FDA · Medical device recall

Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Recall started · FDA enforcement report April 30, 2025 Class II

Reason
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame…
Status
Ongoing
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
April 30, 2025
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
IHB OPERATIONS B.V.
Product
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Reason
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Distribution pattern
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Quantity
112 units
Codes and lots
UDI/DI 10841736116071, All Serial Numbers
Recall started
March 27, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Table, Operating-Room, Ac-Powered
Root cause (FDA)
Process change control
What the company told customers
Skytron issued a Field Correction notice to its service personnel on 03/27/2025 via email. The notice explained the issue, risk to health, and requested the following: Customer Acknowledgement 1. Contact customers to ensure they have received the customer letter. 2. Provide a copy of the customer letter if needed. 3. Document who you spoke with and the date of contact on the Customer Acknowledgement Spreadsheet. Customer Acknowledgement Spreadsheet will be provided by Quality. 4. Return the completed Customer Acknowledgement Spreadsheet to quality@skytron.com. The devices will be corrected on site. Skytron notified the medical facilities (users of the device) by issuing an URGENT: MEDICAL DEVICE RECALL notice on 03/28/2025 via US Mail. This letter explained the issue, risk to health and requested the following: Actions to be taken by Customer/User: Prior to the field correction, you should take the following precautionary measures to mitigate any potential risks: Inspect your 4FXS-60 frames to ensure that all 4 mounting screws are secure. If any loose connections are identified, immediately remove the 4FXS-60 from use and notify your local authorized representative. For questions or concerns, please contact your local authorized representative or Skytron at 1-800-759-8766.
Recall number
Z-1663-2025

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device, the reason the record gives, the record's wordsA simplified drawing of a medical device with a mark for the reason the record gives. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: Skytron Freedom 4FXS-60 Heavy Duty…2 The record describes a part that can loosen or separate3 What the record says4 What the record listsFOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepseparation

“The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.”Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount, FDA enforcement report, 2025-04-30. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Table, Operating-Room, Ac-Powered

Distributed in (FDA's record): Alabama Arizona California Florida Idaho Illinois Iowa Louisiana Maryland Massachusetts Michigan Minnesota New Hampshire New Jersey New Mexico New York North Carolina Pennsylvania Tennessee Texas Utah Virginia West Virginia Wisconsin and nationwide

This recall among the others

  • FDA's enforcement report came 34 days after the firm started the recall. For half of the 972 FDA medical device recall events reported in 2025 with a start date, it took 44 days or fewer (by year).
  • IHB OPERATIONS: 1 other recall or notice on this site in the three years before this one (from May 1, 2022).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-1663-2025. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-1663-2025, April 30, 2025. "Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount." PlainRecords, https://plainrecords.com/recall/device/z-1663-2025/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance