FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software
- Reason
- The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel
- Status
- Ongoing
- Distribution
- US:DE, AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: AD, AE, AI, AR, AT, AU, BD, BE, BF, BG…
- Enforcement report date
- Sept. 20, 2023
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing (FDA has not recorded a termination for this recall; FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
- Filed by
- BioFire Diagnostics, LLC
- Product
- FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software
- Reason
- The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel
- Distribution pattern
- US:DE, AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: AD, AE, AI, AR, AT, AU, BD, BE, BF, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EE, ES, FI, FR, GA, GB, GE, GH, GM, GR, GU, HK, HU, IC, ID, IE, IL, IN, IQ, IT, JP, KE, KR, KW, LV, MO, MX, MY, NL, NO, NP, NZ, PG, PH, PL, PT, QA, RS, SA, SE, SG, SI, TH, TN, TR, TW, VN, ZA, ZW
- Quantity
- 61,069
- Codes and lots
- Model/UDI-DI: RFIT-ASY-0104/00815381020116, RFIT-ASY-0116/00815381020109 using software versions prior to FLM2-SFW-0040 v2.0.3.1/BFS0001-5539 v2.0.200.10
- Recall started
- April 25, 2023
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System
- Root cause (FDA)
- Software design
- What the company told customers
- On 4/25/23 correction notices were mailed to customers who were told to update the GI pouch module software. Instructions for installation can be downloaded here: https://www.biofiredx.com/e-labeling/ITIFA20GI21. This technical note and download cannot be opened via Firefox browser; please use a different browser. Once the pouch module software has been updated, complete and return the webform located at: https://www.biofiredx.com/gisoftware-5747 or return the acknowledgement form via email to FCA5747@biomerieux.com Questions or concerns can be directed to Technical Support at biofiresupport@biomerieux.com or via telephone by dialing 1.800.735.6544 and selecting option 5.
- Recall number
- Z-2588-2023
What the record says about this device
A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.
“The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel”FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software, FDA enforcement report, 2023-09-20. The caption is the record's own words.
A generic device, not the product. The mark follows the record's own words.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System
Distributed in (FDA's record): Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware District of Columbia Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska Nevada New Hampshire New Jersey New Mexico New York North Carolina North Dakota Ohio Oklahoma Oregon Pennsylvania Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Utah Vermont Virginia Washington West Virginia Wisconsin Wyoming
This recall among the others
- FDA's enforcement report came 148 days after the firm started the recall. For half of the 835 FDA medical device recall events reported in 2023 with a start date, it took 51 days or fewer (by year).
- BioFire Diagnostics: 9 other recalls or notices on this site in the three years before this one (from Sept. 20, 2020).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement report, recall number Z-2588-2023. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-2588-2023, Sept. 20, 2023. "FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software." PlainRecords, https://plainrecords.com/recall/device/z-2588-2023/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections