FDA · Medical device recall

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell… and 15 more products

Recall started · FDA enforcement report May 27, 2020 Class II

Reason
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
Status
Terminated
Distribution
Worldwide distribution including the following states: New York, Arkansas, Oklahoma, Michigan, Delaware, Texas, Louisiana, California, Florida, Arizona, Nevada, Ohio, Colorado, Rhode Island, Iowa, Pennsylvania, Massachusetts, North Carolina, Alabama, New…

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
May 27, 2020
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated (FDA's definition: FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard; as in openFDA's data of Nov. 10, 2026)
Filed by
Howmedica Osteonics Corp.
Reason
The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
Distribution pattern
Worldwide distribution including the following states: New York, Arkansas, Oklahoma, Michigan, Delaware, Texas, Louisiana, California, Florida, Arizona, Nevada, Ohio, Colorado, Rhode Island, Iowa, Pennsylvania, Massachusetts, North Carolina, Alabama, New Hampshire, Illinois, New Jersey, Virginia, Utah, Maryland, Mississippi, Missouri, Tennessee, Connecticut, and Kansas

What this record is about

A generic icon of the kind of thing the record is about; the record's own words are below. It is not a picture of the product.

A generic icon of the kind of thing the record is aboutA simplified icon of the kind of thing the record is about and the phrases of the record's own text beside it.1 The record: Stryker Orthopaedics-Restoration Anatomic…2 In the record's words:3 Only the kind of thing is drawn4 The record's own words are below“The Stryker Restoration Anatomic Shell (RAS) may fall off its postwithin its inner packaging which can result in the device loose in…”STRYKER…

“The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.”Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 54 Left, Sterile, Catalog # 504-02-54C-L, UDI (01)07613327012354(17), FDA enforcement report, 2020-05-27. The caption is the record's own words.

The picture shows the kind of thing the record is about (a part, a product, a package), not the product itself; the ring marks it. Nothing in it is a claim beyond the record's own words below.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Distributed in (FDA's record): Alabama Arizona Arkansas California Colorado Connecticut Delaware Florida Illinois Iowa Kansas Louisiana Maryland Massachusetts Michigan Mississippi Missouri Nevada New Hampshire New Jersey New York North Carolina Ohio Oklahoma Pennsylvania Rhode Island Tennessee Texas Utah Virginia

16 products in this recall

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 54 Left, Sterile, Catalog # 504-02-54C-L, UDI (01)07613327012354(17)
Quantity
4895
Codes and lots
Lot #s JD8WN9, 6514K5, AD78D5, TR0M4X, YT7R3V, 2A7V17, Y81WDV, AN2DME, 514WVN, 9P3WE4, LD3LAL, L21A8P, K02EYK, 6K0J1W, XK44M4, H55J1M, T10T38, VT07WW, 7T1WTE, 6544JE, J45R2R, 8963DT, LV2EW7, PN1X3T, DN0276, RP35M3, T245XV, 497H8P, LJ52DA, HP48J3, H98XN3, 9E8M7P, 1J1RDH, R235KJ, 793724, RA44MH, JJ65MR, N12K3P, PX3TTK, 425H7M, 3V6JAM, P52HME, WX80AA, K65W39, 1R40DL
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2084-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 54 Right, Sterile, Catalog # 504-02-54C-R, UDI (01)07613327012415(17)
Quantity
4895
Codes and lots
Lot #s 6A7N7Y, J50WR6, EM4ED2, 4908KM, K14HL3, VN5NTY, 4X7XE4, JK4A37, W20A86, EM5Y28, H81JNL, EE883L, 9L8LMD, 7Y2HM0, MM0JYY, K23J10, DY42DE, 3P3P64, H77Y5E, MX801E, 1L0758, HM5WKX, D058L6, 4L2R4R, VY0PK4, 872EHV, YD53A4, 4T54D0, 4768E2 WJ0W8W, YV5804, P98YT4, 8N512J, A162RN, J35YL4, H089KL, WW76RP, P58D3J, XH7HN8, RV6DJ8, 8Y0XVE, JM34X2, WJ776A, RH1MRT, N28J76, NP2XXL, L7453E, AA1DYJ, 258Y6J, 6N4X1M, E66DMR, V63W1E, 2739DR, YH1P91, 216NDE, A4410X
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2085-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 56 Left, Sterile, Catalog # 504-02-56D-L, UDI (01)07613327012361(17)
Quantity
4895
Codes and lots
Lot #s 4J3N24, W03J5W, N119A4, WK7K99, NM70TR, 3376L4, RK3XRL, RH27YA, M00KML, TM2MMK, PW30H8, DL01A0, WD4T0V, 2D40YA, 4Y5LAE, W63A8E, VJ34EV, AR5H0T, HY6Y45, KV2835, P84X94, KK0AVP, 2M8RW6, AL0AR3, TP1Y1H, W66X1L, PH28V2, P00606, 3Y3MHV, V32P52, EJ32W6, HN8HXP, 2V46PY, TD1A6J, 7R1X76, W62DKJ, N3435R, W92EP4, HJ0X8X, 5E3X1N, P53NP4, DJ2WKH, P32YRY, X62L47, XV86AL, XV2EY5, 108Y9R, T143LR, VH573D, W56T4K, A071R3, V319PM
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2086-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 58 Left, Sterile, Catalog # 504-02-58D-L, UDI (01)07613327012385(17)
Quantity
4895
Codes and lots
Lot #s AX3V4E, DX109D, TD2X0H, JV415X, KW5467, HM4L3L, WK5WV2, LD2HHY, A67ERK, Y579RW, RW1XVW, VJ7TE0, M02DYJ, EW1TJ5, 371T2M, 2W3T75, 345TJ9, N968R7, H386RY, 978TAX, L6266H, JV4M06, Y7169E, M840TN, 9D6PY0, TV06A0, PR2X50, P85T79, 4M5N2M, 9P3RN6, 247VR6, WE2A2A, 65792M, HA59H2, 4N6HT3, 7N01JA, 762PPX, 2D8P93, 2V4609, NE509M, VJ4RT3, KM720L, 7J3TRH, DP1EDA, V2760N, X05E74, D477X9, R372K2, TR0YTY, P720M0, Y63AD0
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2087-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 56 Right, Sterile, Catalog # 504-02-56D-R, UDI (01)07613327012347(17)
Quantity
4895
Codes and lots
Lot #s EE5WWV, DY6V66, 4T75EE, HT3X47, YX8KLL, 99554T, T58TR0, A46AJA, P23W2H, 9X0WHV, 681E6V, XP0AKD, RD53PW, R67PR4, N11PLW, WW20KE, KV6P9X, 6X4R9T, 2X4KLD, M08065, VN205Y, Y68MTM, EP6P1E, E257HE, PP894J, 4X47VL, 7J0H1A, K50T9R, 2E2139, V92X6Y, T42NM1, LV7LYM, RV8RWH, 2E5WH7, 8H3HK1, JV2PT1, PT243R, M30439, RH817R, XT5R2K, 112LE7, 4J485X, D360KK, PM5R68, KN48NR, LT1YKM, H82NXR, VP8MRR, 4Y44YM, V27D3P, N40RWX, YE2L2E, H500TT
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2088-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 58 Right, Sterile, Catalog # 504-02-58D-R, UDI (01)07613327012446(17)
Quantity
4895
Codes and lots
Lot #s 5L0YLY, DN3EJH, Y62465, 288DP4, LW0D2K, 3K4R62, KN2W5X, M44PVH, 3W01XK, 1Y123H, M55NTP, 683N70, JN4PR7, 497YJX, L022L2, 37889P, X86194, 775A9V, X30XTJ, W333HY, H376E6, 1V4A3X, TP5KK8, VA13MW, 9W0HD6, V72LA3, 3V1WAR, 4H7V1M, 173VRV, 9A4D9X, JE6E3M, 9H2LYX, T532A9, WJ72K7, HX5AN0, XM1A2X, 9T3XPK, TR8WMV, 9A8X5J, 5D78YR, RK0MJJ, PL0LA1, 6P4VX9, A83DPD, 8X715Y, WM2AHV, 1M5P7E, 2E36LW, X81L96
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2089-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 60 Left, Sterile, Catalog # 504-02-60D-L, UDI (01)07613327012422(17)
Quantity
4895
Codes and lots
Lot #s Y461NK, 916M1K, 3581JK, 306XWX, AH4AER, YP48JA, 6J7PTL, KT81WD, 1D6NNE, 467JP5, 6M025E, T4286L, 633D81, 6L6ADX, N175H8, WJ41VD, RM697T, A602M0, 840Y1A, 988KM2, LL2VPW, Y37LK3, M35P5P, 280YM6, R62EM9, WM4W56, JE46JM, AK0560, D11734, EK644L, D802YL, HP25TN, L618D8, N532D2, NW0M02, 972H4A, 95171N, JD4934, WN7K01, V8245A, AK00NH, VL1688, 2E7J6X, 5030EN, P13E4K, K94NEN, MA5W63, 8X783D, 5A3RW3, TV13J7
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2090-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 60 right, Sterile, Catalog # 504-02-60D-R, UDI (01)07613327012408(17)
Quantity
4895
Codes and lots
Lot #s 9Y6675, 2J7WNV, A02X70, ME3841, P94D02, TD37HJ, DT8E38, LM49H5, WE24MT, 74064X, A36853, 694L07, YJ3J46, H26D71, 9X2RDX, AV6M23, L686MM, JY3288, MV2HXY, NP416P, 2L349V, XT6WTY, H02LL7, 601W6E, 6X855W, PV8H2X, R846J8, L705V1, D62TJ0, TM88WY, JE7AW0, 6W0DE9, 3400PT, D62345, 2E6E25, 422V52, 5R8KH0, VE2X77, W56APY, 8267AV, YH5N6L, P32YNV, T5091D, 4M46YJ, 7N4HW9, MM0T9T, XL0WKM
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2091-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 62 left, Sterile, Catalog # 504-02-62E-L, UDI (01)07613327012453(17)
Quantity
4895
Codes and lots
Lot #s EK2JXE, V26EXD, K97J79, MV20HR, RM5RX2, 478E97, ET7EXW, J74RKR, L71A2P, P276D0, M15D1R, ND74JJ, 6P6MWY, X500MV, K931PP, 6J36RW, MH415K, T96X19, VM332V, WW65Y5, X13MJV, L97NMP, RL4L86, MR1RLA, MN1P4X, R91A0P, N76VMX, W361J6, RL4L13, Y36651, PA8JD4, NA049P, 8V5K0A, 5M8HYP, 896EHD, 232R3X, 5D2PDD, T137D5, ML43R3, 5R6RN3, DT11P7, N07WMV, VR591J, 7H32J5, 903NRE, 1X5HLV, AA5Y4M, WL5KR3, PR4Y4L, JL7L46, 9K37HW, NA0N94, YA7HX9, AX89T7, RH44YE, 8E30L6, N411HJ, 3V37MK
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2092-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 62 right, Sterile, Catalog # 504-02-62E-R, UDI (01)07613327012439(17)
Quantity
4895
Codes and lots
Lot #s PH230T, 5316PX, 6487DY, MV4KAW, MY6XW7, NY7JW8, XT3NK1, 8J1031, W43YMN, Y363WD, R58ALD, YM0ATP, MD6J43, M04YA2, 1V4RY5, NY85E3, 9439YT, KR3025, NN41PD, W28H8N, 7T0V3R, MT7MA3, PW531T, JT1H7A, WL30XL, LN8LTY, HJ82TV, XN0HR9, RY7TH7, L97718, KN4A66, JE84VV, 5X3H7R, 404547, 86593Y, 7610XX, WY3E23, JH7RH6, HD1H4E, 59725R, T17YD9, KX6P0W, 4A4TW6, 42775A, E02TD6, 3M44J8, DM2L4J, YX2RMM, 9L6EN0, NJ69D3, 4P7MTN, 2P31RY, 935J0K
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2093-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 64 left, Sterile, Catalog # 504-02-64E-L, UDI (01)07613327012484(17)
Quantity
N/A
Codes and lots
Lot #s Y32LP9, AN6ATN, A77J57, EE69A8, 625VEM, K8549T, EV6MXD, L85MMA, 7E2R48, M21Y21, H71N5N, A74YXL, PN2X3H, V2774X, AE6LJR, 7D0YJE, 6957YA, LM5N9M, 477ELH, JN8347, AY8T8N, EM0W77, TL29PE, NX79TH, 956DXP, A86VNH, KX5MX1, WW7VPM, 7A58DL, 37348T, N80109, 1269DY, 2351A5, H58V9W, 5M1K6H, T3509E, HL627T, WY1XP7, A76DE6, PX677J, AK4L6N, EA2MNR, AM5H8X, KM0M7W, D33PLA, 7R165A
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2094-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 66 left, Sterile, Catalog # 504-02-66F-L, UDI (01)07613327012491(17)
Quantity
4895
Codes and lots
Lot #s J4895V, V9556J, RA3188, DH4RW8, WH083M, KY10DV, 937YAE, XJ1WW1, K52EHL, JX5291, M50LV5, 561RR3, 847AX0, V3377L, EJ40D2, W40V6T, TD3H2L, 9X0DHT, LL7R18, 9N0LH6, 6874DA, Y114EE, PP6X0D, NH50PJ, 6P32NN, MA1VYD, 328L1T, 52028A, X800K7, L51VJR, M22WTW, W80LRW, NV83VN, Y54MNA, V74712, EH1VK4, D9414T, PA6RRP, NT8241, 3P112L, NW7K52
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2095-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 64 right, Sterile, Catalog # 504-02-64E-R, UDI (01)07613327012460(17)
Quantity
4895
Codes and lots
Lot #s HP3142, J93277, HK4HK0, N18T96, 590V0J, WA1D8T, 6D4MT7, 2V3A9W, HE19E5, JN2DWN, 777N5V, YY4LXE, JA0986, 213804, 86112L, 8X4H1D, N25J45, EM8Y25, 647TEX, J011V3, N75WT3, H98AT2, JD1K8M, M576M5, 8Y64X7, E810LX, 7J2099, RH6TNE, HR6M1R, 101MM7, V76WAA, TX6X20, 8N36E0, 3T6330, MR025M, WL6T19, 9K2L3R, K53TKW, RT8P7Y, M82N19, 7E1R1Y, TJ56KK, RX8HHY, NJ4XWJ, WM5VVN
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2096-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 66 right, Sterile, Catalog # 504-02-66F-R, UDI (01)07613327012477(17)
Quantity
4895
Codes and lots
Lot #s 1X5DJK, NH4W8Y, 744EH9, AD25RA, 7E4HW3, 1N21NK, E00T60, WY1DD1, D26R36, H74H7L, RT0AK2, 593Y5R, AD00T8, E12H25, 9K66L6, AL7K8E, MP11MY, D04T4A, 8K74RX, 1R57DY, NH0352, KW5K62, 767RKK, V21PPJ, 6580X6, X73EDY, 5A1ANY, LJ0DWY, RR0T4W, JA761W, N11LR0, 1A03DE, 5E7XRM, TW81AP, 7E69MM, M42Y3R, A28827, 5N1DTH, Y6782W, J33J50, D930PJ, XR6WL0, TH4ED5, J54YNJ, D755P9
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2097-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 68 left, Sterile, Catalog # 504-02-68F-L, UDI (01)07613327012521(17)
Quantity
4895
Codes and lots
Lot #s YX05L9, AL6MM6, VX02E2, 5Y1H86, XY5HP7, MA49HV, A26E76, 4V4APT, LD87TH, MH1RLL, 8R05HH, A67K6H, Y00YXH, PA02HR, LE0L4W, 494M8R, 4T4RAV, PW2E41, 3T59A5, T384A0, PD2Y0T, V44JY5, 650WXN, 3D3H6W, DT2YN6, M550AN, 3N0PN7, YA60V9, DV13RY, 4M1NHR, 3M48X6, LR49YE, NT897W, DX8JV1, VN44ME, VR1H7D, TW4EJ0, T722PN, 6V04AX, 390D9L, KW3W1W, 4D6DWK, W079A2, Y81W0J
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2098-2020

Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…

Product
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 68 right, Sterile, Catalog # 504-02-68F-R, UDI (01)07613327012514(17)
Quantity
4895
Codes and lots
Lot #s DD6E5N, 3314LE, EA67YJ, TA05PD, MJ71XA, VH242D, HR43JY, 5E8XPW, KR4YH2, XW7T57, W92MH8, DY1N6E, 4D8EEY, 6M7D2N, 1T81T1, V45AA1, YL2DP4, H017J0, 850EJX, DA1ATY, E78XK4, KN8JXW, VH5DWP, JX71XR, 148L96, KR69DY, M109P1, MK3VNM, X50AR7, KW011V, XM59HJ, 8Y47TR, JD72RX, ND5JDA, T92XV4, 5K7V66, YN3J9D, 8J4YER, A74527, KE1HDR, H5565X, 4L4K4N, PT47V2, 6J83LP, RX1MM4, 5L6P60, L90X5H, 7X117R, MA78Y5
Recall started
April 16, 2020
Recall terminated (FDA)
Nov. 15, 2021
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Packaging
What the company told customers
Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
Recall number
Z-2099-2020

This recall among the others

  • FDA's enforcement report came 41 days after the firm started the recall. For half of the 1,088 FDA medical device recall events reported in 2020 with a start date, it took 54 days or fewer (by year).
  • Howmedica Osteonics: 17 other recalls or notices on this site in the three years before this one (from May 28, 2017).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement reports, recall numbers Z-2084-2020, Z-2085-2020, Z-2086-2020, Z-2087-2020, Z-2088-2020, Z-2089-2020, Z-2090-2020, Z-2091-2020 and others. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-2084-2020, May 27, 2020. "Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell… and 15 more products." PlainRecords, https://plainrecords.com/recall/device/z-2084-2020/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance