U.S. safety records, as the agencies wrote them. Independent.
FDA · Medical device recall

Dentsply Sirona Capsule Mixer Model No…

Recall started · FDA enforcement report Aug. 26, 2026 Class II

Reason
Potential for electrical discharge from unit when unplugged.
Status
Ongoing
Distribution
US: AK AL AR AZ CA CO CT DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV OUS: None

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Aug. 26, 2026
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
SDI Limited
Product
Dentsply Sirona Capsule Mixer Model No. 5546068
Reason
Potential for electrical discharge from unit when unplugged.
Distribution pattern
US: AK AL AR AZ CA CO CT DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV OUS: None
Quantity
1,444 units
Codes and lots
UDI:9336472005542; Serial Numbers 0003-1533
Recall started
July 20, 2026
Voluntary or mandated
Voluntary: Firm initiated
Device type
Amalgamator, Dental, Ac-Powered
Root cause (FDA)
Device Design
What the company told customers
On July 20, 2026, Dentsply Sirona issued a Urgent Medical Device Recall notice to affected consignees via mail. Dentsply Sirona ask consignees to take the following actions: Users 1. Switch off the unit using the rear power switch before removing the plug from the main socket 2. Avoid immediate contact with the plug pins after unplugging. 3. The users should return their unit to Dentsply Sirona to have a corrective modification installed. Please contact Racheal Elliott at rachael.elliott@dentsplysirona.com. 4. Please ensure that all relevant personnel within your organization are informed of the notice. If you have supplied affected units to another organization, please provide this with a copy of the notification. Dealers: 1.Identify all customers who have received the affected units and immediate forward the accompanying communication to those customers. 2.Ensure that all relevant personnel within your organization are informed of this notice. 3.Complete the dealer reply form provided on pages 2 and 3 and rturn it to Dentsply Sirona. 4. Return the completed form via email to Racheal Elliott at rachael.elliott@dentsplysirona.com.
Recall number
Z-2954-2026

What this record is about

A generic icon of the kind of thing the record is about; the record's own words are below. It is not a picture of the product.

A generic icon of the kind of thing the record is aboutA simplified icon of the kind of thing the record is about and the phrases of the record's own text beside it.1 The record: Dentsply Sirona Capsule Mixer Model No.…2 In the record's words:3 Only the kind of thing is drawn4 The record's own words are below“Potential for electrical discharge from unit when unplugged.”DENTSPLY SIRONA CAPSULE…

“Potential for electrical discharge from unit when unplugged.”Dentsply Sirona Capsule Mixer Model No. 5546068, FDA enforcement report, 2026-08-26. The caption is the record's own words.

The picture shows the kind of thing the record is about (a part, a product, a package), not the product itself; the ring marks it. Nothing in it is a claim beyond the record's own words below.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Amalgamator, Dental, Ac-Powered

Distributed in (FDA's record): Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska Nevada New Hampshire New Jersey New Mexico New York North Carolina Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina South Dakota Tennessee Texas Utah Vermont Virginia Washington West Virginia Wisconsin

This recall among the others

  • FDA's enforcement report came 37 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-2954-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-2954-2026, Aug. 26, 2026. "Dentsply Sirona Capsule Mixer Model No." PlainRecords, https://plainrecords.com/recall/device/z-2954-2026/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026