Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.
Filed by: Stryker Leibinger GmbH & Co. KG · Also listed under Stryker Leibinger
Nov. 4, 2026Orthoscan VERSA Accessory Cart…Class IIThe accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure… Recall began Sept. 3, 2026
Nov. 4, 2026OmniaSecure Model 3930M…Class IIAffected leads require additional engineering evaluation due to complaints regarding difficulty in achieving fixation during implantation. Potential… Recall began Sept. 17, 2026
Oct. 28, 2026PNE Lead Implant Kit REF: 1701Class IIRetroactively Reported: 10mL luer lock syringes included in kit may leak , break or experience other issues that may pose a risk to users. Recall began July 15, 2024
Oct. 28, 2026Medtronic NDI Passive Spheres REF: 8801071… and 3 more productsClass IISterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to… Recall began Aug. 31, 2026
Advertisement
Oct. 28, 2026Philips Lumify Diagnostic Ultrasound SystemClass IUltrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which… Recall began Sept. 24, 2026
Oct. 21, 2026LeadCare II Blood Lead Test Kit…Class IIIIn kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the… Recall began Sept. 4, 2026
Oct. 21, 2026DxU 840m Iris… and 1 more productClass IIThe rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage… Recall began Aug. 12, 2026
Oct. 21, 2026JenaValve Trilogy Delivery Catheter… and 2 more productsClass IIDelivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device. Recall began July 31, 2026
Oct. 14, 2026SEDIPLAST…Class IIDuring use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause… Recall began Aug. 25, 2026
Oct. 14, 2026Battle Bandage…Class IICompression bandage may be degraded, which may cause unwanted foreign material to be introduced inside the wound, which may lead to wound exploration… Recall began Sept. 2, 2026
Oct. 14, 2026Spectral CT and Spectral CT 7500 RT…Class IISoftware issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or… Recall began Aug. 21, 2026
Oct. 7, 2026Intrapulmonary Percussive Ventilation… and 6 more productsClass IILabeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV… Recall began Aug. 18, 2026
Oct. 7, 2026Computed Tomography X-ray System… and 11 more productsClass IIPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS… Recall began Aug. 11, 2026
Oct. 7, 2026McKesson CONSULT hCG Urine Cassette… and 3 more productsClass IIhCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result. Recall began Aug. 20, 2026
Oct. 7, 2026HeartSine samaritan PAD (Public Access Defibrillator)…Class IIIProducts were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check. Recall began July 27, 2026
Oct. 7, 2026VentStar Resus Neo…Class IPotential cracks in the hose, which could lead to leakage or the hose falling off. Recall began Sept. 3, 2026
Oct. 7, 2026NovaSeq 6000Dx including DRAGEN Server…Class IIRetrospectively reported correction for an instrument intended for targeted sequencing of DNA libraries from human genomic DNA used with in vitro… Recall began Sept. 17, 2024
Sept. 23, 2026Reign Medical Clench Staple System (Compression)…Class IIBone fixation staple requires a new 510(k); the performance characteristics of this device have not been adequately established. Recall began July 15, 2026
Sept. 23, 2026Happiest Baby… and 1 more productClass IIExtra small (ES) and Extra-large (XL) infant sleep sacks have not been reviewed for safety and effectiveness by the FDA. Recall began Aug. 7, 2026
Sept. 23, 2026GM Helix Implant…Class IIThere is a discrepancy between the product identification on the packaging and the dimensions of the implant contained inside. Recall began Aug. 12, 2026
Sept. 23, 2026Plum administration sets and 7 more productsClass IIIntravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored… Recall began May 12, 2026
Sept. 16, 2026Alfalfa…Class IPotential E. coli and Salmonella contamination
Sept. 16, 2026HeartSine SAM 350P 350P US Device…Class IIDue to a damaged (over stretched) membrane tail that causes an intermittent fault. Recall began July 27, 2026
Sept. 9, 2026Baxter Novum IQ Syringe Pump… and 1 more productClass IIA software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be…
Sept. 9, 2026HEALON EndoCoat PRO…Class IISurgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is… Recall began July 31, 2026
Sept. 2, 2026Medline medical convenience kits… and 3 more productsClass IISpecific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to… Recall began July 8, 2026
Aug. 26, 2026Hamilton Microlab STAR Liquid Handler…Class IISoftware issue that can cause an increase in spontaneous software closures. Recall began June 10, 2026
Aug. 26, 2026BD Needle Kit for Powered Driver 15mm x 15Ga…Class IAfter placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically…
Aug. 19, 2026ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000Class IIA strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector… Recall began June 26, 2026
Aug. 19, 2026OBP Medical…Class IIThere is a potential for vaginal speculums to contain a burr on the proximal plastic blade, which may introduce a sharp edge in the contact regions… Recall began July 13, 2026
