U.S. safety records, as the agencies wrote them. Independent.
FDA · Medical device recall

SnapshotGLO Wound Imaging Device…

Recall started · FDA enforcement report July 29, 2026 Class II

Reason
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when…
Status
Ongoing
Distribution
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
July 29, 2026
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
KENT IMAGING, INC.
Product
SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
Reason
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
Distribution pattern
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
Quantity
77
Codes and lots
UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3
Recall started
June 30, 2026
Voluntary or mandated
Voluntary: Firm initiated
Device type
Autofluorescence Imaging Adjunct Tool For Wounds
Root cause (FDA)
Software design
What the company told customers
On 6/30/2026, correction and removal notices were emailed to customers who were asked to do the following: 1) Discontinue use of the affected unit(s) and remove them from clinical service. 2) Contact Kent Imaging at customerservice@kentimaging.com to confirm return shipping instructions. 3) Pending return, keep the device charged and avoid allowing the battery to fully deplete, to prevent the unit from becoming temporarily inoperable in the interim. 4) Within 5 business days of receipt of this notice, please confirm receipt and current status of the listed unit(s) by replying to this letter or contacting Hitalo Arume directly. Distributor was asked to: a) Identify the current location/holder of each of these 6 units; b) Coordinate scheduling of a brief remote session between Kent Imaging's technical support team and each unit's user; and c) Confirm completion of each update on the attached Device Reconciliation Form. d) For units that require physical return, quarantine and stop distribution. For units distributed, identify each affected customer, coordinate on-site access for collection, and arrange return shipment to Kent Imaging. Questions can be directed to hitaloa@kentimaging.com
Recall number
Z-2775-2026

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device, the reason the record gives, the record's wordsA simplified drawing of a medical device with a mark for the reason the record gives. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: SnapshotGLO Wound Imaging Device, Model…2 The record describes a software issue3 What the record says4 What the record lists!!FOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepsoftware

“The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.”SnapshotGLO Wound Imaging Device, Model KB100, KB100AA., FDA enforcement report, 2026-07-29. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Autofluorescence Imaging Adjunct Tool For Wounds

Distributed in (FDA's record): Arizona Arkansas California District of Columbia Florida Georgia Illinois Indiana Iowa Kansas Nevada Pennsylvania Tennessee Texas Utah and nationwide

This recall among the others

  • FDA's enforcement report came 29 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).
  • KENT IMAGING: 1 other recall or notice on this site in the three years before this one (from July 30, 2023).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-2775-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-2775-2026, July 29, 2026. "SnapshotGLO Wound Imaging Device." PlainRecords, https://plainrecords.com/recall/device/z-2775-2026/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance