FDA · Medical device recall

Quick Connect 272 KG Scale used with Maxi Sky 2 ceiling lift The… and 1 more product

Aug. 15, 2018 Class II

Date
Aug. 15, 2018
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Arjohuntleigh Magog
Reason
The spreader bar can detach from the scale attachment under certain conditions.
Distribution pattern
Worldwide distribution: US (nationwide) to states of: CO, FL, IA, IL, KY, MN, NY, PH, and WA; and countries of: Australia, Belgium, Brazil, Canada, China, France, Germany, Hong Kong, Ireland, Italy, Japan, Netherlands, Norway, Singapore, Swiss, Turkey, United Kingdom, and United Arab Emirates.

2 products in this recall

Quick Connect 272 KG Scale used with Maxi Sky 2 ceiling lift The Scale is intended to be used with patient…

Product
Quick Connect 272 KG Scale used with Maxi Sky 2 ceiling lift The Scale is intended to be used with patient lifters to measure the weight of a patient who cannot stand on a scale. It is designed to be used indoors in hospitals, nursing homes, home care or health care facilities.
Quantity
874 Units (both styles)
Codes and lots
Product Code 700-19485: starting from serial number 300146551 up to serial number 300220716.
Recall started
Feb. 12, 2018
Device type
Lift, Patient, Non-Ac-Powered
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Thefirm, ARJO, sent an "Urgent-A Field Safety Notice" and Customer Response Forms to affected customers by common carrier and signature required. All affected customers need to ensure that all caregivers and users of the affected devices a remade aware of the Field Safety Notice and all listed devices at their facility are available to be corrected during service technician visit. Affected customers need also to review the Field Safety Notice, acknowledge receipt and understanding of the Field Safety Notice and the required actions by completing, signing and returning the Customer Response Form to the local Arjo office. The Arjo Sales and Service office shall arrange Arjo service technician to visit the facilities to carry out the upgrade of the scales attachment on Maxi Sky 2 ceiling lifts. If you have any further questions or require assistance completing the Customer Response Form, please contact Arjo via email at ArjoCustomer.Response-US@Arjo.com
Recall number
Z-2660-2018

Q-Connect 272 KG Scale used with Maxi Sky 2 ceiling lift The Scale is intended to be used with patient…

Product
Q-Connect 272 KG Scale used with Maxi Sky 2 ceiling lift The Scale is intended to be used with patient lifters to measure the weight of a patient who cannot stand on a scale. It is designed to be used indoors in hospitals, nursing homes, home care or health care facilities.
Quantity
874 Units (both styles)
Codes and lots
Product Code 700-19490: starting from serial number 300142013 up to serial number 300219867.
Recall started
Feb. 12, 2018
Device type
Lift, Patient, Non-Ac-Powered
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Thefirm, ARJO, sent an "Urgent-A Field Safety Notice" and Customer Response Forms to affected customers by common carrier and signature required. All affected customers need to ensure that all caregivers and users of the affected devices a remade aware of the Field Safety Notice and all listed devices at their facility are available to be corrected during service technician visit. Affected customers need also to review the Field Safety Notice, acknowledge receipt and understanding of the Field Safety Notice and the required actions by completing, signing and returning the Customer Response Form to the local Arjo office. The Arjo Sales and Service office shall arrange Arjo service technician to visit the facilities to carry out the upgrade of the scales attachment on Maxi Sky 2 ceiling lifts. If you have any further questions or require assistance completing the Customer Response Form, please contact Arjo via email at ArjoCustomer.Response-US@Arjo.com
Recall number
Z-2661-2018

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-2660-2018, Z-2661-2018. The agency record governs; check it and the company before acting.