Aug. 19, 2026GM Helix Implant…Class IIThere may be a mismatch between the implant contained inside the packaging and the information described on the outer carton label. If the the… Recall began July 13, 2026
Aug. 19, 2026navify Digital Pathology (on-prem) v2.5…Class IIDue to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window Recall began April 20, 2026
Aug. 12, 2026REF # D23352… and 1 more productClass IIAnalyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications." Recall began June 26, 2026
July 29, 2026Flexible Hystero-Fiberscope Kits REF: 11264BBU1…Class IIFlexible hysteroscopes indicated for use during diagnostic and operative hysteroscopic procedures may contain an expired single use endoscopic seal… Recall began May 21, 2026
July 29, 2026MiniMed Go App…Class IIFor MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the… Recall began June 16, 2026
July 29, 2026Boston Scientific TheraSphere 360 Y-90 Management Platform…Class IISoftware anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more… Recall began June 19, 2026
July 22, 2026da Vinci SP Instrument Arm Drape…Class IIRobotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the… Recall began May 12, 2025
July 22, 2026Abiomed REF: 0052-3025… and 2 more productsClass IDue to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along… Recall began June 10, 2026
July 15, 2026GSP Neonatal GALT Kit…Class IIReduction in the on-board stability of the GSP GALT assay. This may increase the likelihood of run-QC failures and consequently delay the reporting… Recall began May 18, 2026
July 15, 20265008X Blood Tubing Set…Class IIPotential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the… Recall began May 11, 2026
July 15, 2026Intracept Access Instruments…Class IIIntraosseous nerve ablation system instruments may have sterile barrier breach of the outer tray seal, which may introduce a risk of sterility… Recall began April 22, 2026
July 8, 2026Radius VSM ECG pre-connected Set… and 1 more productClass IITwo devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger… Recall began Feb. 20, 2026
July 8, 2026OTP5000 THORA/PARA TRAY… and 13 more productsClass IIDue to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled… Recall began May 21, 2026
July 8, 2026IntelliCuff…Class IIDue to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to… Recall began May 22, 2026
July 1, 2026Philips Telemetry Monitor 5500 1.4 GHz…Class IIPotential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a…
July 1, 2026Surgify Halo… and 18 more productsClass IIPotential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS). Recall began April 30, 2026
July 1, 2026Thuvera Laser Console…Class IISoftware anomaly that causes a false display of error code 5018. Recall began May 5, 2026
July 1, 2026INZONE DETACHMENT SYSTEM…Class IIEmbolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint… Recall began April 6, 2026
July 1, 2026TMJ Bilateral Implants…Class IICustom temporomandibular joint implant may have contained incorrect patient-specific components that led to a poor fit of the implant, leading to a… Recall began Feb. 11, 2026
July 1, 2026Da Vinci ASSY…Class IIOperative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake… Recall began April 22, 2026
June 24, 2026iLet Bionic Pancreas…Class IIThe reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure… Recall began Sept. 23, 2025
June 24, 2026Mito Red Light…Class IIThe near-infrared therapy device contains a lithium-ion battery that may malfunction, causing it to catch fire or burst into flames. Recall began Feb. 12, 2025
June 24, 2026PARACENTESIS KIT… and 34 more productsClass ILidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have… Recall began April 21, 2026
June 17, 2026Plum Solo Precision IV Pump… and 2 more productsClass IIUnder certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming… Recall began April 29, 2026
June 10, 20261) Cheese Sauce Mix… and 3 more productsClass IProduct contains dry milk powder ingredient recalled by the supplier due to potential Salmonella contamination Recall began May 6, 2026
June 10, 2026Digital Angiography System Trinias:UD150G-40 / D150GC-40…Class IIElectromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off. Recall began April 20, 2026
June 10, 2026Arvis Hip & Knee 3.0 Instrument Set… and 1 more productClass IIThe V-Block Assembly and Trackers A, B, C and E may exhibit signs of early wear and/or corrosion on the exposed magnet face. Recall began April 23, 2026
June 10, 2026Atlan A350… and 1 more productClass IThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation… Recall began May 7, 2026
June 3, 2026Automated Impella Controller (AIC)…Class IA retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware… Recall began April 20, 2026
May 27, 2026Oncology Kit w/60" (152 cm) Appx 2.2 ml…Class IIOncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if… Recall began March 24, 2026
May 27, 2026Swan-Ganz Catheter… and 2 more productsClass IICatheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to… Recall began April 8, 2026
May 13, 2026LIFEPAK 15…Class IISome defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure… Recall began April 13, 2026
May 6, 2026Brand Name: Leksell Vantage Arc System REF: 1053958Class IILocking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures. Recall began March 19, 2026
April 29, 2026Sequencing Agent SEQ0067…Class IIPossible increase in phasing, potentially resulting in false-positive BARD1 splice site indel artifact variants Recall began March 20, 2026
April 29, 2026Flamingo Funnel Small… and 2 more productsClass IICertain lots of product were associated with a manufacturing nonconformance involving use of a non-gamma sterilization compatible drape. Recall began March 16, 2026
April 22, 2026Gentuity HF-OCT Imaging System Model G10-01…Class IIThe potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx… Recall began March 23, 2026
April 22, 2026iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision SoftwareClass IIit was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI)… Recall began Aug. 13, 2025
April 15, 2026Tornier HRS Max parts: 1…Class IIAffected products were incorrectly labeled as Tornier HRS Max products. These parts are compatible with the Tornier HRS system only, and are… Recall began March 5, 2026
April 8, 2026SIGNA Premier systemsClass IIWhole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with… Recall began Feb. 27, 2026
April 8, 2026MiniMed 670G Insulin Pump (MMT-1760… and 13 more productsClass IIAll Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and… Recall began Feb. 13, 2026
April 8, 2026Airway Exam Kit… and 6 more productsClass IIKits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the… Recall began Feb. 16, 2026
April 8, 2026Brand Name: da Vinci S… and 1 more productClass IIDue to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments Recall began Dec. 9, 2025
April 8, 2026Brand Name: da Vinci S… and 3 more productsClass IIDue to increased complaints for broken/frayed grip cables for reusable instruments. Recall began Dec. 9, 2025
April 1, 2026SAVANNAH BEE COMPANY HONEY BBQ SAUCE MUSTARD NET 16 FL OZ (453g)…Class IIUndeclared Allergens: Soy & Wheat. The firm was notified by a customer that the product is labeled as Savannah Bee Company Honey BBQ Sauce "Mustard"… Recall began Feb. 27, 2026
April 1, 2026MiniMed" 780G…Class IIThree software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions… Recall began Jan. 30, 2026
April 1, 2026CentrosFLO Hemodialysis Catheters… and 4 more productsClass I16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies… Recall began Feb. 13, 2026
March 25, 2026Dog Food… and 4 more productsClass IProduct tested positive for positive for one or more types of pathogenic (harmful) bacteria, specifically: Listeria monocytogenes, Salmonella… Recall began Jan. 26, 2026
March 25, 2026Brand Name: Welch Allyn…Class IIDisplay screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in… Recall began Feb. 18, 2026
March 18, 2026Raz-AT (Attendant Tilt) Mobile Shower Commode Chair…Class IIDue to a modification by a vendor and may be reinstalled without properly engaging the outside rear seat brackets that are used to keep the seat in… Recall began Jan. 5, 2026
March 11, 2026Gelsoft Plus…Class IIGelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life.… Recall began Dec. 18, 2025
Feb. 25, 2026Da Vinci 5 ASSY…Class IIAn error event following the release of a software version for the surgical system that can result in loss of user interface content on an external… Recall began Nov. 17, 2025
Feb. 25, 2026MOLLI 2 System System includes: MM1000 (Pack of 10)…Class IIDue to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during… Recall began Jan. 21, 2026
Feb. 25, 2026MONARCH Platform…Class IIDevice for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to… Recall began Jan. 21, 2026
Feb. 18, 2026CREAMY PEANUT BUTTER… and 2 more productsClass IIForeign material. During production, the firm found pieces of blue plastic in a filter. Recall began April 30, 2025
Feb. 18, 2026da Vinci 5 Surgeon Console Viewer Display…Class IIRobotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field… Recall began Dec. 30, 2025
Feb. 18, 2026Brilliance iCT… and 3 more productsClass IIAs part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these… Recall began Jan. 7, 2026
Feb. 11, 2026Willy Pete's Chocolates brand Almond Despair…Class IIIAlmonds are not declared in the ingredient statement. Almond is included in description of product on front label. Recall began Dec. 22, 2025
Feb. 11, 2026Volk Single Use Vitrectomy Lenses 1… and 12 more productsClass IIThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has… Recall began Dec. 16, 2025
Feb. 11, 2026Focalyx FusionClass IISoftware device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not… Recall began Dec. 23, 2025
Feb. 11, 2026CellMek SPS Sample Preparation System…Class IISample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or… Recall began Dec. 23, 2025
Feb. 11, 2026EMPOWR 3D Knee Tibial Insert 5R 14MM… and 1 more productClass II942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices… Recall began Jan. 5, 2026
