FDA · Medical device recall

StealthStation Cranial Software…

Recall started · FDA enforcement report June 14, 2023 Class I

Reason
During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result…
Status
Ongoing
Distribution
Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR, NJ, MN, MO, WI, MA, ND, NY, MD, OR, AK, NE, TN, AZ, IN, GA, GU, VA, PA, TX, SD, MT, NV, MS, KY, AL, CT, WV, KS, IA, NC, UT, CO, NH, DC, WY. OUS (foreign): ALBANIA, ARGENTINA…

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
June 14, 2023
Risk class
Class I (FDA's definition, 21 CFR 7.3(m): a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death)
Status
Ongoing (FDA has not recorded a termination for this recall; FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Medtronic Navigation, Inc.
Product
StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems
Reason
During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury
Distribution pattern
Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR, NJ, MN, MO, WI, MA, ND, NY, MD, OR, AK, NE, TN, AZ, IN, GA, GU, VA, PA, TX, SD, MT, NV, MS, KY, AL, CT, WV, KS, IA, NC, UT, CO, NH, DC, WY. OUS (foreign): ALBANIA, ARGENTINA, Australia, AUSTRIA, BANGLADESH, BARBADOS, BELGIUM, Bosnia and Herzegovina, BULGARIA, CANADA, CHILE, CROATIA, CZECH REPUBLIC, DENMARK, FRANCE, GREECE, INDIA, IRELAND, ITALY, JAPAN, LUXEMBOURG, MARTINIQUE, NEW ZEALAND, POLAND, PORTUGAL, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SINGAPORE, SLOVAKIA, SPAIN, SWITZERLAND, TAIWAN, Turkey, UNITED ARAB EMIRATES, UNITED KINGDOM
Quantity
746
Codes and lots
Model/UDI-DI; Software Version: 9735585/00763000631635, 9735586/00763000631765, 9735587/00763000631826; 3.1.4
Recall started
April 11, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Neurological Stereotaxic Instrument
Root cause (FDA)
Software design
What the company told customers
On 4/11/23, correction notices were sent to healthcare professionals that were advised of the following; 1) Do not utilize the following: a) Biopsy Depth Gauge Graphic in Cranial Biopsy procedures. In accordance with the IFU: Frequently confirm navigation accuracy and system responsiveness during live navigation. Use the navigation instrument to touch several bony anatomical landmarks and confirm that the locations identified on the images match the locations touched on the patient. If accuracy degrades, re-register the patient. Always use the Biopsy Needle Mechanical Depth Stop. If system navigation seems inaccurate and steps to restore accuracy are unsuccessful, abort use of the system. b) Distance to/past Target value for Tumor Resection, Shunt Placement, and DBS procedures. Per the IFU: "Warning: Frequently confirm navigation accuracy and system responsiveness during live navigation. Use the navigation instrument to touch several bony anatomical landmarks and confirm that the locations identified on the images match the locations touched on the patient. If accuracy degrades, re-register the patient." and "If system navigation seems inaccurate and steps to restore accuracy are unsuccessful, abort use of the system." 2) All physician users should review the correction notice. Pass on the notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. 3) Complete and return the customer confirmation form that can be returned to neuro.quality@medtronic.com. Firm will provide warning and instructional placards to be applied to the affected systems. Firm is currently working on a software update and will communicate additional information when it becomes available. If you have any questions related to this issue, contact Technical Services for help at: 1-888-826-5603 or email at rs.navtechsupport@medtronic.com
Recall number
Z-1630-2023

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device, the reason the record gives, the record's wordsA simplified drawing of a medical device with a mark for the reason the record gives. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: StealthStation Cranial Software…2 The record describes a value it shows or uses that may be wrong3 What the record says4 What the record lists??FOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepsoftware anomalies

“During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury”StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems, FDA enforcement report, 2023-06-14. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Neurological Stereotaxic Instrument

Distributed in (FDA's record): Alabama Alaska Arizona Arkansas California Colorado Connecticut District of Columbia Florida Georgia Illinois Indiana Iowa Kansas Kentucky Louisiana Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska Nevada New Hampshire New Jersey New York North Carolina North Dakota Ohio Oklahoma Oregon Pennsylvania South Dakota Tennessee Texas Utah Virginia Washington West Virginia Wisconsin Wyoming and nationwide

This recall among the others

  • FDA's enforcement report came 64 days after the firm started the recall. For half of the 835 FDA medical device recall events reported in 2023 with a start date, it took 51 days or fewer (by year).
  • Medtronic Navigation: 7 other recalls or notices on this site in the three years before this one (from June 14, 2020).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-1630-2023. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-1630-2023, June 14, 2023. "StealthStation Cranial Software." PlainRecords, https://plainrecords.com/recall/device/z-1630-2023/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance