FDA · Medical device recall

SOZO Bilateral Arm L-Dex Software

Recall started · FDA enforcement report Nov. 1, 2023 Class II

Reason
Bilateral L-Dex assessment software does not have the same level of sensitivity to help detect early signs of lymphedema as the unilateral arm L-Dex assessment, which could result in under-recognition of early lymphedema, which could result in delay in early…
Status
Ongoing
Distribution
US Nationwide distribution in the states of IA, CA, IL, AK, NY, FL, KS, WI, MI, NC, GA, SD, KY, TX, MA, LA, AZ, RI, AR, CO, WY, OH, IN, MO, NJ, MD, CT, WA, PA, UT, VA, NM, ME, MN, DC, TN, NE, OR, ND, WV.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Nov. 1, 2023
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA has not recorded a termination for this recall; FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Impedimed Limited
Product
SOZO Bilateral Arm L-Dex Software
Reason
Bilateral L-Dex assessment software does not have the same level of sensitivity to help detect early signs of lymphedema as the unilateral arm L-Dex assessment, which could result in under-recognition of early lymphedema, which could result in delay in early intervention, and more aggressive intervention.
Distribution pattern
US Nationwide distribution in the states of IA, CA, IL, AK, NY, FL, KS, WI, MI, NC, GA, SD, KY, TX, MA, LA, AZ, RI, AR, CO, WY, OH, IN, MO, NJ, MD, CT, WA, PA, UT, VA, NM, ME, MN, DC, TN, NE, OR, ND, WV.
Quantity
354
Codes and lots
UDI-DI: B277SFT0250. Software v4.1 and v5.0
Recall started
Aug. 17, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Monitor, Extracellular Fluid, Lymphedema, Extremity
Root cause (FDA)
Device Design
What the company told customers
On 8/17/23, correction notices were emailed to customers. Customers were asked to take the following actions: We recommend that you confirm the risk profile of your bilateral arm patients in MySOZO and take the following actions: - For patients at unilateral risk for lymphedema (over 90% of breast cancer patients), confirm that the unilateral arm L-Dex assessment is selected in the patient profile and continue to screen for early signs of lymphedema. - For patients at bilateral risk for lymphedema (3-10% of breast cancer patients), guidelines suggest routine screening using clinical exam and symptom assessment. - Complete and return the response form via link: https://forms.office.com/r/kafiuDbw21 In September, the firm will be issuing a SOZO software update to make unilateral L-Dex the sole assessment option for arms. L-Dex assessments for legs will not be impacted. This correction letter should be shared with anyone within your organization who needs to be aware of the correction or with any organization where the device may have been transferred. If you have questions or need to report quality issues, call 1/877/247-0111, Option 4 (Mon-Fri, 8:00 am - 8:00 om ET) or email impedimed156@impedimed.com
Recall number
Z-0155-2024

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device, the reason the record gives, the record's wordsA simplified drawing of a medical device with a mark for the reason the record gives. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: SOZO Bilateral Arm L-Dex Software2 The record describes a software issue3 What the record says4 What the record lists!!FOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepsoftware

“Bilateral L-Dex assessment software does not have the same level of sensitivity to help detect early signs of lymphedema as the unilateral arm L-Dex assessment, which could result in under-recognition of early lymphedema, which could result in delay in early intervention, and more aggressive intervention.”SOZO Bilateral Arm L-Dex Software, FDA enforcement report, 2023-11-01. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Monitor, Extracellular Fluid, Lymphedema, Extremity

Distributed in (FDA's record): Alaska Arizona Arkansas California Colorado Connecticut District of Columbia Florida Georgia Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Missouri Nebraska New Jersey New Mexico New York North Carolina North Dakota Ohio Oregon Pennsylvania Rhode Island South Dakota Tennessee Texas Utah Virginia Washington West Virginia Wisconsin Wyoming and nationwide

This recall among the others

  • FDA's enforcement report came 76 days after the firm started the recall. For half of the 835 FDA medical device recall events reported in 2023 with a start date, it took 51 days or fewer (by year).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-0155-2024. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-0155-2024, Nov. 1, 2023. "SOZO Bilateral Arm L-Dex Software." PlainRecords, https://plainrecords.com/recall/device/z-0155-2024/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance