U.S. safety records, as the agencies wrote them. Independent.
FDA · Medical device recall

Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description…

Recall started · FDA enforcement report Aug. 13, 2025 Class III

Reason
An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31 instead of 2026-01-31). The expiration dates…
Status
Ongoing
Distribution
Worldwide - US Nationwide distribution in the states of CA, MA, PA and the countries of Germany, Hong Kong, Moldova.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Aug. 13, 2025
Risk class
Class III (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
DRG International, Inc.
Product
Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA
Reason
An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31 instead of 2026-01-31). The expiration dates on the bottle labels were correct (i.e., 2026-01-31).
Distribution pattern
Worldwide - US Nationwide distribution in the states of CA, MA, PA and the countries of Germany, Hong Kong, Moldova.
Quantity
47 units
Codes and lots
Model/Catalog Number: EIA-4568; UDI-DI: 04048474045680; Lot number: 304K034;
Recall started
June 4, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Radioimmunoassay, Total Thyroxine
Root cause (FDA)
Reprocessing Controls
What the company told customers
On June 4, 2025 FIELD SAFETY NOTICE letters were sent to customers. Action to be taken by distributors and end-users: Please check your inventory for the respective lot. Please use the CoA provided with this FSN. If you require corrected outer kit box label for your documentation, please contact DRG. If you have transferred any of the affected products to another laboratory, please provide them with a copy of this letter. If you have any questions concerning this issue, please do not hesitate to contact us.
Recall number
Z-2265-2025

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device, the reason the record gives, the record's wordsA simplified drawing of a medical device with a mark for the reason the record gives. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: Brand Name: T4 Total ELISA Product…2 The record describes an expiry date on the label that may be wrong3 What the record says4 What the record listsLABELDATE?LABELDATE?FOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by steplabel

“An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31 instead of 2026-01-31). The expiration dates on the bottle labels were correct (i.e., 2026-01-31).”Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA, FDA enforcement report, 2025-08-13. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Radioimmunoassay, Total Thyroxine

Distributed in (FDA's record): California Massachusetts Pennsylvania and nationwide

This recall among the others

  • FDA's enforcement report came 70 days after the firm started the recall. For half of the 972 FDA medical device recall events reported in 2025 with a start date, it took 44 days or fewer (by year).
  • DRG International: 3 other recalls or notices on this site in the three years before this one (from Aug. 14, 2022).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-2265-2025. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-2265-2025, Aug. 13, 2025. "Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description." PlainRecords, https://plainrecords.com/recall/device/z-2265-2025/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product