Feb. 11, 2026COULTER DxH Diluent…Class IIDue to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers. Recall began Nov. 10, 2025
Feb. 4, 2026Qunol…Class IIMold contamination detected for raw material lots that were produced in July and used in finished goods. Recall began Dec. 16, 2025
Feb. 4, 2026BEAR Implant…Class IIIncorrect expiration date on label that extends beyond the product's documented/approved shelf life. With an incorrect expiration date, the UDI will… Recall began Dec. 31, 2025
Feb. 4, 2026Evolution Upgrade 1.5T… and 12 more productsClass IIThe potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication… Recall began Dec. 3, 2025
Jan. 28, 2026SQ40S Blood Transfusion FilterClass IIDue to products being released for distribution prior to completion of the required quality control release process the sterility assurance cannot be… Recall began Nov. 3, 2025
Jan. 28, 2026The LDL-Cholesterol assay is comprised of two distinct phases…Class IIBeckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL… Recall began Dec. 8, 2025
Jan. 28, 2026Tego Connector: REF: 011-D1000…Class IINeedle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a… Recall began Dec. 19, 2025
Jan. 21, 2026Pulsator Arterial Blood Sampling Kit… and 1 more productClass IIAffected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may… Recall began Dec. 19, 2025
Jan. 21, 2026CentriMag Blood Pump…Class IIAcute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the… Recall began Oct. 10, 2025
Jan. 21, 2026da Vinci SP…Class IIAccess Port System tray may develop cracks potentially resulting in a sterility breach. Recall began Dec. 15, 2025
Jan. 21, 2026Asserachrom HPIA…Class IIPotential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results. Recall began Dec. 15, 2025
Jan. 14, 2026Codman Microsensor Basic Kit… and 1 more productClass IIPotential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits. Recall began Dec. 12, 2025
Jan. 14, 2026BD Pyxis ES Enterprise Server Catalog Numbers 1115-00…Class IIDue a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems. Recall began Dec. 3, 2025
Jan. 14, 2026Philips Tempus Pro Patient Monitor…Class IIPatient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for… Recall began Nov. 26, 2025
Dec. 31, 2025da Vinci 5…Class IISurgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable… Recall began Aug. 21, 2024
Dec. 31, 2025BD Pyxis MedStation ES…Class IIAutomated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to… Recall began Nov. 20, 2025
Dec. 24, 2025Chef brand Milk Pan 24cmClass IIProducts found to contain leachable lead levels (3.00 mg/L) Recall began Nov. 19, 2025
Dec. 24, 2025Alinity ci-series System Control Module (SCM)…Class IIAbbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the… Recall began Nov. 13, 2025
Dec. 24, 2025InPen App…Class IIWhen app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't… Recall began Nov. 13, 2025
Dec. 24, 2025Intelerad InteleShare software…Class IISoftware intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are… Recall began Nov. 18, 2025
Dec. 24, 20255008X CAREsystem +CLiC +CDXClass IISeveral software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if… Recall began Oct. 25, 2025
Dec. 17, 2025Ganirelix Acetate Injection…Class IIFailed impurities/degradation specifications: out of specification results for Ganirelix acetate acrylic acid adduct impurity. Recall began Nov. 13, 2025
Dec. 17, 2025DEXLOCK Achilles Repair Implant Kits…Class IIThere have been multiple complaints of drill bit fusing to the bushing/sleeve within the drill guide during use. Retrieving a replacement drill bit… Recall began Nov. 11, 2025
Dec. 10, 2025Haloperidol Lactate Injection…Class IILabeling: Not Elsewhere Classified:Incorrect RFID tag labels applied to product by repackaging firm. Recall began Oct. 10, 2025
Dec. 10, 2025Lumify Diagnostic Ultrasound System models : 795005…Class IIUltrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining. Recall began Oct. 31, 2025
Dec. 10, 2025InSure ONE Fecal Immunochemical Test…Class IIInSure ONE failed sensitivity testing when used with Test Card Lot T241102, which may result in false negative results. Recall began Sept. 17, 2025
Dec. 3, 2025Cold Form Wrap: Back…Class IIThe IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may… Recall began Sept. 25, 2025
Dec. 3, 2025CareLink Clinic…Class IISoftware error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based… Recall began Oct. 21, 2025
Nov. 26, 2025Brand Name: ACUSON… and 1 more productClass IIAffected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead… Recall began Oct. 8, 2025
Nov. 26, 2025MONARCH Bronchoscope…Class IIPotential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible… Recall began Oct. 18, 2025
Nov. 26, 2025PhD Hemostasis Valve…Class IIHemostasis valve has a manufacturing defect and when used the valve cap may remain depressed, leading to potential leakage, which may result in… Recall began Aug. 4, 2025
