Battery Charging Station…
- Reason
- The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full…
- Status
- Ongoing
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.
- Enforcement report date
- March 25, 2026
- Risk class
- Class III (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences)
- Status
- Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
- Filed by
- Datascope Corp.
- Product
- Battery Charging Station; Model: 0998-00-0802;
- Reason
- The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
- Distribution pattern
- Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.
- Quantity
- 28 units (20 US, 8 OUS)
- Codes and lots
- Model: 0998-00-0802; UDI: 10607567111964; Serial Numbers: 200019347EAI, 200019847EAI, 200019447EAI, 210018625EAI, 210017925EAI, 210017525EAI, 210018425EAI, 210020144EAI, 210020944EAI, 210022544EAI, 210020244EAI, 210022144EAI, 210020044EAI, 210020544EAI, 210021344EAI, 210023244EAI, 210018025EAI, 210019025EAI, 210021144EAI, 210021044EAI, 210022344EAI, 210019225EAI, 210019125EAI, 210018525EAI, 210022844EAI, 200017647EAI, 210019725EAI, 210022744EAI;
- Recall started
- Feb. 6, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- System, Balloon, Intra-Aortic And Control
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On February 6, 2026 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken by the customer: Our records indicate that you may have one or more Battery Charging Station(s) in your facility. Please forward this information to all current Battery Charging Station users within your facility. " If you are currently using one of the affected Battery Charging Stations, please follow the instructions in the Battery Charging Instructions section of this letter. " Complete and sign the attached Response Form to acknowledge that you have received and understand this notification. Return the completed form to recallresponses.grc@getinge.com. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action.
- Recall number
- Z-1587-2026
What this record is about
A generic icon of the kind of thing the record is about; the record's own words are below. It is not a picture of the product.
“The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.”Battery Charging Station; Model: 0998-00-0802;, FDA enforcement report, 2026-03-25. The caption is the record's own words.
The picture shows the kind of thing the record is about (a part, a product, a package), not the product itself; the ring marks it. Nothing in it is a claim beyond the record's own words below.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: System, Balloon, Intra-Aortic And Control
Distributed in (FDA's record): District of Columbia Florida Georgia Illinois Indiana Maryland Michigan North Carolina Texas Washington Wisconsin and nationwide
This recall among the others
- FDA's enforcement report came 47 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).
- Datascope: 17 other recalls or notices on this site in the three years before this one (from March 26, 2023).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement report, recall number Z-1587-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-1587-2026, March 25, 2026. "Battery Charging Station." PlainRecords, https://plainrecords.com/recall/device/z-1587-2026/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections