Five S 5 3x65…
- Reason
- Flexible intubation endoscopes sterility assurance can not be confirmed.
- Status
- Ongoing
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MEX, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and…
- Enforcement report date
- Sept. 27, 2023
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing (FDA has not recorded a termination for this recall; FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
- Filed by
- Karl Storz Endoscopy
- Product
- Five S 5 3x65, REF: 0915612-06, sterile for single use, Sterile EO, Rx Only
- Reason
- Flexible intubation endoscopes sterility assurance can not be confirmed.
- Distribution pattern
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MEX, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of Canada.
- Quantity
- 218 units
- Codes and lots
- UDI: 04048551446690/ All lots with remaining shelf life (Globally) Lots imported into the U.S.: 500472 & 500461
- Recall started
- July 7, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Bronchoscope (Flexible Or Rigid)
- Root cause (FDA)
- Process change control
- What the company told customers
- On August, 4, 2023, Karl Storz Endoscopy issued an Urgent Medical Device Recall Notification via: UPS. Karl Storz asked consignees to take the following actions: 1. Immediately quarantine and discontinue use of Material numbers listed above. 2. Pass on this Urgent Medical Device Recall notice to all users of the products listed above and all other persons who need to be aware within your organization. 3. If you have distributed the products listed, please promptly forward this letter to those recipients, and indicate contact details of the recipient on the Customer Reply From. 4. Return the filled Customer Reply Form by Fax or E-Mail to the indicated contact. 5. Get in touch with your KARL STORZ representative to return affected products. 6. Please report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. 7. (Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.)
- Recall number
- Z-2589-2023
What the record says about this device
A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.
“Flexible intubation endoscopes sterility assurance can not be confirmed.”Five S 5 3x65, REF: 0915612-06, sterile for single use, Sterile EO, Rx Only, FDA enforcement report, 2023-09-27. The caption is the record's own words.
A generic device, not the product. The mark follows the record's own words.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: Bronchoscope (Flexible Or Rigid)
Distributed in (FDA's record): Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware District of Columbia Florida Georgia Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Nebraska Nevada New Hampshire New Jersey New Mexico New York North Carolina Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina Tennessee Texas Utah Vermont Virginia Washington West Virginia Wisconsin Wyoming and nationwide
This recall among the others
- FDA's enforcement report came 82 days after the firm started the recall. For half of the 835 FDA medical device recall events reported in 2023 with a start date, it took 51 days or fewer (by year).
- Karl Storz Endoscopy: 8 other recalls or notices on this site in the three years before this one (from Sept. 27, 2020).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement report, recall number Z-2589-2023. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-2589-2023, Sept. 27, 2023. "Five S 5 3x65." PlainRecords, https://plainrecords.com/recall/device/z-2589-2023/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections