FDA · Medical device recall
VITROS Immunodiagnostic Products High Sample Diluent A Reagent Pack
- Date
- Jan. 26, 2022
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Immunodiagnostic Products High Sample Diluent A Reagent Pack
- Reason
- Lower than expected results for VITROS¿ Immunodiagnostic Products TSH Reagent Pack when diluted with VITROS¿ Immunodiagnostic Products High Sample Diluent A Reagent Pack Lot 2190.
- Distribution pattern
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, IL, IN, KY, LA, MD, ME, MI, MO, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WI, WV and the countries of Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.
- Quantity
- 1416 units
- Codes and lots
- Lot # 2190; Product code(unique Device Identifier): 8430373 (10758750004898)
- Recall started
- Nov. 22, 2021
- Device type
- General Purpose Reagent
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Ortho clinical Diagnostics issued an Important product correction notification on November 22, 2021. Actions to be taken by the Distributors: 1. Please provide the communication (Ref. CL2021-287) to customers who were shipped VITROS HSDA Lot 2190 from your facility. 2. Do not continue to distribute VITROS HSDA Lot 2190. 3. Complete the Confirmation of Receipt form no later than November 30, 2021. 4. Please forward this notification if the product was distributed outside of your facility. If you have questions, please contact our Ortho Care" Technical Solutions Center at 1-800-421-3311. Action to be taken by the customers: 1. Discontinue using VITROS HSDA Lot 2190 if your laboratory utilizes VITROS TSH Reagent Pack. 2.Complete the Confirmation of Receipt form no later than November 30, 2021, even if no replacement product is required. If replacement is required, indicate the quantity of VITROS HSDA Lot 2190 to be replaced. 3. Please forward this notification if the product was distributed outside of your facility. Contact Information If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311
- Recall number
- Z-0503-2022