FDA · Medical device recall

Irvine Scientific Oil for Embryo Culture (Light Mineral Oil) REF Catalog # 9305

April 5, 2023 Class II

Date
April 5, 2023
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Fujifilm Irvine Scientific, Inc.
Product
Irvine Scientific Oil for Embryo Culture (Light Mineral Oil) REF Catalog # 9305
Reason
Due to the detection of oil toxicity in several lots of Oil for Embryo Culture products.
Distribution pattern
U.S.: AZ, AR, CA, CT, DC, FL, HI, IA, ID, IL, IN, KS, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WI and WV. O.U.S.: India, Israel, Japan, Malaysia, Mexico, South Korea, South Africa, Thailand, Vietnam, Brazil, Bosnia, Argentina, Netherlands, and United Kingdom
Quantity
15,131 bottles
Codes and lots
UDI-Di Code: 08937270029305TH Catalog Number / Lot Numbers: 9305-100ml / 0000011351, 9305-500ml / 0000011367, 9305-100ml / 0000015999, 9305-500ml / 0000016001
Recall started
Jan. 16, 2023
Device type
Media, Reproductive
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
On 01/16/2023, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that, the firm has received complaints of reduced blastocyst development, including complete degradation in some cases, during culture of human embryos. Additionally, some clinics performed Human Sperm Survival Assay (HSSA), as part of incoming confirmatory Quality Control testing, and reported low sperm motility. Customers are instructed to: 1. Immediately stop using the product; 2. Quarantine any remaining product in inventory; 3. Return or destroy the product; 4. Complete and return the enclosed MEDICAL DEVICE RETURN RESPONSE FORM; and 5. Email the completed MEDICAL DEVICE RETURN RESPONSE FORM to fisira@fujifilm.com by January 23, 2023. For questions, contact: Customer Service Manager, Medical Business: - Toll free in the US: 1.800.577.6097 - Phone: 1.949.261.7800 x457 - Email: Melissa Gomez, fisitmrequest@fujifilm.com
Recall number
Z-1314-2023

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-1314-2023. The agency record governs; check it and the company before acting.