Menu
FDA · Medical device recall

Brand Name: Elevaris/Spectra Medical Devices Product Name: Sodium Chloride Injection…

Aug. 19, 2026 Class I

Date
Aug. 19, 2026
Risk class
Class I (FDA's definition, 21 CFR 7.3(m): a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death)
Status
Ongoing
Filed by
Spectra Medical Devices, Llc
Product
Brand Name: Elevaris/Spectra Medical Devices Product Name: Sodium Chloride Injection, 0.9%, USP 10mL Ampule Model/Catalog Number: AMPUL10MLK. Sodium Chloride Flush for IV tubing systems and indwelling intravascular access devices.
Reason
Manufacturing firm, Huons, was cited for issues regarding product quality and sterility assurance. If the product sterility is compromised, patients may be exposed to microorganisms, endotoxins, or foreign matter. This may trigger inflammatory responses, infections, micro-clots, and organ failure.
Distribution pattern
Distributed in the United States to medical device kit packers in six (6) states - CA, FL, NC, OK, TX, VA.
Quantity
TBD
Codes and lots
Lot Code: Product codes: 1510-1 (individual ampule) / 1510-2 (carton) Lots: AB3005, AB3006, AB3007, AB3008, AB3009, AB3010, AB4001, AB4002, AB4003, AB4004, AB4005 AB4006, AB5001, AB5002, AB5003, AB5004, AB5005,AB5006, AB5007, AB5008 No individual UDI, as this product is used only within convenience kits.
Recall started
July 6, 2026
Device type
Saline, Vascular Access Flush
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
On July 6, 2026, firm sent letters titled, "Urgent: Medical Device Recall" to customers. Customers were instructed to examine their stock for affected product and discard or return any unused Sodium Chloride Flush ampules from the identified lot numbers. For affected products already in distribution, customers should instruct downstream customers to discard or return unused product. Customers were also instructed to use extra vigilance in detecting and reporting potential patient health concerns that may be related to this product.
Recall number
Z-2873-2026

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2873-2026. The agency record governs; check it and the company before acting.

Report a mistake on this page