Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell… and 15 more products
- Reason
- The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Status
- Terminated
- Distribution
- Worldwide distribution including the following states: New York, Arkansas, Oklahoma, Michigan, Delaware, Texas, Louisiana, California, Florida, Arizona, Nevada, Ohio, Colorado, Rhode Island, Iowa, Pennsylvania, Massachusetts, North Carolina, Alabama, New…
- Enforcement report date
- May 27, 2020
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated (FDA's definition: FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard; as in openFDA's data of Nov. 10, 2026)
- Filed by
- Howmedica Osteonics Corp.
- Reason
- The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.
- Distribution pattern
- Worldwide distribution including the following states: New York, Arkansas, Oklahoma, Michigan, Delaware, Texas, Louisiana, California, Florida, Arizona, Nevada, Ohio, Colorado, Rhode Island, Iowa, Pennsylvania, Massachusetts, North Carolina, Alabama, New Hampshire, Illinois, New Jersey, Virginia, Utah, Maryland, Mississippi, Missouri, Tennessee, Connecticut, and Kansas
What this record is about
A generic icon of the kind of thing the record is about; the record's own words are below. It is not a picture of the product.
“The Stryker Restoration Anatomic Shell (RAS) may fall off its post within its inner packaging which can result in the device loose in the package with coating debris.”Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 54 Left, Sterile, Catalog # 504-02-54C-L, UDI (01)07613327012354(17), FDA enforcement report, 2020-05-27. The caption is the record's own words.
The picture shows the kind of thing the record is about (a part, a product, a package), not the product itself; the ring marks it. Nothing in it is a claim beyond the record's own words below.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Distributed in (FDA's record): Alabama Arizona Arkansas California Colorado Connecticut Delaware Florida Illinois Iowa Kansas Louisiana Maryland Massachusetts Michigan Mississippi Missouri Nevada New Hampshire New Jersey New York North Carolina Ohio Oklahoma Pennsylvania Rhode Island Tennessee Texas Utah Virginia
16 products in this recall
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 54 Left, Sterile, Catalog # 504-02-54C-L, UDI (01)07613327012354(17)
- Quantity
- 4895
- Codes and lots
- Lot #s JD8WN9, 6514K5, AD78D5, TR0M4X, YT7R3V, 2A7V17, Y81WDV, AN2DME, 514WVN, 9P3WE4, LD3LAL, L21A8P, K02EYK, 6K0J1W, XK44M4, H55J1M, T10T38, VT07WW, 7T1WTE, 6544JE, J45R2R, 8963DT, LV2EW7, PN1X3T, DN0276, RP35M3, T245XV, 497H8P, LJ52DA, HP48J3, H98XN3, 9E8M7P, 1J1RDH, R235KJ, 793724, RA44MH, JJ65MR, N12K3P, PX3TTK, 425H7M, 3V6JAM, P52HME, WX80AA, K65W39, 1R40DL
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2084-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 54 Right, Sterile, Catalog # 504-02-54C-R, UDI (01)07613327012415(17)
- Quantity
- 4895
- Codes and lots
- Lot #s 6A7N7Y, J50WR6, EM4ED2, 4908KM, K14HL3, VN5NTY, 4X7XE4, JK4A37, W20A86, EM5Y28, H81JNL, EE883L, 9L8LMD, 7Y2HM0, MM0JYY, K23J10, DY42DE, 3P3P64, H77Y5E, MX801E, 1L0758, HM5WKX, D058L6, 4L2R4R, VY0PK4, 872EHV, YD53A4, 4T54D0, 4768E2 WJ0W8W, YV5804, P98YT4, 8N512J, A162RN, J35YL4, H089KL, WW76RP, P58D3J, XH7HN8, RV6DJ8, 8Y0XVE, JM34X2, WJ776A, RH1MRT, N28J76, NP2XXL, L7453E, AA1DYJ, 258Y6J, 6N4X1M, E66DMR, V63W1E, 2739DR, YH1P91, 216NDE, A4410X
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2085-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 56 Left, Sterile, Catalog # 504-02-56D-L, UDI (01)07613327012361(17)
- Quantity
- 4895
- Codes and lots
- Lot #s 4J3N24, W03J5W, N119A4, WK7K99, NM70TR, 3376L4, RK3XRL, RH27YA, M00KML, TM2MMK, PW30H8, DL01A0, WD4T0V, 2D40YA, 4Y5LAE, W63A8E, VJ34EV, AR5H0T, HY6Y45, KV2835, P84X94, KK0AVP, 2M8RW6, AL0AR3, TP1Y1H, W66X1L, PH28V2, P00606, 3Y3MHV, V32P52, EJ32W6, HN8HXP, 2V46PY, TD1A6J, 7R1X76, W62DKJ, N3435R, W92EP4, HJ0X8X, 5E3X1N, P53NP4, DJ2WKH, P32YRY, X62L47, XV86AL, XV2EY5, 108Y9R, T143LR, VH573D, W56T4K, A071R3, V319PM
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2086-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 58 Left, Sterile, Catalog # 504-02-58D-L, UDI (01)07613327012385(17)
- Quantity
- 4895
- Codes and lots
- Lot #s AX3V4E, DX109D, TD2X0H, JV415X, KW5467, HM4L3L, WK5WV2, LD2HHY, A67ERK, Y579RW, RW1XVW, VJ7TE0, M02DYJ, EW1TJ5, 371T2M, 2W3T75, 345TJ9, N968R7, H386RY, 978TAX, L6266H, JV4M06, Y7169E, M840TN, 9D6PY0, TV06A0, PR2X50, P85T79, 4M5N2M, 9P3RN6, 247VR6, WE2A2A, 65792M, HA59H2, 4N6HT3, 7N01JA, 762PPX, 2D8P93, 2V4609, NE509M, VJ4RT3, KM720L, 7J3TRH, DP1EDA, V2760N, X05E74, D477X9, R372K2, TR0YTY, P720M0, Y63AD0
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2087-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 56 Right, Sterile, Catalog # 504-02-56D-R, UDI (01)07613327012347(17)
- Quantity
- 4895
- Codes and lots
- Lot #s EE5WWV, DY6V66, 4T75EE, HT3X47, YX8KLL, 99554T, T58TR0, A46AJA, P23W2H, 9X0WHV, 681E6V, XP0AKD, RD53PW, R67PR4, N11PLW, WW20KE, KV6P9X, 6X4R9T, 2X4KLD, M08065, VN205Y, Y68MTM, EP6P1E, E257HE, PP894J, 4X47VL, 7J0H1A, K50T9R, 2E2139, V92X6Y, T42NM1, LV7LYM, RV8RWH, 2E5WH7, 8H3HK1, JV2PT1, PT243R, M30439, RH817R, XT5R2K, 112LE7, 4J485X, D360KK, PM5R68, KN48NR, LT1YKM, H82NXR, VP8MRR, 4Y44YM, V27D3P, N40RWX, YE2L2E, H500TT
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2088-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 58 Right, Sterile, Catalog # 504-02-58D-R, UDI (01)07613327012446(17)
- Quantity
- 4895
- Codes and lots
- Lot #s 5L0YLY, DN3EJH, Y62465, 288DP4, LW0D2K, 3K4R62, KN2W5X, M44PVH, 3W01XK, 1Y123H, M55NTP, 683N70, JN4PR7, 497YJX, L022L2, 37889P, X86194, 775A9V, X30XTJ, W333HY, H376E6, 1V4A3X, TP5KK8, VA13MW, 9W0HD6, V72LA3, 3V1WAR, 4H7V1M, 173VRV, 9A4D9X, JE6E3M, 9H2LYX, T532A9, WJ72K7, HX5AN0, XM1A2X, 9T3XPK, TR8WMV, 9A8X5J, 5D78YR, RK0MJJ, PL0LA1, 6P4VX9, A83DPD, 8X715Y, WM2AHV, 1M5P7E, 2E36LW, X81L96
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2089-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 60 Left, Sterile, Catalog # 504-02-60D-L, UDI (01)07613327012422(17)
- Quantity
- 4895
- Codes and lots
- Lot #s Y461NK, 916M1K, 3581JK, 306XWX, AH4AER, YP48JA, 6J7PTL, KT81WD, 1D6NNE, 467JP5, 6M025E, T4286L, 633D81, 6L6ADX, N175H8, WJ41VD, RM697T, A602M0, 840Y1A, 988KM2, LL2VPW, Y37LK3, M35P5P, 280YM6, R62EM9, WM4W56, JE46JM, AK0560, D11734, EK644L, D802YL, HP25TN, L618D8, N532D2, NW0M02, 972H4A, 95171N, JD4934, WN7K01, V8245A, AK00NH, VL1688, 2E7J6X, 5030EN, P13E4K, K94NEN, MA5W63, 8X783D, 5A3RW3, TV13J7
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2090-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 60 right, Sterile, Catalog # 504-02-60D-R, UDI (01)07613327012408(17)
- Quantity
- 4895
- Codes and lots
- Lot #s 9Y6675, 2J7WNV, A02X70, ME3841, P94D02, TD37HJ, DT8E38, LM49H5, WE24MT, 74064X, A36853, 694L07, YJ3J46, H26D71, 9X2RDX, AV6M23, L686MM, JY3288, MV2HXY, NP416P, 2L349V, XT6WTY, H02LL7, 601W6E, 6X855W, PV8H2X, R846J8, L705V1, D62TJ0, TM88WY, JE7AW0, 6W0DE9, 3400PT, D62345, 2E6E25, 422V52, 5R8KH0, VE2X77, W56APY, 8267AV, YH5N6L, P32YNV, T5091D, 4M46YJ, 7N4HW9, MM0T9T, XL0WKM
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2091-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 62 left, Sterile, Catalog # 504-02-62E-L, UDI (01)07613327012453(17)
- Quantity
- 4895
- Codes and lots
- Lot #s EK2JXE, V26EXD, K97J79, MV20HR, RM5RX2, 478E97, ET7EXW, J74RKR, L71A2P, P276D0, M15D1R, ND74JJ, 6P6MWY, X500MV, K931PP, 6J36RW, MH415K, T96X19, VM332V, WW65Y5, X13MJV, L97NMP, RL4L86, MR1RLA, MN1P4X, R91A0P, N76VMX, W361J6, RL4L13, Y36651, PA8JD4, NA049P, 8V5K0A, 5M8HYP, 896EHD, 232R3X, 5D2PDD, T137D5, ML43R3, 5R6RN3, DT11P7, N07WMV, VR591J, 7H32J5, 903NRE, 1X5HLV, AA5Y4M, WL5KR3, PR4Y4L, JL7L46, 9K37HW, NA0N94, YA7HX9, AX89T7, RH44YE, 8E30L6, N411HJ, 3V37MK
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2092-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 62 right, Sterile, Catalog # 504-02-62E-R, UDI (01)07613327012439(17)
- Quantity
- 4895
- Codes and lots
- Lot #s PH230T, 5316PX, 6487DY, MV4KAW, MY6XW7, NY7JW8, XT3NK1, 8J1031, W43YMN, Y363WD, R58ALD, YM0ATP, MD6J43, M04YA2, 1V4RY5, NY85E3, 9439YT, KR3025, NN41PD, W28H8N, 7T0V3R, MT7MA3, PW531T, JT1H7A, WL30XL, LN8LTY, HJ82TV, XN0HR9, RY7TH7, L97718, KN4A66, JE84VV, 5X3H7R, 404547, 86593Y, 7610XX, WY3E23, JH7RH6, HD1H4E, 59725R, T17YD9, KX6P0W, 4A4TW6, 42775A, E02TD6, 3M44J8, DM2L4J, YX2RMM, 9L6EN0, NJ69D3, 4P7MTN, 2P31RY, 935J0K
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2093-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 64 left, Sterile, Catalog # 504-02-64E-L, UDI (01)07613327012484(17)
- Quantity
- N/A
- Codes and lots
- Lot #s Y32LP9, AN6ATN, A77J57, EE69A8, 625VEM, K8549T, EV6MXD, L85MMA, 7E2R48, M21Y21, H71N5N, A74YXL, PN2X3H, V2774X, AE6LJR, 7D0YJE, 6957YA, LM5N9M, 477ELH, JN8347, AY8T8N, EM0W77, TL29PE, NX79TH, 956DXP, A86VNH, KX5MX1, WW7VPM, 7A58DL, 37348T, N80109, 1269DY, 2351A5, H58V9W, 5M1K6H, T3509E, HL627T, WY1XP7, A76DE6, PX677J, AK4L6N, EA2MNR, AM5H8X, KM0M7W, D33PLA, 7R165A
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2094-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 66 left, Sterile, Catalog # 504-02-66F-L, UDI (01)07613327012491(17)
- Quantity
- 4895
- Codes and lots
- Lot #s J4895V, V9556J, RA3188, DH4RW8, WH083M, KY10DV, 937YAE, XJ1WW1, K52EHL, JX5291, M50LV5, 561RR3, 847AX0, V3377L, EJ40D2, W40V6T, TD3H2L, 9X0DHT, LL7R18, 9N0LH6, 6874DA, Y114EE, PP6X0D, NH50PJ, 6P32NN, MA1VYD, 328L1T, 52028A, X800K7, L51VJR, M22WTW, W80LRW, NV83VN, Y54MNA, V74712, EH1VK4, D9414T, PA6RRP, NT8241, 3P112L, NW7K52
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2095-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 64 right, Sterile, Catalog # 504-02-64E-R, UDI (01)07613327012460(17)
- Quantity
- 4895
- Codes and lots
- Lot #s HP3142, J93277, HK4HK0, N18T96, 590V0J, WA1D8T, 6D4MT7, 2V3A9W, HE19E5, JN2DWN, 777N5V, YY4LXE, JA0986, 213804, 86112L, 8X4H1D, N25J45, EM8Y25, 647TEX, J011V3, N75WT3, H98AT2, JD1K8M, M576M5, 8Y64X7, E810LX, 7J2099, RH6TNE, HR6M1R, 101MM7, V76WAA, TX6X20, 8N36E0, 3T6330, MR025M, WL6T19, 9K2L3R, K53TKW, RT8P7Y, M82N19, 7E1R1Y, TJ56KK, RX8HHY, NJ4XWJ, WM5VVN
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2096-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 66 right, Sterile, Catalog # 504-02-66F-R, UDI (01)07613327012477(17)
- Quantity
- 4895
- Codes and lots
- Lot #s 1X5DJK, NH4W8Y, 744EH9, AD25RA, 7E4HW3, 1N21NK, E00T60, WY1DD1, D26R36, H74H7L, RT0AK2, 593Y5R, AD00T8, E12H25, 9K66L6, AL7K8E, MP11MY, D04T4A, 8K74RX, 1R57DY, NH0352, KW5K62, 767RKK, V21PPJ, 6580X6, X73EDY, 5A1ANY, LJ0DWY, RR0T4W, JA761W, N11LR0, 1A03DE, 5E7XRM, TW81AP, 7E69MM, M42Y3R, A28827, 5N1DTH, Y6782W, J33J50, D930PJ, XR6WL0, TH4ED5, J54YNJ, D755P9
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2097-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 68 left, Sterile, Catalog # 504-02-68F-L, UDI (01)07613327012521(17)
- Quantity
- 4895
- Codes and lots
- Lot #s YX05L9, AL6MM6, VX02E2, 5Y1H86, XY5HP7, MA49HV, A26E76, 4V4APT, LD87TH, MH1RLL, 8R05HH, A67K6H, Y00YXH, PA02HR, LE0L4W, 494M8R, 4T4RAV, PW2E41, 3T59A5, T384A0, PD2Y0T, V44JY5, 650WXN, 3D3H6W, DT2YN6, M550AN, 3N0PN7, YA60V9, DV13RY, 4M1NHR, 3M48X6, LR49YE, NT897W, DX8JV1, VN44ME, VR1H7D, TW4EJ0, T722PN, 6V04AX, 390D9L, KW3W1W, 4D6DWK, W079A2, Y81W0J
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2098-2020
Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell…
- Product
- Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell, Size 68 right, Sterile, Catalog # 504-02-68F-R, UDI (01)07613327012514(17)
- Quantity
- 4895
- Codes and lots
- Lot #s DD6E5N, 3314LE, EA67YJ, TA05PD, MJ71XA, VH242D, HR43JY, 5E8XPW, KR4YH2, XW7T57, W92MH8, DY1N6E, 4D8EEY, 6M7D2N, 1T81T1, V45AA1, YL2DP4, H017J0, 850EJX, DA1ATY, E78XK4, KN8JXW, VH5DWP, JX71XR, 148L96, KR69DY, M109P1, MK3VNM, X50AR7, KW011V, XM59HJ, 8Y47TR, JD72RX, ND5JDA, T92XV4, 5K7V66, YN3J9D, 8J4YER, A74527, KE1HDR, H5565X, 4L4K4N, PT47V2, 6J83LP, RX1MM4, 5L6P60, L90X5H, 7X117R, MA78Y5
- Recall started
- April 16, 2020
- Recall terminated (FDA)
- Nov. 15, 2021
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Root cause (FDA)
- Packaging
- What the company told customers
- Stryker Orthopaedics sent an Urgent Medical Device Recall Letter to consignees on April 16, 2020, to alert them of the packaging failure for the Restoration Anatomic Shell (RAS) implant, sizes 54mm 68mm, (Left and Right). The firm explained that the implant can become loose within the packaging which can result in packaging and/or coating debris. Stryker Orthopaedics instructed the consignees to immediately check all stock areas and/or operating room storage for affected product and inform all users of the recall notice. The notice states that use of affected products should be discontinued. The affected products should be quarantined and returned to Stryker through their Sales Representative or Distributor. The firm requested that consignees return the Urgent Medical Device Recall Business Reply Form within 5 days via fax or email, regardless of whether they have any physical inventory onsite anymore.
- Recall number
- Z-2099-2020
This recall among the others
- FDA's enforcement report came 41 days after the firm started the recall. For half of the 1,088 FDA medical device recall events reported in 2020 with a start date, it took 54 days or fewer (by year).
- Howmedica Osteonics: 17 other recalls or notices on this site in the three years before this one (from May 28, 2017).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement reports, recall numbers Z-2084-2020, Z-2085-2020, Z-2086-2020, Z-2087-2020, Z-2088-2020, Z-2089-2020, Z-2090-2020, Z-2091-2020 and others. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-2084-2020, May 27, 2020. "Stryker Orthopaedics-Restoration Anatomic Hemispherical Shell… and 15 more products." PlainRecords, https://plainrecords.com/recall/device/z-2084-2020/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections