Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
- Reason
- The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame…
- Status
- Ongoing
- Distribution
- US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
- Enforcement report date
- April 30, 2025
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
- Filed by
- IHB OPERATIONS B.V.
- Product
- Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
- Reason
- The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
- Distribution pattern
- US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
- Quantity
- 112 units
- Codes and lots
- UDI/DI 10841736116071, All Serial Numbers
- Recall started
- March 27, 2025
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Table, Operating-Room, Ac-Powered
- Root cause (FDA)
- Process change control
- What the company told customers
- Skytron issued a Field Correction notice to its service personnel on 03/27/2025 via email. The notice explained the issue, risk to health, and requested the following: Customer Acknowledgement 1. Contact customers to ensure they have received the customer letter. 2. Provide a copy of the customer letter if needed. 3. Document who you spoke with and the date of contact on the Customer Acknowledgement Spreadsheet. Customer Acknowledgement Spreadsheet will be provided by Quality. 4. Return the completed Customer Acknowledgement Spreadsheet to quality@skytron.com. The devices will be corrected on site. Skytron notified the medical facilities (users of the device) by issuing an URGENT: MEDICAL DEVICE RECALL notice on 03/28/2025 via US Mail. This letter explained the issue, risk to health and requested the following: Actions to be taken by Customer/User: Prior to the field correction, you should take the following precautionary measures to mitigate any potential risks: Inspect your 4FXS-60 frames to ensure that all 4 mounting screws are secure. If any loose connections are identified, immediately remove the 4FXS-60 from use and notify your local authorized representative. For questions or concerns, please contact your local authorized representative or Skytron at 1-800-759-8766.
- Recall number
- Z-1663-2025
What the record says about this device
A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.
“The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.”Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount, FDA enforcement report, 2025-04-30. The caption is the record's own words.
A generic device, not the product. The mark follows the record's own words.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: Table, Operating-Room, Ac-Powered
Distributed in (FDA's record): Alabama Arizona California Florida Idaho Illinois Iowa Louisiana Maryland Massachusetts Michigan Minnesota New Hampshire New Jersey New Mexico New York North Carolina Pennsylvania Tennessee Texas Utah Virginia West Virginia Wisconsin and nationwide
This recall among the others
- FDA's enforcement report came 34 days after the firm started the recall. For half of the 972 FDA medical device recall events reported in 2025 with a start date, it took 44 days or fewer (by year).
- IHB OPERATIONS: 1 other recall or notice on this site in the three years before this one (from May 1, 2022).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement report, recall number Z-1663-2025. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-1663-2025, April 30, 2025. "Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount." PlainRecords, https://plainrecords.com/recall/device/z-1663-2025/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections