FDA · Medical device recall
OPMI Pentero…
- Date posted by the FDA
- Sept. 23, 2011
- Status
- Terminated
- Filed by
- Carl Zeiss Meditec, Inc.
- Product
- OPMI Pentero, OPMI Pentero C, OPMI Pentero 900, OPMI Vario/S8, OPMI Vario/S88, OPMI Neuro Multivision/NC4, OPMI Vario/NC33, OPMI Neuro/NC4, OPMI Neuro/NC4 Ceiling Mount. A surgical microscope and accessories is an AC powered device intended for use during surgery to provide a magnified view of the surgical area.
- Reason
- Some third party surgical drapes have optically active protective lenses that could affect the ZEISS surgical microscope position information that is passed on to third party navigation systems.
- Distribution pattern
- Nationwide distribution: USA only.
- Quantity
- 1677 consignees
- Codes and lots
- All-OPMI¿ Pentero¿, OPMI¿ Pentero¿ C, OPMI¿ Pentero¿ 900, OPMI¿ Vario/S8, OPMI¿ Vario/S88, OPMI¿ Neuro Multivision/NC4, OPMI¿ Vario/NC33, OPMI¿ Neuro/NC4, OPMI¿ Neuro/NC4 Ceiling Mount.
- Recall started
- Sept. 12, 2011
- Device type
- Microscope, Surgical
- Root cause (FDA)
- Pending
- What the company told customers
- The firm, Zeiss, sent an "IMPORTANT SAFETY ALERT FOR ZEISS SURGICAL MICROSCOPES" letter dated September 7, 2011, to all affected customers. The letter described the product, problem and action to be taken. The customers were instructed to use surgical drapes compatible with ZEISS surgical microscopes and always follow the instructions for use of the third party navigation system manufacturer for calibration, calibration verification and operation. In addition, the firm supplied customers with a supplement to their Zeiss surgical microscope user documentation and requested that they read the information and keep a copy with their original instructions also the customers were instructed to complete and return the Acknowledgement of Receipt form within 5 business days via fax to 1-866-809-6044 or USPS to Stericycle, Inc., Attn: Event 2480, 2670 Executive Dr. Suite A, Indianapolis, IN 46241. Should you have any questions about this Safety Alert, please contact your Carl Zeiss Customer Care representative at 1-888-871-7120.
- Recall number
- Z-3275-2011
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-3275-2011. The agency record governs; check it and the company before acting.