FDA · Medical device recall

Allon 2001…

Recall started · FDA enforcement report June 18, 2025 Class II

Reason
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms…
Status
Ongoing
Distribution
Worldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM, PA, TX, VA, WA and the countries of Belgium, Bolivia (Plurinational State of), Brazil, Canada, Colombia, Israel, Italy, Portugal, Romania, Sweden, Taiwan (Province of China)…

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
June 18, 2025
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Belmont Instrument LLC
Product
Allon 2001. Part Number: 200-00263
Reason
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.
Distribution pattern
Worldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM, PA, TX, VA, WA and the countries of Belgium, Bolivia (Plurinational State of), Brazil, Canada, Colombia, Israel, Italy, Portugal, Romania, Sweden, Taiwan (Province of China), Thailand, United Kingdom.
Quantity
229 units
Codes and lots
Part Number: 200-00263. UDI-DI: 007290107581103, Serial Numbers: 99111143000002, 99111143000014, 99111627300034, 9993901600001, 9993901600002, 9993901600004, 9993901600006, 9993901600007, 9993901600009, 9995356300005, 9995356300006, 9995356300007, 9995356300008, 9996673800002, 99111143000001, 99111143000003, 99111143000004, 99111143000007, 99111143000008, 99111143000009, 99111143000011, 99111143000012, 99111143000013, 99111143000015, 99111143000018, 99111143000019, 99111288700016, 99111288700021, 99111288700035, 99111288700036, 99111627300011, 99111627300017, 99111627300022, 99111627300029, 99111627300040, 99111627300041, 99111627300045, 99111627300046, 99111627300054, 99111627300057, 99111627300064, 99111627300067, 99113168300004, 99113168300005, 99113168300008, 99113168300010, 99113168300012, 99128623400008, 99129614700002, 99129614700005, 99131046400001, 99131046400003, 99131046400004, 99131046400005, 99131046400006, 99131046400009, 99131046400011, 99131046500007, 99132732400009, 99132732400018, 99132732400019, 99134232100011, 99132121000010, 99113168300003, 9983349800008, 9991451200002, 9991451200004, 9992175000001, 9995356300010, 99111627300001, 99111627300008, 99113168300013, 99132121000014, 99132732400005, 99134232100001, 99134232100002, 99134232100003, 99134232100009, 99134232100022, 99134232100035, 99111627300042, 99111627300056, 99111627300058, 99111627300059, 99113168300014, 99111627300020, 99111627300049, 99113168300006, 99128623400003, 99132121000013, 99134232100018, 9991451200001, 99111143000006, 99111627300005, 99111627300035, 99113168300001, 99113168300002, 99113168300007, 99111627300053, 99111288700014, 99111288700017, 99111288700018, 99111288700019, 99111288700020, 99111627300016, 99111627300032, 99111143000017, 99111143000021, 99111143000022, 99111288700001, 99111288700002, 99129614700009, 9962076700011, 99132121000012, 99132121000015, 99132121000005, 99132121000007, 99111627300009, 99134232100017, 99131046400008, 9995356100003, 99113168300011, 99111627300025, 99132121000009, 99111627300061, 99111627300038, 9974083900001, 99111627300030, 9991451200006, 99111627300018, 9976534100006, 9945767700003, 9983349700008, 9991451200003, 9991451200009, 99111288700005, 99111627300031, 99128623400004, 99128623400005, 99128623400007, 99129614700001, 99129614700003, 99129614700004, 99129614700007, 99129614700008, 99131046400002, 99131046500001, 99131046500002, 99132121000001, 99128623400006, 99111143000016, 9969414000006, 99134232100013, 99134232100021, 99111627300036, 99132121000017, 99132121000018, 99132121000019, 99132121000020, 99132732400002, 99132732400006, 99132732400013, 99132732400014, 99134232100005, 99134232100010, 99132121000016, 99111627300013, 99111627300033, 99111627300043, 99111627300044, 99111627300051, 99111627300063, 99129614700006, 99111627300007, 99111627300024, 99111627300037, 99111627300047, 99111627300014, 99111627300052, 99113168300009, 99128623400001, 99128623400002, 99128623400009, 99128623400010, 99134232100008, 99111143000020, 99131046400010, 99132121000011, 99132732400007, 99132732400011, 99132732400012, 99132732400016, 99134232100034, 99132732400008, 99132732400015, 99132732400017, 99132732400020, 99134232100014, 9959060500014, 99111627300026, 99134232100004, ", 99134232100006", 9991451200005, 9995356200008, 99111143000010, 99111288700004, 99131046500003, 99131046500005, 9973536800007, 99131046500004, 99131046500008, 99131046500009, 99132121000004, 99132121000006, 99132121000008, 99134232100026, 99132732400001, 99132732400003, 99132732400004, 99134232100007, 99134232100019, 9940577300005, 99111143000005, 99111627300048, 99111627300062, 99132121000002, 9979841500003, 99132121000003, 99131046500010
Recall started
May 13, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
System, Thermal Regulating
Root cause (FDA)
Device Design
What the company told customers
Belmont Medical Technologies notified consignees via letter on about 05/13/2025. Consignees were instructed to ensure users of the devices are made aware of the correction, identify affected units on hand and check their serial numbers, ensure all devices have the Temp Sensor Alarms always set to ON to ensure that the user is alerted to errors and conditions requiring actions and attach the provided Step by Step Guide (P/N DDT-063-027) to each device. Consignees were also instructed to complete and return the response form and notify customers if the units were further distributed.
Recall number
Z-1966-2025

What this record is about

A generic icon of the kind of thing the record is about; the record's own words are below. It is not a picture of the product.

A generic icon of the kind of thing the record is aboutA simplified icon of the kind of thing the record is about and the phrases of the record's own text beside it.1 The record: Allon 2001. Part Number: 200-002632 In the record's words:3 Only the kind of thing is drawn4 The record's own words are below“An investigation into a report of the patient wrap (ThermoWrap),used with the Allon device, exceeding the set temperature…”ALLON 2001. PART NUMBER…

“An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.”Allon 2001. Part Number: 200-00263, FDA enforcement report, 2025-06-18. The caption is the record's own words.

The picture shows the kind of thing the record is about (a part, a product, a package), not the product itself; the ring marks it. Nothing in it is a claim beyond the record's own words below.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: System, Thermal Regulating

Distributed in (FDA's record): California District of Columbia Louisiana Massachusetts New Mexico Pennsylvania Texas Virginia Washington and nationwide

This recall among the others

  • FDA's enforcement report came 36 days after the firm started the recall. For half of the 972 FDA medical device recall events reported in 2025 with a start date, it took 44 days or fewer (by year).
  • Belmont Instrument: 4 other recalls or notices on this site in the three years before this one (from June 19, 2022).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-1966-2025. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-1966-2025, June 18, 2025. "Allon 2001." PlainRecords, https://plainrecords.com/recall/device/z-1966-2025/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance