SunMed Tracheal Tube…

Sept. 23, 2026 Class II

Date
Sept. 23, 2026
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Well Lead Medical Co. Ltd
Product
SunMed Tracheal Tube, Size 2.5; Model Number: 1-7333-25
Reason
The 6Fr suction catheter could not pass through the samples smoothly.
Distribution pattern
US Nationwide distribution in the state of MI.
Quantity
27200
Codes and lots
Model Number: 1-7333-25; UDI-DI:10889483048108; Lot No.: 2108011462, 2108011506, 2109011963, 2111012039, 2112012040, 2204010568, 2204010569, 2204010734, 2205010833, 2207011263, 2208011729, 2208011789, 2304010830, 2306011372, 2306011425, 2308011907, 2308012011, 2308012012, 2309012128, 2310012358, 2310012599, 2311013016, 2311012924, 2312013223, 2401010069, 2401010197, 2402010378, 2403010466, 2403010889, 2404010999, 2406011822, 2407012282, 2408012389, 2409012800, 2409012882, 2508013006, 2510013653, 2511014081, 2512014080, 2602010272, 2603010274, 2603010900, 2604011014, 2604011416, 2605011923, 2605011924;
Recall started
July 31, 2026
Device type
Tube, Tracheal (W/Wo Connector)
Root cause (FDA)
Process control
What the company told customers
On July 31, 2026 URGENT: MEDICAL DEVICE RECALL letters were emailed to customers. Actions to be taken: 1. We request that you check your inventory for product within the scope of this recall. Quarantine all impacted product immediately. Cease use and distribution immediately. 2. If the affected product is currently in use on a patient, do not perform a prophylactic endotracheal tube exchange solely due to this notification. Continue to manage the patient according to institutional protocols and clinical judgment. 3. If you do have stock in scope of this recall, mark the applicable checkbox on the Affected Product Information Table and contact Well Lead by E-mail . 4. If you do not have stock in scope of this recall, mark the applicable checkbox on the Affected Product Information Table and return the form to the E-mail address . 5. For affected products, please contact your distributor to arrange product replacement.
Recall number
Z-3172-2026

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-3172-2026. The agency record governs; check it and the company before acting.