FDA · Medical device recall
Centricity Universal Viewer Zero Footprint Client and 4 more products
- Date
- June 15, 2022
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- GE Healthcare, LLC
- Reason
- Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution.
- Distribution pattern
- Worldwide distribution. Nationwide distribution to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International distribution to Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Bolivia, Plurinational State of, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, South, Kuwait, Lebanon, Malaysia, Malta, Mexico, Namibia, Netherlands, Nigeria, Norway, Pakistan, Panama, Peru, Poland, Portugal, Puerto Rico, Qatar, Russia, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Bolivarian Republic of, Vietnam
5 products in this recall
Centricity Universal Viewer Zero Footprint Client
- Product
- Centricity Universal Viewer Zero Footprint Client
- Quantity
- 488 units
- Codes and lots
- Software Version 6.0 SP11 through 6.0 SP11.4; GTIN/DI: 00840682102988; Codes: 2089507-1XX
- Recall started
- April 28, 2022
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design
- What the company told customers
- GE Healthcare notified consignees on about 04/28/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION." The letter outlined two safety issues and instructed customers to continue to use the device in accordance with the User Manuals, but for issue #1, do not rely on measurements displayed in the viewer. They should manually calibrate the image to create a measurement calibration reference and then perform necessary measurements. For issue #2, do not perform measurements on lossy images. In both instances, customers were requested to complete and return the acknowledgment form to Recall.85460@ge.com. A second letter was sent to customers on about 02/24/2023, to provide instructions for downloading the user manual addendum for CWeb.
- Recall number
- Z-1229-2022
Centricity Enterprise Web
- Product
- Centricity Enterprise Web
- Quantity
- 536 units
- Codes and lots
- Software Versions 3.0 through 3.0 SP14d, 4.0 through 4.0 Spa6c and 4.0 Spa7b; Codes: 2027870-0XX, 2049944-0XX, 2068692-0XX.
- Recall started
- April 28, 2022
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design
- What the company told customers
- GE Healthcare notified consignees on about 04/28/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION." The letter outlined two safety issues and instructed customers to continue to use the device in accordance with the User Manuals, but for issue #1, do not rely on measurements displayed in the viewer. They should manually calibrate the image to create a measurement calibration reference and then perform necessary measurements. For issue #2, do not perform measurements on lossy images. In both instances, customers were requested to complete and return the acknowledgment form to Recall.85460@ge.com. A second letter was sent to customers on about 02/24/2023, to provide instructions for downloading the user manual addendum for CWeb.
- Recall number
- Z-1230-2022
Centricity Radiology RA600
- Product
- Centricity Radiology RA600
- Quantity
- 1,332 units
- Codes and lots
- a. Software Version 7.0 through 7.0 SP, No GTIN/DI (Code: 2022296-0XX); b. Software Version 8.0 through 8.0 SP14H, GTIN/DI: 00840682125260 (Codes: 2038437-0XX, 2097190-0XX)
- Recall started
- April 28, 2022
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design
- What the company told customers
- GE Healthcare notified consignees on about 04/28/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION." The letter outlined two safety issues and instructed customers to continue to use the device in accordance with the User Manuals, but for issue #1, do not rely on measurements displayed in the viewer. They should manually calibrate the image to create a measurement calibration reference and then perform necessary measurements. For issue #2, do not perform measurements on lossy images. In both instances, customers were requested to complete and return the acknowledgment form to Recall.85460@ge.com. A second letter was sent to customers on about 02/24/2023, to provide instructions for downloading the user manual addendum for CWeb.
- Recall number
- Z-1231-2022
Centricity Cardiology CA1000
- Product
- Centricity Cardiology CA1000
- Quantity
- 601 units
- Codes and lots
- a. Software Version 1.0 through 1.0 SP, No GTIN/DI (Code: 2022296-0XX); b. Software Version 2.0 through 2.0 SP14H, GTIN/DI: 00840682125260 (Codes: 2038437-0XX, 2097190-0XX)
- Recall started
- April 28, 2022
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design
- What the company told customers
- GE Healthcare notified consignees on about 04/28/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION." The letter outlined two safety issues and instructed customers to continue to use the device in accordance with the User Manuals, but for issue #1, do not rely on measurements displayed in the viewer. They should manually calibrate the image to create a measurement calibration reference and then perform necessary measurements. For issue #2, do not perform measurements on lossy images. In both instances, customers were requested to complete and return the acknowledgment form to Recall.85460@ge.com. A second letter was sent to customers on about 02/24/2023, to provide instructions for downloading the user manual addendum for CWeb.
- Recall number
- Z-1232-2022
Centricity PACS RA1000
- Product
- Centricity PACS RA1000
- Quantity
- 676 units
- Codes and lots
- a. Software Version 3.0 through 3.2 SP8, No GTIN/DI (Code: 2051372-0XX, 2064678-0XX) b. Software Version 4.0 through 4.0 SP14, GTIN/DI 00840682124447 (Code: 2062264-0XX); c. 6.0 through 6.0 SP10.3, GTIN/DI: 00840682104821 (Code: 2094628-0XX); d. 7.0 through 7.0 SP0.0.4.7, GTIN/DI: 00840682145558 (Code: 5824366-0XX)
- Recall started
- April 28, 2022
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design
- What the company told customers
- GE Healthcare notified consignees on about 04/28/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION." The letter outlined two safety issues and instructed customers to continue to use the device in accordance with the User Manuals, but for issue #1, do not rely on measurements displayed in the viewer. They should manually calibrate the image to create a measurement calibration reference and then perform necessary measurements. For issue #2, do not perform measurements on lossy images. In both instances, customers were requested to complete and return the acknowledgment form to Recall.85460@ge.com. A second letter was sent to customers on about 02/24/2023, to provide instructions for downloading the user manual addendum for CWeb.
- Recall number
- Z-1233-2022
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-1229-2022, Z-1230-2022, Z-1231-2022, Z-1232-2022, Z-1233-2022. The agency record governs; check it and the company before acting.