ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters…

Nov. 15, 2010

Date posted by the FDA
Nov. 15, 2010
Status
Terminated
Filed by
Ethicon Endo-Surgery Inc
Product
ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Model # ATS45. Manfuactured by Ethicon Endo-Survery, LLC; 475 Calle C; Guaynabo PR 00969.
Reason
The mechanism that connects the articulation joint to the device shaft may cause the jaws of the cutters to remain closed and clamped down on tissue after the device is fired. If the jaws of the device remain clamped down on the tissue, there is a risk the surgery will need to be modified to remove the device. This may lead to an increase in procedure time, tissue manipulation, and a change in t
Distribution pattern
Worldwide Distribution: USA only, in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI. IA. ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Quantity
4,113 UNITS
Codes and lots
Lot #'s: E4MD1T, E4MD5W, E4MD3Y, E4ME7J, E4MD4Z, E4ME7U, E4MD50 & E4MF0A
Recall started
Jan. 13, 2009
Device type
Staple, Implantable
Root cause (FDA)
Other
What the company told customers
Ethicon Endo-Surgery issued "Urgent Device Recall Event 1952" notification letters dated January 21, 2009 to their customers. They instructed their custormers to immediately discontinue use of the product. They are to complete and fax back the Business Reply Form. Affected product is to be returned with a copy of the completed Business Reply Form to Stericycle. Replacement product should be received within 30-60 days after return. Customers can contact Stericycle concerning this recall at 1-866-861-6921, reference Event 1952.
Recall number
Z-0361-2011

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0361-2011. The agency record governs; check it and the company before acting.