FDA · Medical device recall

epoc BGEM BUN Test Card with epoc Host SW v3.37.3… and 2 more products

Recall started · FDA enforcement report Sept. 6, 2023 Class II

Reason
There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Status
Ongoing
Distribution
US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Sept. 6, 2023
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA has not recorded a termination for this recall; FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Siemens Healthcare Diagnostics Inc
Reason
There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Distribution pattern
US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device or a product, the reason the record gives, the record's wordsA simplified drawing for a device record; the mark follows the reason the record gives, not a photograph of the product. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: epoc BGEM BUN Test Card with epoc Host…2 The record describes a test result that moves out of its range3 What the record says4 What the record listsFOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepdiscrepant

“There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.”epoc BGEM BUN Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1, FDA enforcement report, 2023-09-06. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Distributed in (FDA's record): Alabama Arizona Arkansas California Colorado Delaware District of Columbia Florida Idaho Illinois Iowa Kansas Kentucky Louisiana Michigan Minnesota Missouri New Hampshire New York North Carolina North Dakota Ohio Oklahoma South Carolina South Dakota Texas

3 products in this recall

epoc BGEM BUN Test Card with epoc Host SW v3.37.3…

Product
epoc BGEM BUN Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1
Quantity
3201 ea
Codes and lots
UDI-DI: 00809708121860 Siemens Material Number: 10736515 Lots: 04-23003-60 (expiry 20-Jun-2023), 11-22337-60 (expiry 20-May-2023), 04-23034-40 (expiry 21-Jul-2023), 04-23059-50 (expiry 15-Aug-2023), 04-23059-60 (expiry 15-Aug-2023), and 04-23069-40 (expiry 25-Aug-2023)
Recall started
June 1, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Root cause (FDA)
Under Investigation by firm
What the company told customers
Consignees were notified via an Urgent Medical Device Correction letter beginning June 5, 2023. Consignees were informed of the issue and were provided a workaround and actions to take. The next sensor configuration/software version, sensor configuration 42.1/epoc Host SW v3.38.2 /epoc NXS SW v4.11.11, is available to customers in Siemens Healthineers Document Library or the epoc Live Update Service (eLUS). This next sensor configuration, 42.1, mitigates the issue.
Recall number
Z-2476-2023

epoc BGEM Crea Test Card with epoc Host SW v3.37.3…

Product
epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1
Quantity
473 ea
Codes and lots
UDI-DI: 00809708072254 Siemens Material Number: 10736382 Lots: 03-23002-60 (expiry 19-Jun-2023) and 03-23056-50 (expiry 12-Aug-2023)
Recall started
June 1, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Root cause (FDA)
Under Investigation by firm
What the company told customers
Consignees were notified via an Urgent Medical Device Correction letter beginning June 5, 2023. Consignees were informed of the issue and were provided a workaround and actions to take. The next sensor configuration/software version, sensor configuration 42.1/epoc Host SW v3.38.2 /epoc NXS SW v4.11.11, is available to customers in Siemens Healthineers Document Library or the epoc Live Update Service (eLUS). This next sensor configuration, 42.1, mitigates the issue.
Recall number
Z-2477-2023

epoc BGEM Test Card (BUN & TCO2),Veterinary with epoc Host SW v3.37.3…

Product
epoc BGEM Test Card (BUN & TCO2),Veterinary with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1
Quantity
348 ea
Codes and lots
UDI-DI: 00630414243788 Siemens Material Number: 10736516 Lots: 04-23043-51 (expiry 30-Jul-2023)
Recall started
June 1, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Root cause (FDA)
Under Investigation by firm
What the company told customers
Consignees were notified via an Urgent Medical Device Correction letter beginning June 5, 2023. Consignees were informed of the issue and were provided a workaround and actions to take. The next sensor configuration/software version, sensor configuration 42.1/epoc Host SW v3.38.2 /epoc NXS SW v4.11.11, is available to customers in Siemens Healthineers Document Library or the epoc Live Update Service (eLUS). This next sensor configuration, 42.1, mitigates the issue.
Recall number
Z-2478-2023

This recall among the others

  • FDA's enforcement report came 97 days after the firm started the recall. For half of the 835 FDA medical device recall events reported in 2023 with a start date, it took 51 days or fewer (by year).
  • Siemens Healthcare Diagnostics: 50 other recalls or notices on this site in the three years before this one (from Sept. 6, 2020).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement reports, recall numbers Z-2476-2023, Z-2477-2023, Z-2478-2023. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-2476-2023, Sept. 6, 2023. "epoc BGEM BUN Test Card with epoc Host SW v3.37.3… and 2 more products." PlainRecords, https://plainrecords.com/recall/device/z-2476-2023/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance