Presource Convenience Kits: 1) KIT… and 1 more product
- Reason
- The kits contain applicators that may exhibit an open or incomplete seal on the packaging of the applicator. Use of a potentially contaminated applicator could result in no patient harm or may lead to localized skin or soft tissue infection. Less commonly…
- Status
- Ongoing
- Distribution
- US Nationwide distribution in the states of FL, KY, MO, NC, OH, TX.
- Enforcement report date
- Sept. 30, 2026
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
- Filed by
- Cardinal Health 200, LLC
- Reason
- The kits contain applicators that may exhibit an open or incomplete seal on the packaging of the applicator. Use of a potentially contaminated applicator could result in no patient harm or may lead to localized skin or soft tissue infection. Less commonly, systemic infection or other infectious complications may occur.
- Distribution pattern
- US Nationwide distribution in the states of FL, KY, MO, NC, OH, TX.
What the record says about this device
A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.
“The kits contain applicators that may exhibit an open or incomplete seal on the packaging of the applicator. Use of a potentially contaminated applicator could result in no patient harm or may lead to localized skin or soft tissue infection. Less commonly, systemic infection or other infectious complications may occur.”Presource Convenience Kits: 1) KIT, PRE OP , Catalog Number: AMN548616A; 2) KIT, NICU ADMIT NEW HAN , Catalog Number: BA1DH0116B; 3) KIT, PRE PROCEDURE , Catalog Number: IVSJV8302E; 4) KIT, IV SUPPLIES , Catalog Number: IVS3223531; 5) KIT, IV SUPPLIES , Catalog Number: IVS3223532; 6) KIT, IV SUPPLIES , Catalog Number: IVS3223533; 7) KIT, IV SUPPLIES , Catalog Number: IVS3223534; 8) KIT, IV SUPPLIES , Catalog Number: IVS3223535; 9) KIT, IV SUPPLIES , Catalog Number: IVS322358M; 10) KIT, IV SUPPLIES , Catalog Number: IVS322358N; 11) KIT, IV SUPPLIES , Catalog Number: IVS322358O; 12) KIT, PATIENT ADMISSIONS , Catalog Number: AMN333032A; 13) KIT, C SECTION , Catalog Number: MAT42267AL; 14) KIT, C SECTION , Catalog Number: MAT4226725; 15) KIT, VAGINAL DELIVERY , Catalog Number: MAT422677G; 16) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226792; 17) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226793; 18) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226794; 19) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226795, FDA enforcement report, 2026-09-30. The caption is the record's own words.
A generic device, not the product. The mark follows the record's own words.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: Tray, Surgical
Distributed in (FDA's record): Florida Kentucky Missouri North Carolina Ohio Texas and nationwide
2 products in this recall
Presource Convenience Kits: 1) KIT…
- Product
- Presource Convenience Kits: 1) KIT, PRE OP , Catalog Number: AMN548616A; 2) KIT, NICU ADMIT NEW HAN , Catalog Number: BA1DH0116B; 3) KIT, PRE PROCEDURE , Catalog Number: IVSJV8302E; 4) KIT, IV SUPPLIES , Catalog Number: IVS3223531; 5) KIT, IV SUPPLIES , Catalog Number: IVS3223532; 6) KIT, IV SUPPLIES , Catalog Number: IVS3223533; 7) KIT, IV SUPPLIES , Catalog Number: IVS3223534; 8) KIT, IV SUPPLIES , Catalog Number: IVS3223535; 9) KIT, IV SUPPLIES , Catalog Number: IVS322358M; 10) KIT, IV SUPPLIES , Catalog Number: IVS322358N; 11) KIT, IV SUPPLIES , Catalog Number: IVS322358O; 12) KIT, PATIENT ADMISSIONS , Catalog Number: AMN333032A; 13) KIT, C SECTION , Catalog Number: MAT42267AL; 14) KIT, C SECTION , Catalog Number: MAT4226725; 15) KIT, VAGINAL DELIVERY , Catalog Number: MAT422677G; 16) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226792; 17) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226793; 18) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226794; 19) KIT, VAGINAL DELIVERY , Catalog Number: MAT4226795
- Quantity
- 17503 kits
- Codes and lots
- 1) AMN548616A, UDI-DI: 10197106721834(each), 10197106721834(case), Lot Number: 4249818 ; 2) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5124984 ; 3) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5129488 ; 4) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5129488 ; 5) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5188103 ; 6) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5188103 ; 7) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5211212 ; 8) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5155824 ; 9) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5211985 ; 10) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5279226 ; 11) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5155824 ; 12) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5188103 ; 13) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5211212 ; 14) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5246952 ; 15) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5279226 ; 16) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5279226 ; 17) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 6062168 ; 18) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5155824 ; 19) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 5279226 ; 20) BA1DH0116B, UDI-DI: 10198956006331(each), 10198956006331(case), Lot Number: 6062168 ; 21) BA1DH0116B, UDI-DI: 0(each), 0(case), Lot Number: 6051588 ; 22) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5124984 ; 23) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5124984 ; 24) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 25) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5124984 ; 26) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 27) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 28) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5188103 ; 29) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5195818 ; 30) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211907 ; 31) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211985 ; 32) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211212 ; 33) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211985 ; 34) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5279226 ; 35) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 36) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5211985 ; 37) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 6037918 ; 38) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5129488 ; 39) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5155824 ; 40) IVSJV8302E, UDI-DI: 10198956022683(each), 10198956022683(case), Lot Number: 5279226 ; 41) IVS3223531, UDI-DI: 10198956173941(each), 10198956173941(case), Lot Number: 5124984 ; 42) IVS3223532, UDI-DI: 10198956173958(each), 10198956173958(case), Lot Number: 5124984 ; 43) IVS3223533, UDI-DI: 10198956362536(each), 10198956362536(case), Lot Number: 5188103 ; 44) IVS3223533, UDI-DI: 10198956362536(each), 10198956362536(case), Lot Number: 5211985 ; 45) IVS3223534, UDI-DI: 10198956470347(each), 10198956470347(case), Lot Number: 5129488 ; 46) IVS3223534, UDI-DI: 10198956470347(each), 10198956470347(case), Lot Number: 5155824 ; 47) IVS3223534, UDI-DI: 10198956470347(each), 10198956470347(case), Lot Number: 5310701 ; 48) IVS3223535, UDI-DI: 10198956527010(each), 10198956527010(case), Lot Number: 5188103 ; 49) IVS3223535, UDI-DI: 10198956527010(each), 10198956527010(case), Lot Number: 60622168 ; 50) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4249818 ; 51) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4248961 ; 52) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4249818 ; 53) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4249818 ; 54) IVS322358M, UDI-DI: 10195594809362(each), 10195594809362(case), Lot Number: 4249818 ; 55) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5124984 ; 56) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 4249818 ; 57) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5124984 ; 58) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5129488 ; 59) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5188103 ; 60) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5188103 ; 61) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5195818 ; 62) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5188103 ; 63) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5188103 ; 64) IVS322358N, UDI-DI: 10198956106086(each), 10198956106086(case), Lot Number: 5129488 ; 65) IVS322358O, UDI-DI: 10198956517073(each), 10198956517073(case), Lot Number: 6051588 ; 66) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5248375 ; 67) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5309623 ; 68) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5309623 ; 69) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5310701 ; 70) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 6051588 ; 71) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5150160 ; 72) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5188102 ; 73) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5188103 ; 74) AMN333032A, UDI-DI: 10198956367791(each), 10198956367791(case), Lot Number: 5350797 ; 75) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5126398 ; 76) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5126398 ; 77) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5181074 ; 78) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5188102 ; 79) MAT42267AL, UDI-DI: 10198956120129(each), 10198956120129(case), Lot Number: 5188102 ; 80) MAT42267AL, UDI-DI: 5276968 (each), 10198956120129(case), Lot Number: 10198956120129; 81) MAT4226725, UDI-DI: 5126398 (each), 10198956181557(case), Lot Number: 10198956181557; 82) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5126398 ; 83) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5126398 ; 84) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5129058 ; 85) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5126398 ; 86) MAT422677G, UDI-DI: 10198956120136(each), 10198956120136(case), Lot Number: 5210853 ; 87) MAT4226792, UDI-DI: 10198956207974(each), 10198956207974(case), Lot Number: 5129058 ; 88) MAT4226793, UDI-DI: 10198956426405(each), 10198956426405(case), Lot Number: 5350797 ; 89) MAT4226794, UDI-DI: 10198956484559(each), 10198956484559(case), Lot Number: 5309623 ; 90) MAT4226794, UDI-DI: 10198956484559(each), 10198956484559(case), Lot Number: 5310701 ; 91) MAT4226795, UDI-DI: 10198956538023(each), 10198956538023(case), Lot Number: 6051588
- Recall started
- Aug. 10, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Tray, Surgical
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- An URGENT: PRODUCT CORRECTION notification letter dated 8/10/26 was sent to customers. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: gmb-FieldCorrectiveAction@cardinalhealth.com For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits please notify your customers of this recall. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.
- Recall number
- Z-3254-2026
Presource Convenience Kits: 1) KIT…
- Product
- Presource Convenience Kits: 1) KIT, USGPIV , Catalog Number: PVMHIVMHD; 2) KIT, USGPIV , Catalog Number: PVMHIVM14; 3) KIT, USGPIV , Catalog Number: PVMHIVM15
- Quantity
- 9480 kits
- Codes and lots
- 1) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 20 ; 2) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5126398 ; 3) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5126398 ; 4) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5129488 ; 5) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5181074 ; 6) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5181074 ; 7) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5182660 ; 8) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5188103 ; 9) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5126398 ; 10) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5181074 ; 11) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5188103 ; 12) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5188102 ; 13) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5199118 ; 14) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5188103 ; 15) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5195818 ; 16) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5150160 ; 17) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5195818 ; 18) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5210853 ; 19) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5248375 ; 20) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5276968 ; 21) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5350797 ; 22) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5279226 ; 23) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5350797 ; 24) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5307185 ; 25) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 5310701 ; 26) PVMHIVMHD, UDI-DI: 10197106488959(each), 50197106488957(case), Lot Number: 6051588 ; 27) PVMHIVM14, UDI-DI: 10198956165434(each), 50198956165432(case), Lot Number: 5126398 ; 28) PVMHIVM15, UDI-DI: 10198956184459(each), 50198956184457(case), Lot Number: 5129058
- Recall started
- Aug. 10, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Peripheral Catheter Insertion Kit
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- An URGENT: PRODUCT CORRECTION notification letter dated 8/10/26 was sent to customers. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Additional labels will be sent to your facility upon request. Please send an email to: gmb-FieldCorrectiveAction@cardinalhealth.com For component credit or replacement, please contact the following: " Hospitals Cardinal Customer Service at 866.551.0533 " Distributors Distributor Management at 800.635.6021 " Federal Government facilities Government Customer Service at 800.444.1166 If you are a Distributor and have further distributed affected kits please notify your customers of this recall. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.
- Recall number
- Z-3255-2026
This recall among the others
- FDA's enforcement report came 51 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).
- Cardinal Health 200: 23 other recalls or notices on this site in the three years before this one (from Oct. 1, 2023).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement reports, recall numbers Z-3254-2026, Z-3255-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-3254-2026, Sept. 30, 2026. "Presource Convenience Kits: 1) KIT… and 1 more product." PlainRecords, https://plainrecords.com/recall/device/z-3254-2026/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections