INTRABEAM Shielding Material Flat 20 x 20 cm…

Aug. 10, 2011

Date posted by the FDA
Aug. 10, 2011
Status
Terminated
Filed by
Carl Zeiss Meditec, Inc.
Product
INTRABEAM Shielding Material Flat 20 x 20 cm, Sterile/FO 0.05 mm Pb Equivalent (50keV) Carl Zeiss Meditec AG Indicated for radiation therapy treatments.
Reason
Carl Zeiss Meditec received a report that the INTRABEAM 20 x 20 cm sterile shield's packaging looked wet inside. Investigation found the substance to be Triethylene Glycol,
Distribution pattern
Nationwide (USA) Distribution including the states of FL, CA, NY, VA and AL.
Quantity
8 boxes of 10 sterile packages.
Codes and lots
Lot 110851; Catalog number 304534-6000-585.
Recall started
July 1, 2011
Device type
System, Therapeutic, X-Ray
Root cause (FDA)
Process control
What the company told customers
Carl Zeiss Meditec, Inc. sent an "Important Field Correction Action" letter dated July 13, 2011 to all affected customers. The letter included description of affected product and problem. Customers are asked to notify the firm if they have experienced any problems with the product and to return any remaining product for replacement upon completion of required testing. Customers are asked to fill out and return an Acknowledgement Form. For additional information contact Carl Zeiss Meditec at (925) 570-4844.
Recall number
Z-2989-2011