Cascade Abrazo aPTT Test Card…

Oct. 9, 2019 Class II

Date
Oct. 9, 2019
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Helena Laboratories, Corp.
Product
Cascade Abrazo aPTT Test Card, Model 5722, For Export Only. The firm name on the label is Helena Laboratories, Beaumont, TX.
Reason
Some of the cards show occasional shorter clot times than the expected variation resulting in a potential false negative result on the screening assay.
Distribution pattern
Distribution was made to Germany and Australia. There was no U.S. distribution.
Quantity
21 kits
Codes and lots
Lot number 2-17-5722, Exp. 2018-09-30
Recall started
Dec. 7, 2017
Device type
Activated Partial Thromboplastin
Root cause (FDA)
Unknown/Undetermined by firm
What the company told customers
The recalling firm issued letters dated 11/12/2017 via email on 12/7/2017 explaining the issue and that they were doing a withdrawal of the lot.
Recall number
Z-2761-2019

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2761-2019. The agency record governs; check it and the company before acting.