Nov. 26, 2025BD Luer Tip Caps…Class IIBD has confirmed through internal testing that certain device trays failed routine biocompatibility testing. Recall began Oct. 14, 2025
Nov. 19, 2025CEREPAK Detachable Coil System…Class IDetachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional… Recall began Oct. 2, 2025
Nov. 12, 2025Frozen Crab Paste…Class IUndeclared allergen ingredients (fish, soy, and wheat) Recall began Oct. 3, 2025
Nov. 12, 2025SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent…Class IIBeckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166… Recall began Sept. 17, 2025
Nov. 5, 2025MIXED NUTS Roasted & Salted… and 1 more productClass IIUndeclared Tree Nut. The firm discovered that Hazelnuts was not listed in the allergen statement Recall began Sept. 15, 2025
Nov. 5, 2025CellMek SPS Sample Preparation System…Class IIAutomated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead… Recall began Sept. 8, 2025
Nov. 5, 2025SAFE-T-FILL Micro Capillary Blood Collection…Class IIAll RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when… Recall began May 5, 2025
Nov. 5, 2025ChemoLock Bag Spike… and 37 more productsClass IIPort weld of drug transfer device may separate or break during use and potentially result in a leak. Recall began Sept. 8, 2025
Nov. 5, 2025Alcon PRECISION1 One-Day Contact Lenses…Class IIPotential for a weak seal in some units resulting in compromise in sterility. Recall began Sept. 30, 2025
Oct. 29, 2025Arctic Shores… and 4 more productsClass IIProduct manufactured under insanitary conditions whereby it may have been contaminated with Cesium-137 (Cs-137). This is an expansion to the original… Recall began Sept. 23, 2025
Oct. 29, 2025Floria Dried Apricots…Class IProduct contains undeclared sulfites. Recall began April 29, 2025
Oct. 29, 2025LOW PROF LCKNG SCREW… and 1 more productClass IIAlthough the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged… Recall began Sept. 26, 2025
Oct. 22, 2025Airvo 2…Class IIWhen unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound… Recall began Sept. 18, 2025
Oct. 22, 2025XERF EFFECTOR 60…Class IIManufacturing defects in devices manufactured from 4/1/25 to 8/20/25 can on rare occasions result in adverse events for patients. Recall began Sept. 4, 2025
Oct. 22, 2025Zenith Alpha 2 Thoracic Endovascular Graft…Class IAffected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular… Recall began Sept. 18, 2025
Oct. 15, 2025Kirkland Signature brand Ahi Tuna Wasabi Poke…Class IListeria monocytogenes. Recall was initiated after being notified by green onion supplier of a Listeria monocytogenes positive test result in the…
Oct. 15, 2025FDR Visionary Suite…Class IIIt was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not… Recall began Sept. 15, 2025
Oct. 15, 2025CorNeat EverPatch… and 1 more productClass IIReviewed complaints have identified that conjunctival wound dehiscence may result in early exposure of the surgical patch. Recall began Aug. 18, 2025
Oct. 8, 2025Venclose digiRF Generators…Class IISoftware version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring… Recall began Aug. 21, 2025
Oct. 8, 2025JUNCTIONAL EMERGENCY TREATMENT TOOL (JETT)…Class IIDevice used for unilateral or bilateral occlusion of blood flow to the lower limbs has a potential manufacturing defect that may compromise the… Recall began July 1, 2025
Oct. 8, 2025GORE ACUSEAL Vascular Graft…Class IIReports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to… Recall began Aug. 26, 2025
Oct. 8, 2025Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616…Class IIThe bovine carotid arteries used to produce the impacted grafts from affected lots were sourced using raw material from a supplier that had not been… Recall began Sept. 2, 2025
Oct. 8, 2025IMPLANT EPIKUT PLUS MT16 3,5X10,0MM…Class IIDental implants were imported under temperature and storage conditions that do not comply with the instructions for use and product labeling so there… Recall began July 30, 2025
Oct. 8, 2025Getinge 88-Series Washer- Disinfector…Class IIPotential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and… Recall began Aug. 29, 2025
Oct. 1, 2025Extractor Pro RX Retrieval Balloon Catheter…Class IIThe product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and… Recall began Aug. 28, 2025
Oct. 1, 2025FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative…Class IIFilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead… Recall began July 10, 2025
Sept. 24, 2025(Bulk) Oyster…Class IIOyster product potentially contaminated with Norovirus Recall began Aug. 7, 2025
Sept. 24, 2025Sand Bar… and 1 more productClass IIProduct manufactured under insanitary conditions whereby it may have been contaminated with Cesium-137 (Cs-137) Recall began Aug. 19, 2025
Sept. 24, 2025SureStep Foley Tray System Bardex I.C… and 2 more productsClass IIFoley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection… Recall began July 17, 2025
Sept. 24, 2025Exactech Equinoxe… and 1 more productClass IIReverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling. Recall began Aug. 15, 2025
Sept. 24, 2025Vercise Deep Brain Stimulation Systems Surgical Implant Manual…Class IIWire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire… Recall began July 8, 2025
Sept. 17, 2025QuickVue: Dipstick Strep A Test: 50T…Class IIDipstick strep A test has potential for false positive results. Recall began June 17, 2025
Sept. 17, 2025CorNeat EverPatch…Class IIRetroactive; Due to complaints, labeling correction for permanent tissue-integrating surgical matrices to include prevention and management of would… Recall began Oct. 16, 2024
Sept. 10, 2025Hans Kissle Red Bliss Potato SaladClass IIContains undeclared wheat. The top label identifies the product as potato salad, but the unit contains pasta salad. Recall began Aug. 4, 2025
Sept. 10, 2025Medshape Universal Joints…Class IIBone external fixation system may have the pin that can become dislodged resulting in frame instability, the use of this product may cause a delay in… Recall began June 2, 2025
Sept. 10, 2025Superion IDS Kit Superion Indirect Decompression System 8MM…Class IIAs a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected… Recall began June 6, 2025
Sept. 3, 2025BD Pyxis Product Name / UDI-DI code / Catalog No…Class IIAntivirus software was not consistently installed on impacted devices during the implementation process. Recall began May 8, 2025
Sept. 3, 2025Bardex Lubri-Sil I.C…Class IIFoley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or abdominal… Recall began June 18, 2025
Sept. 3, 2025BD PhaSeal Injector Luer (N30C)…Class IIClosed system drug transfer devices were shipped to the U.S. market without a U.S. version of the instructions for use, which could be an issue if a… Recall began June 16, 2025
Aug. 27, 2025Sri Sri Tattva…Class IIProduct contains undeclared soy lecithin.
Aug. 27, 2025Burr Hole Cover Kit…Class IIResistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of… Recall began July 8, 2025
Aug. 27, 2025Pentax Medical Video Processor…Class IIDuring endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become… Recall began July 16, 2025
Aug. 20, 2025Signature Cafe Premium Tuna Salad… and 6 more productsClass ITuna Salad products are recalled due to potential contamination with Listeria monocytogenes because products containing Newly Weds Foods recalled… Recall began July 16, 2025
Aug. 20, 2025Zvu Anorectal Manometry Procedure Kit…Class IIIAnorectal manometry procedure kit outer kit label is mislabeled with an incorrect expiration date; the manufacturing date was used in place of the… Recall began June 4, 2025
Aug. 20, 20253mensio Workstation (Vascular Fenestrated) softwareClass IIWhen fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular… Recall began July 14, 2025
Aug. 20, 2025High Pressure Tubing: REF: 502211002… and 3 more productsClass IIHigh pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use… Recall began June 5, 2025
Aug. 13, 2025One-piece Guedel airway… and 1 more productClass IIPotential contamination with small burrs, which if detach could be inhaled and result in potential complications such as airway obstruction, tissue… Recall began June 20, 2025
Aug. 13, 2025BD Trucount Controls… and 1 more productClass IIControl kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead… Recall began July 8, 2025
Aug. 13, 2025BD Alaris Pump Module Model… and 1 more productClass IInfusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy… Recall began July 8, 2025
Aug. 13, 2025Microbore Extension Set with 0.2 Micron Filter…Class ILabel is incorrect. The product label indicates that the device contains an air eliminating filter. However, the filter utilized on this extension… Recall began July 9, 2025
Aug. 6, 2025Volcano Visions Digital IVUS Catheter: PV.014P (Platinum)…Class IIReports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or… Recall began June 13, 2025
Aug. 6, 2025Precision Sampling Set…Class IISampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds… Recall began April 21, 2025
Aug. 6, 2025Phasor Drill…Class IIDrill for use in neurosurgical procedures may have a reverse battery orientation, which may increase surgical time. Recall began March 21, 2025
Aug. 6, 2025HeartSine SAM 350P…Class IIDue to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks) Recall began June 30, 2025
Aug. 6, 2025Breathing circuit set…Class IVentilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise… Recall began June 26, 2025
Aug. 6, 2025SafeStar 90 Plus Filter… and 3 more productsClass IReports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters. Recall began June 16, 2025
July 30, 2025Stryker… and 1 more productClass IIDue to complaints their is the potential that irrigation solution may leaked into the handpiece and battery pack causing it to malfunction. Recall began June 18, 2025
July 30, 2025ACUVUE OASYS MAX 1-Day MULTIFOCALClass IIDue to defects (bubbles/voids) identified during standard finished goods testing Recall began June 12, 2025
July 30, 2025HSG Procedure Tray… and 2 more productsClass IIThe HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify… Recall began June 11, 2025
July 30, 2025Azurion 7 M20…Class IIPotential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose. Recall began June 4, 2025
July 23, 2025SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve…Class IIDue to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container… Recall began June 3, 2025
July 16, 2025Brand Name: Alcon Laboratories… and 2 more productsClass IIDue to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area… Recall began April 28, 2025
July 16, 2025TrueDent Clear… and 4 more productsClass IIICustomers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will… Recall began Jan. 19, 2024
July 2, 2025FIREHOOK ARTISAN BAKED CRACKERS CLASSIC SEA SALT…Class IIUndeclared Sesame. The firm was notified by a retailer that product labeled as Sea Salt Bagel Crackers contained Everything Bagel Crackers. The… Recall began May 30, 2025
July 2, 2025Spectral CT…Class IIDevices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or… Recall began May 29, 2025
July 2, 2025Product Name [REF]: Supply Server…Class IIDuring automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then… Recall began May 8, 2025
June 25, 2025Lancaster LF Authentic Greek Salad Kit…Class ISalmonella. Greek Salad Kit made with recalled cucumbers sourced from Bedner Grower's Company which may be contaminated with Salmonella and… Recall began May 21, 2025
June 25, 20255008X HDF Hemodialysis Blood Tubing Set…Class IIPotential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD)… Recall began April 29, 2025
June 25, 2025ICU Medical Plum Duo… and 1 more productClass IIInternal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The… Recall began April 22, 2025
June 18, 2025Allon 2001…Class IIAn investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The… Recall began May 13, 2025
June 18, 2025SCD Comfort Knee Length Small… and 2 more productsClass IIIncorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy. Recall began May 13, 2025
June 18, 2025Newport HT70 Ventilator…Class IVentilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down…
June 11, 2025AMCE GREEK SALAD AUTHENTIC FS…Class ISalmonella. Greek Salad products were made with recalled cucumbers from Bedner Grower's Company which are contaminated with Salmonella and implicated…
June 4, 2025Catalog No… and 5 more productsClass IIPackaging defect may compromise sterility of the product. Use of a non-sterile product on the patient could lead to infection in neonates. Recall began May 2, 2025
June 4, 2025GM Helix Acqua Implant… and 1 more productClass IIIt is possible that a package labelled as an 11.5 mm implant may contain a 13 mm implant instead. Recall began May 1, 2025
June 4, 2025DORO Easy-Connect Navigation Adaptor…Class IIPotential of compromised compatibility resulting in influence on the positioning of the reference array and an influence of the function of the… Recall began March 14, 2025
June 4, 2025Atellica IM Homocysteine Calibrator (2 Pack)… and 1 more productClass IICertain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference… Recall began April 15, 2025
May 28, 2025PhenoMATRIXClass IIAI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo)…
May 28, 2025META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM…Class IIIntegrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional… Recall began March 18, 2025
May 28, 2025True 3D Viewer…Class IIWith certain software versions, import of enhanced ultrasound formatted data/ultrasound images into 3D viewer, (for digital image… Recall began Feb. 21, 2018
May 21, 2025DxC 500 AU Clinical Chemistry Analyzer…Class IIBeckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested… Recall began March 20, 2025
May 21, 2025LifeShield Drug Library Management (DLM)…Class IIFirm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate…
May 14, 2025Philips Tempus Pro Patient Monitor…Class IIPatient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required…
May 7, 2025Enhanced Verify Evaluation Handset (CFN HH90130FA)Class IIEvaluation handsets may not be able to communicate with the neurostimulator due to handsets having a faster processing speed than prior handsets. An… Recall began April 3, 2025
May 7, 20253.0 Liter Disposable Set (4 sets/case)…Class IIPotential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during… Recall began March 31, 2025
May 7, 2025Hyperthermia Pump Procedure Kit… and 2 more productsClass IIPotential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of… Recall began March 31, 2025
May 7, 2025DxC 500 AU Clinical Chemistry Analyzer…Class IIThe reason for the recall is incorrect sample barcode identification and processing errors that result in the reporting of erroneous and delayed… Recall began March 17, 2025
April 30, 2025Tempus Pro Patient Monitor…Class IIsoftware issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
April 30, 2025Skytron Freedom 4FXS-60 Heavy Duty Monitor MountClass IIThe mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the… Recall began March 27, 2025
April 30, 2025Radiometer ABL90 Series…Class IIABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer… Recall began March 21, 2025
April 30, 2025InPen smart insulin pen…Class IIInsulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the… Recall began March 4, 2025
April 30, 2025Merit Medical… and 1 more productClass IIUpdate to IFU to provide additional information related to risk with device over-rotation and multiple deployment of fasteners. Recall began March 14, 2025
April 30, 2025BD PYXIS MEDBANK (Name…Class IIUsers trying to restock a single bin location of an automated dispensing cabinet, during pocket exchange restocking workflow, may experience a… Recall began March 11, 2025
April 30, 2025NIM Vital Nerve Monitoring System: CONSOLE NIM4CM01 NIM 4.0…Class IINerve Monitoring System with certain software versions has potential for increased stimulus artifact, which may require troubleshooting, and may lead… Recall began March 10, 2025
April 23, 2025HeartMate Mobile Power Unit: REF: 107754…Class ILeft Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or… Recall began March 13, 2025
April 23, 2025MEDFLEX…Class IIIf "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first… Recall began Oct. 5, 2023
April 23, 2025BD 4 Fr Single-Lumen PowerPICC Catheters…Class ICatheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher)… Recall began March 11, 2025
April 16, 2025Dried Cranberries (21.75 lb/case…Class IIMisbranded - label states 2g total carbohydrates instead of 27g Recall began March 17, 2025
April 16, 2025BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower…Class IIDue to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect… Recall began March 11, 2025
April 16, 2025GenetiSure Dx Labeling Kit…Class IIDNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot…
April 9, 2025DxC 500 AU Clinical Chemistry Analyzer…Class IIBeckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent… Recall began Feb. 26, 2025
April 9, 2025BD Connecta BD Luer-Lok 360…Class IIDue to specific lots of luer-lok devices were shipped to the U.S. market with an Outside US Instruction for Use (IFU) Recall began Feb. 20, 2025
April 9, 2025Boppli: Bedside Device Kit…Class IISingle-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in… Recall began Feb. 24, 2025
April 2, 2025NexGen LPS Flex…Class IIThe "Use with plate 7, 8, 9, 10" compatibility statement on top of the label does not match with the compatibility statements "USE WITH PLATE 5, 6"… Recall began Feb. 27, 2025
April 2, 2025CVAC Aspiration System…Class IAspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions… Recall began Feb. 19, 2025
March 26, 2025Synapse PACS Software Version 7.4.x…Class IIThe incorrect computed patient age is showing in VX for patients less than 3 months old. Recall began Feb. 11, 2025
March 26, 2025BD Pyxis: Anesthesia Station ES…Class IISync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount… Recall began Feb. 6, 2025
March 12, 2025BD BBL TB Stain Kit K…Class IIOne component of the kit was shipped with an expiration date of 30Sep2024 while the Certificate of Analysis (COA) indicated an expiration date of…
March 12, 2025Paradigm REF: MMT-712 and MMT-715 and 11 more productsClass IIAll Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin… Recall began Jan. 31, 2025
March 12, 2025Philips IntelliSpace Cardiovascular software…Class IIStudy data is not able to be archived, copied, or exported with the cardiovascular software version. Recall began April 4, 2024
March 12, 2025Access Erythropoietin (EPO)…Class IIAffected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested… Recall began Feb. 3, 2025
March 12, 2025Rotarex Atherectomy System and Instructions for Use…Class IAtherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter… Recall began Feb. 5, 2025
March 5, 2025DxC 500 AU Clinical Chemistry Analyzer…Class IIBeckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates… Recall began Jan. 10, 2025
March 5, 2025BD Pyxis: MedFlex…Class IIAutomated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential… Recall began Jan. 10, 2025
Feb. 26, 2025DeRoyal AMG Suction Connector Tubing REF: 71-2083…Class IIThe tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in… Recall began Jan. 15, 2025
Feb. 19, 2025Biological…Class IISpore test strip monitoring system, designed for routine monitoring of sterilization processes, may have a printing error on the return envelope that… Recall began Dec. 11, 2024
Feb. 12, 2025Temporary Titanium AbutmentsClass IICertain definitive dental implant restoration cases may have included a temporary titanium abutment (cylinder), rather than a definitive titanium… Recall began Dec. 6, 2024
Feb. 12, 2025Nobel Biocare N1 TiUltra TCC RP 4.0x9mm…Class IIDue to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching… Recall began Jan. 10, 2025
Feb. 12, 2025PROLENE Polypropylene Suture… and 2 more productsClass IISutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to… Recall began Dec. 20, 2024
Feb. 12, 2025SenTiva… and 1 more productClass IIVagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a… Recall began Dec. 12, 2024
Feb. 5, 2025Infinity Central Station…Class IIEdifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the…
Feb. 5, 2025Medtronic PIN…Class IIDue to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter. Recall began Dec. 4, 2024