U.S. safety records, as the agencies wrote them. Independent.
FDA · Medical device recall

Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R…

Recall started · FDA enforcement report March 11, 2026 Class II

Reason
There is the potential for the length of the trocar shaft to be too long.
Status
Ongoing
Distribution
US Nationwide distribution in the states of AZ, AR, CA, CO, FL, GA, HI, IA, ID, IL, KS, LA, MA, MD, MI, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TX, UT, WA.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
March 11, 2026
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Aesculap Inc
Product
Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Component: N/A
Reason
There is the potential for the length of the trocar shaft to be too long.
Distribution pattern
US Nationwide distribution in the states of AZ, AR, CA, CO, FL, GA, HI, IA, ID, IL, KS, LA, MA, MD, MI, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TX, UT, WA.
Quantity
126 units
Codes and lots
Mode No. FF399R; UDI-DI 04038653065547; Serial Numbers: 4595, 4619, 4657, 4764, 4592, 4596, 4598, 4601, 4606, 4607, 4608, 4612, 4616, 4618, 4620, 4621, 4623, 4659, 4661, 4662, 4667, 4670, 4673, 4674, 4688, 4689, 4746, 4748, 4757, 4758, 4760, 4761, 4762, 4767, 4768, 4771, 4773, 4774, 4779, 4780, 4784, 4786, 4789, 4791, 4792, 4795, 4799, 4811, 4812, 4818, 4819, 4821, 4822, 4824, 4825, 4826, 4827, 4836, 4838, 4841, 4842, 4843, 4854, 4856, 4858, 4864, 4868, 4869, 4880, 4882, 4886, 4890, 4891, 4892, 4893, 4899, 4901, 4902, 4907, 4914, 4921, 4927, 4940, 4951, 4959, 4967, 4968, 4970, 4971, 4972, 4973, 4974, 4975, 4976, 4990, 4994, 4995, 5000, 5004, 5015, 5024, 5035, 5036, 5042, 5057, 5058, 5059, 5062, 5063, 5064, 5072, 5076, 5081, 5084, 5106, 5112, 5115, 5117, 5118, 5119, 5121, 5123, 5165, 5166, 5168, 5172;
Recall started
Jan. 15, 2026
Voluntary or mandated
Voluntary: Firm initiated
Device type
Endoscope, Neurological
Root cause (FDA)
Under Investigation by firm
What the company told customers
on January 28, 2026 URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Review the attachment Aesculap AG Urgent Field Safety Notice in its entirety. - If it is determined that a portion of the visual field is obscured, the manufacturer recommends rotating the device to compensate for the restricted view (refer to the attached Urgent Field Safety Notice from the manufacturer). Note: Devices are not being removed from the market, this is a field notification. - Return the completed Urgent Medical Device Correction Acknowledgement Form to Aesculap, Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to recalls@bbraunusa.com within two (2) weeks of receipt. Action planned by Aesculap AG to correct the problem: To address this issue, the manufacturer is implementing additional inspections of the optical system and introducing clearly defined acceptance criteria, along with other corrective actions. Should there be any questions, comments, or concerns regarding the product you have received, please feel free to contact (844-903-6417) or recalls@bbraunusa.com
Recall number
Z-1485-2026

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device or a product, the reason the record gives, the record's wordsA simplified drawing for a device record; the mark follows the reason the record gives, not a photograph of the product. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: Brand Name: AESCULAP Product Name…2 The record describes a size or a dimension that is off3 What the record says4 What the record listslimitlimitFOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by steptoo long

“There is the potential for the length of the trocar shaft to be too long.”Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Component: N/A, FDA enforcement report, 2026-03-11. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Endoscope, Neurological

Distributed in (FDA's record): Arizona Arkansas California Colorado Florida Georgia Hawaii Idaho Illinois Iowa Kansas Louisiana Maryland Massachusetts Michigan Mississippi Missouri Nebraska New Jersey New York North Carolina Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina South Dakota Texas Utah Washington and nationwide

This recall among the others

  • FDA's enforcement report came 55 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).
  • Aesculap: 6 other recalls or notices on this site in the three years before this one (from March 12, 2023).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-1485-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-1485-2026, March 11, 2026. "Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R." PlainRecords, https://plainrecords.com/recall/device/z-1485-2026/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product