FDA · Medical device recall

Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit… and 7 more products

Recall started · FDA enforcement report July 7, 2021 Class II

Reason
There is a potential that the pressure tubing may detach from the blood sampling system
Status
Terminated
Distribution
US: AZ CA CO CT DC DE FL GA ID IL IN KY LA MA MD ME MI MN MO MS MT NC ND NE NJ PR SD TX UT VA WA WI WV OUS: Canada, United Kingdom, Italy

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
July 7, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated (FDA's definition: FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard; as in openFDA's data of Nov. 10, 2026)
Filed by
Edwards Lifesciences, LLC
Reason
There is a potential that the pressure tubing may detach from the blood sampling system
Distribution pattern
US: AZ CA CO CT DC DE FL GA ID IL IN KY LA MA MD ME MI MN MO MS MT NC ND NE NJ PR SD TX UT VA WA WI WV OUS: Canada, United Kingdom, Italy

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device, the reason the record gives, the record's wordsA simplified drawing of a medical device with a mark for the reason the record gives. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: Edwards TruWave (3 cc)/VAMP Jr 48…2 The record describes a part that can loosen or separate3 What the record says4 What the record listsFOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepdetach

“There is a potential that the pressure tubing may detach from the blood sampling system”Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507, FDA enforcement report, 2021-07-07. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Transducer, Pressure, Catheter Tip

Distributed in (FDA's record): Arizona California Colorado Connecticut Delaware District of Columbia Florida Georgia Idaho Illinois Indiana Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska New Jersey North Carolina North Dakota Puerto Rico South Dakota Texas Utah Virginia Washington West Virginia Wisconsin

8 products in this recall

Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit…

Product
Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
Quantity
460 units
Codes and lots
Lot #s: 63505216 and 63590087
Recall started
May 26, 2021
Recall terminated (FDA)
Dec. 14, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Transducer, Pressure, Catheter Tip
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
Recall number
Z-1959-2021

Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit…

Product
Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
Quantity
700 units
Codes and lots
Lot #s: 63470654, 63505217 and 63592430
Recall started
May 26, 2021
Recall terminated (FDA)
Dec. 14, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Transducer, Pressure, Catheter Tip
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
Recall number
Z-1960-2021

Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit…

Product
Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***Not Marketed within the US
Quantity
290 units
Codes and lots
Lot # 63479666- Not Marketed within the US
Recall started
May 26, 2021
Recall terminated (FDA)
Dec. 14, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Transducer, Pressure, Catheter Tip
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
Recall number
Z-1961-2021

Edwards REF: VJ0990…

Product
Edwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
Quantity
2,000 units
Codes and lots
Lot # 63456332
Recall started
May 26, 2021
Recall terminated (FDA)
Dec. 14, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Transducer, Pressure, Catheter Tip
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
Recall number
Z-1962-2021

Edwards VAMP Jr…

Product
Edwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
Quantity
1,679 units
Codes and lots
Lot # 63445142 and 63630816
Recall started
May 26, 2021
Recall terminated (FDA)
Dec. 14, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Transducer, Pressure, Catheter Tip
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
Recall number
Z-1963-2021

Edwards TruWave (30 cc)/VAMP Jr Kit…

Product
Edwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
Quantity
990 units
Codes and lots
Lot # 63438489
Recall started
May 26, 2021
Recall terminated (FDA)
Dec. 14, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Transducer, Pressure, Catheter Tip
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
Recall number
Z-1964-2021

Edwards TruWave 3 cc/VAMP Jr Kit…

Product
Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
Quantity
50 units
Codes and lots
Lot # 63598812
Recall started
May 26, 2021
Recall terminated (FDA)
Dec. 14, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Transducer, Pressure, Catheter Tip
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
Recall number
Z-1965-2021

Edwards TruWave (3 cc)/VAMP Jr…

Product
Edwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
Quantity
90 units
Codes and lots
Lot # 63545959
Recall started
May 26, 2021
Recall terminated (FDA)
Dec. 14, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Transducer, Pressure, Catheter Tip
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 26, 2021 Edwards Lifesciences issued a "Urgent Medical Device Recall" Notification via FedEx to all affected consignees. In addition to informing consignees about the recalled product, the firm asked consignees to take the following actions: 1. Action to be taken by user: We request that you return all unused product in your inventory with the model and lot numbered reference above. For product that has been used and will be returned, please contact Customer Service at 1-800-424-3278 to request a biokit. Please review your inventory and quarantine any affected product until prepared for return to Edwards Lifesciences. Please follow the instructions included in the enclosed acknowledgement form an return within 5 days of receipt of this notification. 2. Action to be taken by distributor: Please complete the acknowledgement form and return to US.FCA@edwards.com. Please forward this customer communication to any of your customers who have purchased the impacted Edwards VAMP Jr. and TruWave/VAMP Jr. Products and ask them to return the product toe Edwards with the enclosed acknowledgement form. 3. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product and Edwards will communicate directly with your customers to facilitate the recall and acknowledgement process. Please forward your customer list to US.FCA@edwards.com 4. Adverse Event Reporting in the US: - Please contact the FDA's MedWatch Adverse Event reporting program either online, by regular mail, or by fa: Complete and submit the report Online: WWW.fda.gov/medwatch/report.htm - Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178 5. Transmission of the Product Recall: This notice need to be passed on to all individuals within your organization who need to be aware of th
Recall number
Z-1966-2021

This recall among the others

  • FDA's enforcement report came 42 days after the firm started the recall. For half of the 892 FDA medical device recall events reported in 2021 with a start date, it took 50 days or fewer (by year).
  • Edwards Lifesciences: 12 other recalls or notices on this site in the three years before this one (from July 8, 2018).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement reports, recall numbers Z-1959-2021, Z-1960-2021, Z-1961-2021, Z-1962-2021, Z-1963-2021, Z-1964-2021, Z-1965-2021, Z-1966-2021. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-1959-2021, July 7, 2021. "Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit… and 7 more products." PlainRecords, https://plainrecords.com/recall/device/z-1959-2021/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance