GE Healthcare Centricty PACS…

Oct. 2, 2024 Class II

Date
Oct. 2, 2024
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare
Product
GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.
Reason
GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.
Distribution pattern
Worldwide distribution.
Quantity
2138 units
Codes and lots
All versions of Centricity PACS: V2, UDI/DI Not Applicable; V3, UDI/DI Not Applicable; V4, UDI/DI 00840682124430; V6, UDI/DI 00840682104807; V7, UDI/DI 00840682145572 installed on the following systems: CARTI02312611, CARTI22214087T, 501268SSASW, 501268SSATSW, 928757VPASW, 928757VPBSW, 928757VPTSW, 602344PVSW2, 602344REPSITEB, MARICOPAC02308029, 602344TIMSVM, 480323PVSW, 661633PVSW, 805395PVSW, 805395PVTSW, V5SPORTS01804913, 310349PACSSWPV, UCLAMC2259090, 310349PACSSWBPV, UCLAMC2259063, 831624PVASW, UHSD02184496, 916781CPACSSRWT, PALMDALEM01810792, PALMDALEM16023955, UHSD16084343, PALMDALEM20508297, PALMDALEM21178954, MHL02186470, 909788HCAPACS1, 916781CPACSCKS, 916781CPACSBCKS, 650696PACSEESW, MPHS2259091, GSHCACPSW, GSH320037250, VSMY01820423, 209532SESW, SONORAREG19607823, RANCHO02255039, RANCHOPISU2, RANCHOPTESTISU, To be Provided , 805370HCALR, VMPACS69, LOSROBLES19666790, SOUTHWEST01815383, MCANPACSW, MCASPACSW, CENPACSW, 303329HCAPSWS, 303584HCAPVSW, 303584REPBSW, 303584HCAPVSWTESTIMS, SRMCPACSW, 719584VPSW, PKV2259020, 719584VPBSW, NSMCPACSW, 970476PVSW, VVMC15111181, 302674PACSSW, 302674PACSBSW, 302736SAEATSW, JFKSW, JFKR19980220B, INFINITYW01452121, 727462PVSW, 727462PVBSW, MPMH2259089T, 772461HCAPACSW, 772461HCAPACSW1, 772461HCAPACSWT, FWMCCPTSW, 352333PACSSW, NFRH19663978B, 904202BMCDSW, 904202BMCSSW, MJPACSW, MJPACTSW, 904202BMCDTSW, BIHT20915070-IMSTest2, BIHT20915071-IMSTEST4x, 407518COREPACS, HH1001398036, 407518CPREPB, 863680PVSW, KRMCOREPACSSW, KRMC19478073, 407888HCASWSS, 615344HCASW, PCMC420084442, 561694CAPACSSW, WFLHIV01, WESTFLORI19661467, 772335PACSSW, 772335PVSW, 941637PVSW, 941637TIMS90, 850875PVSW2, 904819PVSW, 904819PV2SW, 904819PVTSW, 727519PVBSW, 850325PACSCORE, CRMC820084433, 888663CPBSW, USFIMSSSASW, 727519PVSW, VSMAE01817090, NGDC01893569, 888663CPTSW, 888663CPB2SW, USFIMSSSABSW, USFIMSSSATSW1, USFIMSSSATSW2, 727519PVMSW, 727519PVTSW, VAMAF02345966, VSMI18957442DCC, 407518SCPACSHC, 850325HCASWOPT, WFLHIVT01, PCMC42259036, 352333PACSTSW, VSMAM02004646, WELLINGTON01810730, UHSD02289750, WELLINGTON19678279T, WELLINGTON20036598, TRMPACSW, TRMNC20021605, 100115-8-Centricity PACS-21222171, TRMTESTPACS, Doctors Ho22222417, 407772CORESW, 404605PVBSW, 404851PVSW, 404851PVTSW, 404851PVBSW, 404605PVTSW, 407772COREBSW, 407772CORETSW, 706651HCACORE, 912466PVSW, SGHS2259039, 912276PVSW, 912276PVTSW, 706278PACSSWPV, 706278PACSSWBPV, 