Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit… and 7 more products
- Reason
- There is a potential that the pressure tubing may detach from the blood sampling system
- Status
- Terminated
- Distribution
- US: AZ CA CO CT DC DE FL GA ID IL IN KY LA MA MD ME MI MN MO MS MT NC ND NE NJ PR SD TX UT VA WA WI WV OUS: Canada, United Kingdom, Italy
- Enforcement report date
- July 7, 2021
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated (FDA's definition: FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard; as in openFDA's data of Nov. 10, 2026)
- Filed by
- Edwards Lifesciences, LLC
- Reason
- There is a potential that the pressure tubing may detach from the blood sampling system
- Distribution pattern
- US: AZ CA CO CT DC DE FL GA ID IL IN KY LA MA MD ME MI MN MO MS MT NC ND NE NJ PR SD TX UT VA WA WI WV OUS: Canada, United Kingdom, Italy
What the record says about this device
A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.
“There is a potential that the pressure tubing may detach from the blood sampling system”Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507, FDA enforcement report, 2021-07-07. The caption is the record's own words.
A generic device, not the product. The mark follows the record's own words.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: Transducer, Pressure, Catheter Tip
Distributed in (FDA's record): Arizona California Colorado Connecticut Delaware District of Columbia Florida Georgia Idaho Illinois Indiana Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska New Jersey North Carolina North Dakota Puerto Rico South Dakota Texas Utah Virginia Washington West Virginia Wisconsin
8 products in this recall
Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit…
- Product
- Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
- Quantity
- 460 units
- Codes and lots
- Lot #s: 63505216 and 63590087
- Recall started
- May 26, 2021
- Recall terminated (FDA)
- Dec. 14, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Transducer, Pressure, Catheter Tip
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
- Recall number
- Z-1959-2021
Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit…
- Product
- Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
- Quantity
- 700 units
- Codes and lots
- Lot #s: 63470654, 63505217 and 63592430
- Recall started
- May 26, 2021
- Recall terminated (FDA)
- Dec. 14, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Transducer, Pressure, Catheter Tip
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
- Recall number
- Z-1960-2021
Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit…
- Product
- Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***Not Marketed within the US
- Quantity
- 290 units
- Codes and lots
- Lot # 63479666- Not Marketed within the US
- Recall started
- May 26, 2021
- Recall terminated (FDA)
- Dec. 14, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Transducer, Pressure, Catheter Tip
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
- Recall number
- Z-1961-2021
Edwards REF: VJ0990…
- Product
- Edwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
- Quantity
- 2,000 units
- Codes and lots
- Lot # 63456332
- Recall started
- May 26, 2021
- Recall terminated (FDA)
- Dec. 14, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Transducer, Pressure, Catheter Tip
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
- Recall number
- Z-1962-2021
Edwards VAMP Jr…
- Product
- Edwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
- Quantity
- 1,679 units
- Codes and lots
- Lot # 63445142 and 63630816
- Recall started
- May 26, 2021
- Recall terminated (FDA)
- Dec. 14, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Transducer, Pressure, Catheter Tip
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
- Recall number
- Z-1963-2021
Edwards TruWave (30 cc)/VAMP Jr Kit…
- Product
- Edwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
- Quantity
- 990 units
- Codes and lots
- Lot # 63438489
- Recall started
- May 26, 2021
- Recall terminated (FDA)
- Dec. 14, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Transducer, Pressure, Catheter Tip
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
- Recall number
- Z-1964-2021
Edwards TruWave 3 cc/VAMP Jr Kit…
- Product
- Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
- Quantity
- 50 units
- Codes and lots
- Lot # 63598812
- Recall started
- May 26, 2021
- Recall terminated (FDA)
- Dec. 14, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Transducer, Pressure, Catheter Tip
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
- Recall number
- Z-1965-2021
Edwards TruWave (3 cc)/VAMP Jr…
- Product
- Edwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
- Quantity
- 90 units
- Codes and lots
- Lot # 63545959
- Recall started
- May 26, 2021
- Recall terminated (FDA)
- Dec. 14, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Transducer, Pressure, Catheter Tip
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
- Recall number
- Z-1966-2021
This recall among the others
- FDA's enforcement report came 42 days after the firm started the recall. For half of the 892 FDA medical device recall events reported in 2021 with a start date, it took 50 days or fewer (by year).
- Edwards Lifesciences: 12 other recalls or notices on this site in the three years before this one (from July 8, 2018).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement reports, recall numbers Z-1959-2021, Z-1960-2021, Z-1961-2021, Z-1962-2021, Z-1963-2021, Z-1964-2021, Z-1965-2021, Z-1966-2021. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-1959-2021, July 7, 2021. "Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit… and 7 more products." PlainRecords, https://plainrecords.com/recall/device/z-1959-2021/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections