Menu
FDA · Medical device recall

Brand Name: ACUSON… and 1 more product

Nov. 26, 2025 Class II

Date
Nov. 26, 2025
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Sterilmed, Inc.
Reason
Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.
Distribution pattern
United States Only: Kentucky, Maryland, Massachusetts, Mississippi, New Jersey, Pennsylvania, Texas, Utah

2 products in this recall

Brand Name: ACUSON…

Product
Brand Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Product Name: CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135910 Software Version: N/A Product Description: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Component: N/A
Quantity
6
Codes and lots
Lot Code: GTIN/UDI: 10888551004725 LOT# SERIAL #: 2196112 999253 2200877 937157 2219499 1182771 2219926 1048031 2220881 1126584 2234882 2304200759
Recall started
Oct. 8, 2025
Device type
Reprocessed Intravascular Ultrasound Catheter
Root cause (FDA)
Reprocessing Controls
What the company told customers
Initial phone call notification was performed 10/10/2025 to affected customers. Customers were instructed to immediately stop use of identified devices. Affected products should be segregated. Follow-up written notification was sent to customers on 10/17/25 which included instructions on how to return affected products.
Recall number
Z-0573-2026

Brand Name: Sterilmed Reprocessed Imaging Catheter Product Name: ACUSON…

Product
Brand Name: Sterilmed Reprocessed Imaging Catheter Product Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135936 Software Version: N/A Product Description: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR FOR USE WITH SIEMENS SYSTEM Component: N/A
Quantity
4
Codes and lots
Lot Code: GTIN: 10888551004732 LOT# SERIAL#: 2188147 131065 2207154 916179 2220232 1050886 2234795 1170283
Recall started
Oct. 8, 2025
Device type
Reprocessed Intravascular Ultrasound Catheter
Root cause (FDA)
Reprocessing Controls
What the company told customers
Initial phone call notification was performed 10/10/2025 to affected customers. Customers were instructed to immediately stop use of identified devices. Affected products should be segregated. Follow-up written notification was sent to customers on 10/17/25 which included instructions on how to return affected products.
Recall number
Z-0574-2026

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0573-2026, Z-0574-2026. The agency record governs; check it and the company before acting.