706278PACSSWTPV, 404501PACSSWPV, 404501PACSSWBPV, 404501PACSSWATPV, 404501PACSSWBTPV, 706754PACSSW, 404605PVSW, 478633PACSSWPV, MCG2259011, MCG20084728, VSMQL01987331, 404325PACSSWPV, 404325PACSSWBPV, 404325PACSTSW, 406238PVSW, VSMZ01812192, 641422PVSW, PROSPECT2259057, PROSPECT2259084, PROSPECT22128566, 319272SCPSWULP, 319272CPDRSW, 319272CPDRBSW, 319272SCPTSWULP, 208COREPVSW, EID19660790, 208CORETIMS, 208743PVSW, 208743PVTSW, 847NWCPVSW, 847NWCPVSW2, 773880PVSW, 795079PACSEESW1, 795079PACSEESW2, 312926PACSSWPV, 312996PVSW2, JSHPACSSW, 312996PVSW1, 312996PVSWT1, 312996PVSWT2, 795079PACST, JHSJHOCC02259030, 773880PVBSW, 773880PVTSW, JSHPACSTSW, 217876PVSW, 217876PVBSW, 815285PACUV100, 630963PVSW, ADVPACSTSW1, 847437PVSW, 847437DCPVSW, 847437SAPVTSW, 847990PVSW, S046PACSSWPV, S801PACSSWPV, S046PACSSWTPV, ADVPACSSW, ADVPACSBSW, ADVPACSTSW, 815398CPACSSW, 219325PVSW1, 219325PVBSW1, 219325PVTSW1, 765983PVSW, 765983PVSW2, RHHCS02253964T, 812232PACSPV, 812232PACSTPV, 812232PACS2PV, THRH20084364, FRNKHAPACSPV, FRNKHAPACSTPV, FRANKFORT19633100, RRMCPACSW, RGMCPACSTW, RRMC120080697, VTP01810617T, VHT02334235P, VTP19617532, 985867CORESW, 985230PVSW, NOMC2259032, CFO18227240, 985230PVSWS, 504903CPACSSW, 504988SSASW, 504903CPACSTSW, 504988SSATSW, MCOLANO20039031, HTMC20085961, 504903CPACSIMS, 617638SSASW, 617638SSBSW, 617638SSATSW, 781979PVSW, 781979PVBSW, 781979PVTSW, 413447BCPVSW, 413447HCPVSW, 413447PVSW, 413748PVSW, 413748PVTSW, 413748PVBSW, 301698PVSW, 301698PVTSW, NORTHE02315652, NORTHE02370307T, 207795PVSWC3, 207795PVSWC1, 207795PVSWC2, 517353PACSEESW, 517353PACSEETSW, 906387VPSW, 248858PVASW, 248858PVBSW, 248858PVTSW, 906358PACUV100, 507288PVSW, 507288PVSWT, OMCR19625746, 651855PVSW, 651855PVBSW, 612831PVTSW, 660885PVSW, 660885PVTSW, S011PACSSWPV, S011PACSBSW, S011PACSTSW, 417623PVSW, 417623PVTSW, 417623PVBSW, BARTONCOU01819621, S009PACSSWPV, S009PACSSWBPV, S009PACSSWTPV1, S009PACSSWTPV2, 601824PVSW, 601859PVSW, MCMC2259045T, 601200PACSEESW, 601200PACSSWPV, 601376PVSW, 601936PACSSWPV, SDJM20032054, 601883PVSW, 601883PVTSW, DBCSW, 406345SCPACSTSW, 828326PVSW, 704660PVSW, 704660PVTSW, 402484PACSSW, AMIM02031832, 603433TSW1, 603433SW, PRH119666811, 609652PACSSW, 609652PACSBSW, 609652PACSTSW, 609835PACSSWPV, 609835PACSTSW, 609835PACSSWBPV, VHGPVSW, VHGPVBSW, VHGPVTSW, 856342CPACSSW, 856342CPACSBSW, 856342CPACSTSW, 973490PVSW, 973490PVTSW2, 973490PVTSW1, UMDNJPVPSW, UMDNJPVPBSW, UMDNJPVPTSW, 609383UVPACS100, 908522PVSW2, ARTHRITIS01452981, 505609PVSW, SJRMIMST20655929, 702733CPACSSW, SHPACSW, SUNPACSW, VALLEYHEA14460287, SUNPACSREPB, 20910879CPACST, NORTHERNN20284710, NORTHERNN20838223, NORTHERNN20838225, BLH01442255TST, BLH01451970, STO01440342, STO02157029, STO02157030T2, 718206PVSW, 718206PVTSW, 2435-1-Centricity PACS-20608371, 718206REPBSW, 315798PVSW1, 315798PTISU, COLPVPSWIMSA, MSKPVSW, MTSPACSSW, NEWYORK01976546, VSMAT01986937, MSI02099022, COLPVPSWIMSB, COLPVPSWIMSAT, COLPVPSWIMSBT, MSKPVBSW, MSKPVTSW2, MSKPVTSW3, MTSPACSBSW, ERM16295285, MSI19614991, 718920PVSW, 718920PVTSW1, BROOKHAVEN15531196, OAR01373900, 516763PVSW, 516763PVBSW, 516763PVTSW1, 516763PVTSW2, STO18806599, 914357PVSW, GSHP01370055, 914357PVBSW, 914357PVTSW, 315470CPACSASW, 315470CPACSBSW, 315470CPACSTSW, CRH21306236, 914681PVSW, 914681PVTSW, 718920PVYSW, 718920PVTSW2, 330543PACSSWPV, CHPA2259061, CHPA2259088T, 513747PVSW, 614234PVSW, 614722PVSW, MARKHZAN01465406, 614722PVBSW, SAH52259083, SAH52259110T, 614722PVTSW, 740348PACSSW, 740348PACSBSW, 740348PACSTSW, 419383PVSW, MUOO2259017, 419383PVSWB, STMARYS-TISU, STMARYS-PISU, S040PACSSWBPV, S040PACSSWAPV, S040PACSSWTPV, 503257PACSSWPV, UHS101383446, PAMC19604625T, TMGSSASW, 215913CPBSW, 484526PVSW1, 484526PVTSW, 484526CPREPB, 215710PVSW, SMMC14898807, QUANTUMIM14640640A, QUANTUMIM14776702B, QUANTUMIM14776726T, 570523PVSW, EVAN02302786T, EVCH22176253, EVANVTPACSIMS, 724728SSASW, 724728SSABSW, 724728SSATSW2, STMARYME01476581, 412575CPACSSW, 412623CPBSW, UHS100555969, UHS101404909, 412623CPTSW, 215453PACSSWPV, GVH102398920T, GVH121323962, 401444PVBSW, 401444PVSW, 401444PVTSW, AIMC01810707, AIMC19632995, 864716PACSSWPV, 864716PACSSWT, 864716PACSSWBPV, TRMT2PACSW, 803461PACSSWPV, SCOA2259023, 843347PVSW, 843347PVTSW, CNOS02186078, 605353SSASW, 605333PVSW, 605882PVSW, 423493PACSW, 423553VPSW, 423855PACSW, 423493REPB, WTN00104692T, WTN2259070, WTN2259097, WTN21176121, 423837HCASWSU, 423431PVSW, 865977SSASW, 865977SSTSW, 806354CPACS, 806354CPACSB, NORTHWEST21314320, 806354CPACST7, 361694VPSW, DRISCOLLC02257225TIMS, 214820EEPVSW1, BSWNPACIMSDB1TR, SCW2259068, FORTDUNCA01810773, FORTDUNCA20854738T, FORTDUNCA21337445, DSMCCOREPACSSW, DSMCCOREPACSDRSW, DSMC-22886533, 214820PVSW, 972906GSPACSSW, HCA-NORT02332691, MEDIC01812756, 281274MSLPVSW, 713441MHPVSWA, 713704MHMSWA, 713704MHMSWB, 713704MHMSWTA, 713704MHMSWTB, 713741MHHPVSWB, VSMHE01893552, DHOL01810756, DHOL15291375T, DHOL20631410, DHOL23253841, 713770CVPACS, 210670PACSSWPV1, 254918PACSUV100, 214820EEPVSW2, BSWNPACIMSDB2T1, 254724PVSW, SCW2259095, VASOY02315718, 361572SSASW, CMDC2259015, HH1001979502, 801DASSW, 801DSWSP1, 801377PVSW1, 801SOFTWARE, 801SOFTWAREOPT, SMH820071869, 801924HCADASSW, MNTSW, 434654PVSW, APICPAX1CS07, 540433PVSW, CFHCAPACSW, RHC123290092, CMCPACSW, 352867PACSSW, HDHPACSW, CMCPACSTW, FHS219676614, CH721013826, CH721013832T, LGMSW, LGHI20084372, LGMSWOPT, RESTRAPVSW, RHC120038618, STSP21736781, STSP22125863, 802775SSASW, 802775SSATSW, 802775SSATSW4X, KVCH01462777, 206292PACSSWPV, 206292PACSSWTPV, 206364PVSW, 206545EESW, 206364PVBSW, UWTDB101, UWH20021873REPB, UWTDB110, 414465PVSW, 414465PVTASW, 414465PVTBSW, 414465PVBSW, PCHT01721604, 608745PVSW, COMMUNITY01973590, 304487PVSW, PCH12259035-TISU, 304536UVSW, DZ1906PP02, DZ1381PP01, DZ27??PP03, DZ2112CC01, To be Provided , To be Provided , To be Provided , HCIT4691855CPACS, HCIT1555339PACS, HCIT4825134PACS, HCIT1524287PACS, HCIT4691855PACS, HCIT1596840PACS, HCIT4430774CPACS, HCIT5184728CPACS, HCIT4859945PACS, 0910620047, 0910620047, 0910620047, 0910552308, 0910552310, 0910553037, 0910553037, ANZVICCAUL, ANZVICSAND, AH0104CS01, AH5045CS01, BH1038PC01, 22303605, To be Provided , 130041ST02, 030151ST09, 010120ST03, 010120ST52, 300013ST04, 300013ST09, 300013ST11, 300013ST21, 300013ST24, 441236VPSW, VSMAI01812924, HCIT4691866CPACS, HCIT4759526PACS, HCIT7102800PACS, HCIT7125379CPACS, To be Provided , HCIT5245250CPACS, HCIT5245250CPACS, HCIT4860936PACS, HCIT4432783PACS, HCIT5317135CPACS, HCIT4290686PACS, HCIT5245275CPACS, 323782-Centricity PACS-21288445, HCIT4790060PACS, HCIT4769017PACS, HCIT7203852PACS, HCIT4478566PACS, HCIT7134190CPACS, HCIT4568867PACS, HCIT7104315PACS, HCIT7103264IMS, 656687-Centricity PACS-21288460, HCIT5273606CPACS, HCIT5245248CPACS, HCIT5245248TESTCPACS, HCIT7229698PACS, 604520REPSITEA, 604520GEPACS, 604520TESTIMS, 604586REPSITEB, 709737PACS, 709752CORDMODA, 709752IMSAsw, 709752IMSBsw, 709752ITS1A, 709752ITS2B, 709752REPSRVRAsw, 709752GEPACS, 709752DCSDAS02sw, 709752DCSDAS04sw, 709752TESTIMSsw, 709752DASTST01sw, 709737DASEHOUTsw, 709752REPSERVERsw, 709752DCSDAS01sw, 709752DCSDAS03sw, 709752TDASNL01, 709752TDASNL02, 709752DASNLA01, 709752DASNLA02, 709752DASNLA03, 709752DASNLA04, 709752DASNLA05, 709752DASNLA06, 709752DASNLA07, 709752DASNLA08, 709752DASNLA09, 709752DASNLA10, 709752DASNLA11, 709752DASNLA12, 709752DASNLA13, 709752DASNLA14, 709752DASNLA15, 709752DASNLA16, 709752DASNLB01, 709752DASNLB02, 709752DASNLB03, 709752DASNLB04, 709752DASNLB05, 709752DASNLB06, 709752DASNLB07, 709752DASNLB07, 709752DASNLB08, 709752DASNLB09, 709752DASNLB10, 709752DASNLB11, 709752DASNLB12, 709752DASNLB13, 709752DASNLB14, 709752DASNLB15, 709752DASNLB16, 709737UTILITYsw, 709752ITS2sw, 709752ITS1sw, 709752CORDMODsw, 709752DASTST02sw, 709637DAS01sw, 709637DAS02sw, 709637DAS03sw, 709637DAS04sw, 709637DAS05sw, 709637DAS06sw, 709637DAS07sw, 709637DAS09sw, 709637DASWH10sw, 709637DASWH11sw, 709637DASWH12sw, 709637DAS10sw, 709637DAS8sw, 709637DAS11sw, 709637DAS12sw, 709637DAS13sw, 709637UTILITYsw, 709637CDPSsw, 709651DASGA01, 709651DASGA02, 709651DASGA03, 709651DAS01sw, 709651DAS02sw, 709651DAS03sw, 709651UTILITYsw, 709292DAS01sw, 709292DAS02sw, 709292DAS03sw, 709292DAS04sw, 709292UTILITYsw, 709292PACS, 709454PACS, 519843IMS, 519843TESTIMS, 519843GEPACS, 905522GEPACS, 905521GEPACS, 905521IMSA, 905522IMSB, 905521TESTIMS, 613548IMSA, 613548IMSB, 613548TESTIMS, 613548GEPACS, 519685REFERENCEIMS, 519685REFERENCEDAS, 905575IMS, 905576REPSITEA, 905576REPSITEB, 905576GEPACS, 905576TESTIMS, 613732GEPACS, 613732IMS, 613732TESTIMS, 519426IMS, 519426TESTIMS, 519426GEPACS, 416813REPSITEB, 416813REPSITEA, 416813TESTIMS, 416813GEPACS, 416326IMS1, 416326IMS2, 416326GEPACS, 416756GEPACS, 416756TESTIMS, 416756IMS, 905372GEPACS, 905372IMSA, 905372IMSB, 905372TESTIMS, HCIT4602438PACS, HCIT4358052PACS, HCIT4949579PACS, HCIT4920063PACS, HCIT4096494PACS, HCIT4567313PACS, HCIT4601968PACS, HCIT5026141PACS, HCIT5183890CPACS, HCIT4888543PACS, To be Provided , To be Provided , To be Provided , To be Provided , 082499110172, 082499310155, 082455210055, 082455150052, 082455150052, 082455160221, 082455160221, 082455160134, 082499180109, 082499180102, 082455160255, 082427190180, 082445110005, 082403010015, 082455190146, 082499130790, 082499130125, 082499130790, 082499020074, 082455040339, 082407040684, 082407040684, 082455040204, 82455040334, 082499040230, 082455040337, 082455040339, 082455040147, 082455040274, 082455040352, 082445040028, 082445040028, 082455040317, SiteID-Centricity PACS-20608297, 082407040684, 082499110183, 082455190147, 82499220115, 082499100382, 082455150074, 082455150068, 082455150039, To be Provided , 082455180040…
Recall started
Aug. 5, 2024
Device type
System, Image Processing, Radiological
Root cause (FDA)
Device Design
What the company told customers
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/05/2024 by letter delivered via traceable means. The notice explained the issue, potential risk to the patient, and requested the following: "You can continue to use your device if it has not reached the maximum number of Ckeys. In order to determine the remaining number of available Ckeys: " If GE HealthCare is remotely monitoring your Centricity PACS system as part of an existing service agreement, the number of available Ckeys is assessed routinely. If your device is close to maximum capacity, the GE HealthCare support team will contact you to discuss next steps. " If GE HealthCare is not remotely monitoring your system, please contact the GE HealthCare support team who will help you check your Ckey status. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions above." "GE HealthCare will provide an updated manual with CKey monitoring instructions at no cost to you. A GE HealthCare representative will contact you with further information." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-3214-2024

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-3214-2024. The agency record governs; check it and the company before acting.