market clearance. Recall began Jan. 14, 2022
March 9, 2022Alinity m Integrated Reaction Units (IRU)Class IIThere is potential for the Reaction Vessels (RVs) detaching from the RV cap during transport by the pipettor robot.
March 9, 2022da Vinci Xi and X 12-8 mm Conductive Cannula Reducer…Class IIConductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal… Recall began Dec. 23, 2021
March 9, 2022AU/DxC AU Magnesium…Class IILipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive… Recall began Dec. 8, 2021
March 9, 2022Acorn 130 Stairlift…Class IIThe design of the seats did not consistently have the adequate strength required to support the customers weight with continued use. Recall began April 20, 2016
March 2, 2022Endurant…Class IIStent Graft Systems built with specific batches of taper tip assemblies have the potential for the taper tip to detach from the delivery system.… Recall began Dec. 21, 2021
Feb. 23, 2022bellavista 1000 ventilator…Class IPotential cessation in ventilation can occur under specific conditions. Recall began Dec. 20, 2021
Feb. 2, 2022GE Vscan Extend… and 6 more productsClass IIThe firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to… Recall began Nov. 4, 2021
Jan. 19, 2022Top Quality Dog Food… and 1 more productClass ITop Quality Dog Food, LLC. 3630 East St. Hyattsville, MD 20785 has initiated a recall of Top Quality Dog Food, Beef HVM, 1 lb. pkgs in film casings… Recall began Aug. 19, 2021
Jan. 12, 2022Alinity m System…Class IIThere is an issue with the installation of updated camera firmware on the system. Recall began Dec. 6, 2021
Jan. 12, 2022Simplexa Direct Amplification Disc Kit…Class IIA subset of affected discs are at risk of leakage when it is used more than one time with Simplexa COVID-19 Direct and/or Simplexa Flu A/B & RSV… Recall began Nov. 19, 2021
Jan. 5, 2022Karl Storz Udel Sterilization Tray Instructions for Use…Class IITo ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible… Recall began Oct. 15, 2021
Dec. 29, 2021Leica Microsystems M220 F12 Microscope SystemsClass IIIn May 2021, Leica Microsystems received one (1) complaint pertaining to the M220 F12 Surgical Microscope system. There was no patient injury… Recall began Nov. 3, 2021
Dec. 29, 2021WIRION EMBOLIC PROTECTION SYSTEM…Class IUnder certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under such circumstances, alternative means may be… Recall began Nov. 22, 2021
Dec. 22, 2021Miami J Select Collar…Class IIInstructions for use were updated to include additional information to facilitate the selection and use of the collar device. Recall began Nov. 9, 2021
Dec. 8, 2021SHIP KIT…Class IIITop label of the outer package of the collection kit Item 5820-3118 showed expiration date of 2022/05/01 (correct expiration date) while side label… Recall began Oct. 14, 2021
Dec. 8, 2021Abbott HeartMate Touch Communication System…Class IIIf LVAS communication system is trying to establish Bluetooth connection and another Bluetooth-enabled device is nearby and advertising for Bluetooth… Recall began Oct. 22, 2021
Dec. 1, 2021Kings Hawaiian Hawaiian Sweet Hamburger Buns8 buns…Class IIProducts that were on HOLD for potential foreign materials was accidently released into commerce. The products released may be contaminated with…
Dec. 1, 2021ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87… and 1 more productClass IITwelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect… Recall began Sept. 29, 2021
Nov. 10, 2021IPL Coupling Gel…Class IUltrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination. On August 18, 2021, FDA published a…
Nov. 10, 2021FIVE S 3.5x65…Class IIDuring a regular requalification review, affected flexible intubation video endoscopes did not achieve the required sterility assurance level. There… Recall began Sept. 10, 2021
Nov. 3, 2021REF/Catalogue Number GSX0030A… and 8 more productsClass IISeptal Occluders contain the incorrect expiration date. Product is labeled with a 3 yr. expiration date instead of 2 yrs. Recall began Sept. 2, 2021
Oct. 20, 2021Elekta Medical Linear Accelerator (EMLA)…Class IIThere is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube… Recall began Sept. 17, 2021
Oct. 20, 2021Ion Flexison Biopsy Needle…Class IIPotential for the sheath tip on the biopsy needle to separate from the sheath shaft. Recall began Aug. 6, 2021
Oct. 13, 2021CardioQuip… and 3 more productsClass IILabeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices. Recall began May 12, 2021
Oct. 13, 2021Access hsTnI High Sensitivity Troponin I…Class IIThrough customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent… Recall began Aug. 9, 2021
Oct. 13, 2021Nutrifill Scleral…Class IIContact lens solution may contain foreign material Recall began July 27, 2021
Oct. 13, 2021GEL CONDUCTOR… and 1 more productClass IUltrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination. On August 18, 2021, FDA published a… Recall began Sept. 7, 2021
Sept. 15, 2021Nihon Kohden NKV-550 Ventilator…Class IIPotential defective gas inlet pressure regulator within ventilator which could result in gas releasing noise from the bottom of the ventilator, a… Recall began July 20, 2021
Sept. 15, 2021Pipeline Flex Embolization Device and 1 more productClass IDue to potential push wire fractures in the delivery system during use. The issue presents as a fracture in the spiral cut area of the Hypotube. Recall began July 13, 2021
Sept. 1, 2021AQUABEAM Handpiece…Class IIScope tube tip may detach from the telescoping tube, which is attached to a single-use handpiece, a component of an aquablation robotic system. Both… Recall began July 9, 2021
Aug. 25, 2021Cell Marque Antibodies…Class IIIMouse Monoclonal Antibody labelled with an incorrect expiry date of 31 Dec 2023. The correct expiry date is 31 Dec 2022. If used beyond expiry, the… Recall began July 15, 2021
Aug. 25, 2021EPIQ Ultrasound Systems…Class IIUltrasound system control panel arm swivel lock could fail, which could cause the control panel and monitor to swivel freely while a user is… Recall began July 22, 2021
Aug. 18, 2021Chassis Label…Class IIDue to failure to acquire pre-market clearance for its high resolution tomographic/biomicroscope imaging device (Model 6000) prior to commercializing… Recall began July 8, 2021
Aug. 11, 2021GlideScope Go MonitorsClass IIIHandheld video monitor failed to meet the labeled IP67 rating which may allow fluid ingress into the monitor resulting in loss of functionality. Recall began July 1, 2021
Aug. 4, 2021OEC 3D Imaging… and 1 more productClass IIThere is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable. Recall began June 10, 2021
July 28, 2021Infant Heel Warmer with Attachment Tab… and 1 more productClass IIA caution statement is being added to remind the user to activate away from face and infant as reports of pack rupture or leaking during activation… Recall began June 9, 2021
July 7, 2021Lyra SARS-CoV-2 Assay (M120) in vitro diagnostic test reagentClass IIn rare situations patients infected with high viral load of SARS CoV 2 may have samples that generate Ct values less than or equal to (d) 5 when… Recall began April 27, 2021