SnapshotGLO Wound Imaging Device…
- Reason
- The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when…
- Status
- Ongoing
- Distribution
- US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
- Enforcement report date
- July 29, 2026
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
- Filed by
- KENT IMAGING, INC.
- Product
- SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
- Reason
- The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
- Distribution pattern
- US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
- Quantity
- 77
- Codes and lots
- UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3
- Recall started
- June 30, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Autofluorescence Imaging Adjunct Tool For Wounds
- Root cause (FDA)
- Software design
- What the company told customers
- On 6/30/2026, correction and removal notices were emailed to customers who were asked to do the following: 1) Discontinue use of the affected unit(s) and remove them from clinical service. 2) Contact Kent Imaging at customerservice@kentimaging.com to confirm return shipping instructions. 3) Pending return, keep the device charged and avoid allowing the battery to fully deplete, to prevent the unit from becoming temporarily inoperable in the interim. 4) Within 5 business days of receipt of this notice, please confirm receipt and current status of the listed unit(s) by replying to this letter or contacting Hitalo Arume directly. Distributor was asked to: a) Identify the current location/holder of each of these 6 units; b) Coordinate scheduling of a brief remote session between Kent Imaging's technical support team and each unit's user; and c) Confirm completion of each update on the attached Device Reconciliation Form. d) For units that require physical return, quarantine and stop distribution. For units distributed, identify each affected customer, coordinate on-site access for collection, and arrange return shipment to Kent Imaging. Questions can be directed to hitaloa@kentimaging.com
- Recall number
- Z-2775-2026
What the record says about this device
A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.
“The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.”SnapshotGLO Wound Imaging Device, Model KB100, KB100AA., FDA enforcement report, 2026-07-29. The caption is the record's own words.
A generic device, not the product. The mark follows the record's own words.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: Autofluorescence Imaging Adjunct Tool For Wounds
Distributed in (FDA's record): Arizona Arkansas California District of Columbia Florida Georgia Illinois Indiana Iowa Kansas Nevada Pennsylvania Tennessee Texas Utah and nationwide
This recall among the others
- FDA's enforcement report came 29 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).
- KENT IMAGING: 1 other recall or notice on this site in the three years before this one (from July 30, 2023).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement report, recall number Z-2775-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-2775-2026, July 29, 2026. "SnapshotGLO Wound Imaging Device." PlainRecords, https://plainrecords.com/recall/device/z-2775-2026/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections