Brand Name: CleverCut Product Name: Single Use 3-Lumen Sphincterotome… and 29 more products
- Reason
- Devices which did not undergo thermoforming could deform and lose performance.
- Status
- Ongoing
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, GU, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and…
- Enforcement report date
- Feb. 11, 2026
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
- Filed by
- Olympus Corporation of the Americas
- Reason
- Devices which did not undergo thermoforming could deform and lose performance.
- Distribution pattern
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, GU, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and International: TBD.
What the record says about this device
A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.
“Devices which did not undergo thermoforming could deform and lose performance.”Brand Name: CleverCut Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0320; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current., FDA enforcement report, 2026-02-11. The caption is the record's own words.
A generic device, not the product. The mark follows the record's own words.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Distributed in (FDA's record): Alabama Arizona Arkansas California Colorado Delaware District of Columbia Florida Georgia Hawaii Illinois Indiana Iowa Kentucky Louisiana Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska Nevada New Hampshire New Jersey New Mexico New York North Carolina Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Washington West Virginia Wisconsin Wyoming and nationwide
30 products in this recall
Brand Name: CleverCut Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0320…
- Product
- Brand Name: CleverCut Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0320; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 72023 units (3920 US, 68103 OUS)
- Codes and lots
- Model/Catalog Number: KD-V411M-0320; Material REF: (1) N1089430, (2) N1089410, (3) N5411130; UDI-DI: (1) 04953170183973, (2) 04953170466274, (3) 04953170380556; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1268-2026
Product Name: Single Use Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0720 Product…
- Product
- Product Name: Single Use Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 12641 units (8789 US, 3852 OUS)
- Codes and lots
- Model/Catalog Number: KD-VC431Q-0720; Material REF: (1) N5391530, (2) N5391510, (3) N5777830; UDI-DI: (1) 04953170466427, (2) 04953170466427, (3) 04953170466427; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1269-2026
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-431Q-0720 Product Description…
- Product
- Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-431Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 7689 units (7501 US, 188 OUS)
- Codes and lots
- Model/Catalog Number: KD-431Q-0720; Material REF: (1) 5859230, (2) N6221831, (3) N6221840; UDI-DI: (1) 04953170042362, (2) 04953170463778, (3) 04953170466564; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1270-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0330…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0330; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 33433 units (36 US, 33397 OUS)
- Codes and lots
- Model/Catalog Number: KD-V411M-0330; Material REF: (1) N1089530, (2) N1089510, (3) N5411230; UDI-DI: (1) 04953170380563, (2) 04953170380563, (3) 04953170380563; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1271-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0720…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 65117 units (9707 US, 55410 OUS)
- Codes and lots
- Model/Catalog Number: KD-V411M-0720; Material REF: (1) N1089610, (2) N5411330; UDI-DI: (1) 04953170184017, (2) 04953170380570; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1272-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0725…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0725; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 136037 units (9959 US, 126078 OUS)
- Codes and lots
- Model/Catalog Number: KD-V411M-0725; Material REF: (1) N1089710, (2) N5411430; UDI-DI: (1) 04953170380587, (2) 04953170380587; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1273-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 86303 units (1803 US, 84500 OUS)
- Codes and lots
- Model/Catalog Number: KD-V411M-0730; Material REF: (1) N1089810, (2) N5411530; UDI-DI: (1) 04953170380594, (2) 04953170380594; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1274-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0720…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 30489 units (11779 US, 18710 OUS)
- Codes and lots
- Model/Catalog Number: KD-V431M-0720; Material REF: (1) N1090410, (2) N1090430, (3) N5411930; UDI-DI: (1) 04953170184178, (2) 04953170184178, (3) 04953170184178; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1275-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0730…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 17655 units (2411 US, 15244 OUS)
- Codes and lots
- Model/Catalog Number: KD-V431M-0730; Material REF: (1) N1090530, (2) N1090510, (3) N5412030; UDI-DI: (1) 04953170380648, (2) 04953170380648, (3) 04953170380648; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1276-2026
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0730…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 4183 units (821 US, 3362 OUS)
- Codes and lots
- Model/Catalog Number: KD-VC431Q-0730; Material REF: (1) N5391610, (2) N5391630, (3) N5777930; UDI-DI: (1) 04953170466434, (2) 04953170466434, (3) 04953170466434; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1277-2026
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0720…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 10008 units (3038 US, 6970 OUS)
- Codes and lots
- Model/Catalog Number: KD-VC411Q-0720; Material REF: (1) N5391210, (2) N5391230, (3) N5777530; UDI-DI: (1) 04953170371110, (2) 04953170371110, (3) 04953170371110; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1278-2026
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0730…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 10257 units (274 US, 9983 OUS)
- Codes and lots
- Model/Catalog Number: KD-VC411Q-0730; Material REF: (1) N5391330, (2) N5391310, (3) N5777630; UDI-DI: (1) 04953170399329, (2) 04953170399329, (3) 04953170399329; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1279-2026
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0320…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0320; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 5946 units (2208 US, 3738 OUS)
- Codes and lots
- Model/Catalog Number: KD-VC411Q-0320; Material REF: (1) N5391030, (2) N5391010, (3) N5777330; UDI-DI: (1) 04953170466403, (2) 04953170466403, (3) 04953170466403; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1280-2026
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0330…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0330; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 5089 units (30 US, 5059 OUS)
- Codes and lots
- Model/Catalog Number: KD-VC411Q-0330; Material REF: (1) N5391130, (2) N5391110, (3) N5777430; UDI-DI: (1) 04953170371097, (2) 04953170371097, (3) 04953170371097; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1281-2026
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC412Q-0215…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC412Q-0215; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 1918 units (413 US, 1505 OUS)
- Codes and lots
- Model/Catalog Number: KD-VC412Q-0215; Material REF: (1) N5391410, (2) N5391430, (3) N5777730; UDI-DI: (1) 04953170371158, (2) 04953170371158, (3) 04953170371158; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1282-2026
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC433Q-0720…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC433Q-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 1504 units (350 US, 1154 OUS)
- Codes and lots
- Model/Catalog Number: KD-VC433Q-0720; Material REF: (1) N5391710, (2) N5391730, (3) N5778030; UDI-DI: (1) 04953170371219, (2) 04953170371219, (3) 04953170371219; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1283-2026
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number:KD-VC433Q-0730…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number:KD-VC433Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 93 units (5 US, 88 OUS)
- Codes and lots
- Model/Catalog Number: KD-VC433Q-0730; Material REF: (1) N5391830, (2) N5391810, (3) N5778130; UDI-DI: (1) 04953170371233, (2) 04953170371233, (3) 04953170371233; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1284-2026
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0720 Product Description…
- Product
- Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 125 units (7 US, 118 OUS)
- Codes and lots
- Model/Catalog Number: KD-411Q-0720; Material REF: 5858030; UDI-DI: Under Review; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1285-2026
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0730 Product Description…
- Product
- Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 887 units (650 US, 237 OUS)
- Codes and lots
- Model/Catalog Number: KD-411Q-0730; Material REFs: (1) 5858230, (2) N6221731, (3) N6221740; UDI-DI: (1)04953170466557, (2)04953170466557, (3)04953170466557; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1286-2026
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description…
- Product
- Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 886 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-401Q-0330; Material REF: 5856730; UDI-DI: 04953170041617; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1287-2026
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0720 Product Description…
- Product
- Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 1072 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-401Q-0720; Material REF: 5856830; UDI-DI: 04953170041648; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1288-2026
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description…
- Product
- Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 1490 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-401Q-0730; Material REF: 5857030; UDI-DI: 04953170041709; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1289-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 3954 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-V411M-1520; Material REF: (1) N5411630, (2) N1089910; UDI-DI: (1)04953170380600, (2) N/A; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1290-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1530 Product Description…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1530 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 7046 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-V411M-1530; Material REF: (1) N5411730, (2) N1090010; UDI-DI: (1)04953170380617, (2)N/A; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1291-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3030 Product Description…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3030 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 2344 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-V411M-3030; Material REF: (1) N5411830, (2) N1090310; UDI-DI: (1)04953170184154, (2)N/A; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1292-2026
Brand Name: Disposable Triple Lumen Sphincterotome Product Name: Disposable Triple Lumen Sphincterotome…
- Product
- Brand Name: Disposable Triple Lumen Sphincterotome Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0320 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 999 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-401Q-0320; Material REF: 5856630; UDI-DI: 04953170041587; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1293-2026
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0725 Product Description…
- Product
- Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0725 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 1980 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-401Q-0725; Material REF: 5856930; UDI-DI: 04953170041587; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1294-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3020 Product Description…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3020 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 650 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-V411M-3020; Material REF: N1090110; UDI-DI: N/A; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1295-2026
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3025 Product Description…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3025 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 1025 units (OUS only)
- Codes and lots
- Model/Catalog Number: KD-V411M-3025; Material REF: N1090210; UDI-DI: N/A; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1296-2026
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0725…
- Product
- Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0725; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.
- Quantity
- 1474 units (all OUS)
- Codes and lots
- Model/Catalog Number: KD-VC411Q-0725; Material REF: N5424710; UDI-DI: N/A; All Lots with a valid expiration date
- Recall started
- Jan. 7, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
- Root cause (FDA)
- Process control
- What the company told customers
- On or around January 7, 2026, URGENT: MEDICAL DEVICE REMOVAL letters were sent to customers. Actions to be taken: the following actions: 1. Examine your inventory and quarantine any identified devices immediately. 2. Immediately cease usage of any affected product in your inventory. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. Olympus will issue a credit to your facility upon receipt of your affected product. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0487 c. Complete the form as instructed. 5. If you have further distributed it, please forward this notice to other users who may have the affected products. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@olympus.com or by phone at (647) 999-3203.
- Recall number
- Z-1297-2026
This recall among the others
- FDA's enforcement report came 35 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).
- Olympus Corporation of the Americas: 73 other recalls or notices on this site in the three years before this one (from Feb. 12, 2023).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement reports, recall numbers Z-1268-2026, Z-1269-2026, Z-1270-2026, Z-1271-2026, Z-1272-2026, Z-1273-2026, Z-1274-2026, Z-1275-2026 and others. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-1268-2026, Feb. 11, 2026. "Brand Name: CleverCut Product Name: Single Use 3-Lumen Sphincterotome… and 29 more products." PlainRecords, https://plainrecords.com/recall/device/z-1268-2026/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections