U.S. safety records, as the agencies wrote them. Independent.
FDA · Medical device recall

Arrow-LOK Implant…

Recall started · FDA enforcement report Aug. 26, 2026 Class II

Reason
Implant did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.
Status
Ongoing
Distribution
US Nationwide distribution in the states of VT, FL, TN, NY, GA, TX, HI.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Aug. 26, 2026
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Arrowhead DE LLC
Product
Arrow-LOK Implant, REF: 21-253519, 21-253522, 21-253525, 21-253528, 21-253530, 21-253533, 21-253535, 21-253540, 21-253545, 21-253550, 21-353513, 21-353516, 27-253519, 27-253522, 27-253525, 27-253528, 27-253530, 27-253533, 27-253535. Implant indicated for fixation of osteotomies, arthrodeses and reconstruction in the lesser toes following corrective procedures.
Reason
Implant did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.
Distribution pattern
US Nationwide distribution in the states of VT, FL, TN, NY, GA, TX, HI.
Quantity
4,171
Codes and lots
REF/UDI-DI/Lot: 21-253519/00855967006035/24662119, 242162; 21-253522/00855967006042/24662122, 242163, 220702; 21-253525/00855967006059/24662125, 220703, 242164; 21-253528/00855967006066/24662128, 242165, 230704; 21-253530/00855967006073/242166, 230705, 220705; 21-253533/00855967006080/24662133, 242167, 230706; 21-253535/00855967006097/24662135, 242168, 230707; 21-253540/00855967006110/242169; 21-253545/00855967006127/242170; 21-253550/00855967006134/24662150; 21-353513/00855967006004/24662113, 242160; 21-353516/00855967006011/24662116, 230701, 242161; 27-253519/00855967006028/24662719, 242171; 27-253522/00855967006141/242172; 27-253525/00855967006158/242173; 27-253528/00855967006165/242174; 27-253530/00855967006172/242175; 27-253533/00855967006189/24662733, 242176; 27-253535/00855967006196/24662735, 242177
Recall started
June 23, 2026
Voluntary or mandated
Voluntary: Firm initiated
Device type
Pin, Fixation, Smooth
Root cause (FDA)
Process change control
What the company told customers
On 6/23/2026 customers were phoned and informed to return affected implants. On 7/7/2026, recall notices were mailed to risk managers who were asked to do the following: 1) Cease use, distribution, and quarantine affected product. 2) Contact firm customer service to obtain instructions for returning product. 3) Email mail tracking numbers to firm. 4) Complete and return the acknowledgement and receipt form. Questions can be directed to the firm at 470-419-1429
Recall number
Z-2958-2026

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device, the reason the record gives, the record's wordsA simplified drawing of a medical device with a mark for the reason the record gives. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: Arrow-LOK Implant, REF: 21-253519…2 The record describes a lack of assurance of sterility3 What the record says4 What the record listsSTERILE POUCH?STERILE POUCH?FOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepsterility assurance

“Implant did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.”Arrow-LOK Implant, REF: 21-253519, 21-253522, 21-253525, 21-253528, 21-253530, 21-253533, 21-253535, 21-253540, 21-253545, 21-253550, 21-353513, 21-353516, 27-253519, 27-253522, 27-253525, 27-253528, 27-253530, 27-253533, 27-253535. Implant indicated for fixation of osteotomies, arthrodeses and reconstruction in the lesser toes following corrective procedures., FDA enforcement report, 2026-08-26. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Pin, Fixation, Smooth

Distributed in (FDA's record): Florida Georgia Hawaii New York Tennessee Texas Vermont and nationwide

This recall among the others

  • FDA's enforcement report came 64 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-2958-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-2958-2026, Aug. 26, 2026. "Arrow-LOK Implant." PlainRecords, https://plainrecords.com/recall/device/z-2958-2026/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance