FDA · Medical device recall

VITROS Anti-SARS-CoV-2 Total Reagent… and 1 more product

Aug. 19, 2020 Class II

Date
Aug. 19, 2020
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Ortho Clinical Diagnostics
Reason
The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality control failures.
Distribution pattern
Worldwide distribution - US Nationwide distribution.

2 products in this recall

VITROS Anti-SARS-CoV-2 Total Reagent…

Product
VITROS Anti-SARS-CoV-2 Total Reagent, COV2TOT, VITROS CoV2T. Catalog number 6199922, UDI 10758750033386 - Product Usage: Rx ONLY. For in vitro diagnostic and Laboratory Professional use. For emergency authorization use only.
Quantity
47,271 sales units
Codes and lots
Lots 0010-0037
Recall started
July 17, 2020
Device type
Reagent, Coronavirus Serological
Root cause (FDA)
Process control
What the company told customers
On July 17, 2020, the firm issued Urgent Product Correction Notification letters to affected customers. Customers were informed of the potential for erroneous patient sample results, calibration errors, and QC failures when using the impacted product. Customers were asked to do the following; " Visually inspect affected VITROS Anti-SARS-CoV-2 Total (Lots 0010-0037) and IgG (Lots 0010-0013) Reagent Packs to ensure correct reagent bottle configuration " Set aside any pack in which the reagent bottles are positioned incorrectly, so the lot number and pack id can be provided to Ortho. Complete and submit the enclosed Request for Credit Form, including the pack information, to receive credit for an affected pack. " Complete the Confirmation of Receipt form no later than July 28, 2020. " Post this notification by each system that processes an affected lot. " Please forward this notification if the product was distributed outside of your facility. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Recall number
Z-2770-2020

VITROS Anti-SARS-CoV-2 IgG Reagent Pack…

Product
VITROS Anti-SARS-CoV-2 IgG Reagent Pack, COV2IgG, VITROS CoV2G. Catalog number 6199919, UDI 10758750033355 - Product Usage: Rx ONLY. For in vitro diagnostic and Laboratory Professional use. For emergency authorization use only.
Quantity
2349 sales units
Codes and lots
Lots 0010 - 0013
Recall started
July 17, 2020
Device type
Reagent, Coronavirus Serological
Root cause (FDA)
Process control
What the company told customers
On July 17, 2020, the firm issued Urgent Product Correction Notification letters to affected customers. Customers were informed of the potential for erroneous patient sample results, calibration errors, and QC failures when using the impacted product. Customers were asked to do the following; " Visually inspect affected VITROS Anti-SARS-CoV-2 Total (Lots 0010-0037) and IgG (Lots 0010-0013) Reagent Packs to ensure correct reagent bottle configuration " Set aside any pack in which the reagent bottles are positioned incorrectly, so the lot number and pack id can be provided to Ortho. Complete and submit the enclosed Request for Credit Form, including the pack information, to receive credit for an affected pack. " Complete the Confirmation of Receipt form no later than July 28, 2020. " Post this notification by each system that processes an affected lot. " Please forward this notification if the product was distributed outside of your facility. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Recall number
Z-2771-2020

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-2770-2020, Z-2771-2020. The agency record governs; check it and the company before acting.