FDA · Medical device recall

Omnipod 5 Pods… and 2 more products

Recall started · FDA enforcement report July 8, 2026 Class I

Reason
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an…
Status
Ongoing
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
July 8, 2026
Risk class
Class I (FDA's definition, 21 CFR 7.3(m): a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Insulet Corporation
Reason
External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Switzerland, Germany, Denmark, Finland, France, UK, Greece, Israel, Italy, Kuwait, Leichtenstein, Netherlands, Norway, Saudi Arabia, Sweden and Turkey.

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device, the reason the record gives, the record's wordsA simplified drawing of a medical device with a mark for the reason the record gives. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: Omnipod 5 Pods. Includes the below…2 The record describes a leak3 What the record says4 What the record listsFOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepleaking

“External soft cannula damage during manufacturing that results in insulin leaking around the Pod instead of being delivered to the user regardless of basal or bolus delivery. The primary failure mode is pump under-delivery due to loss of insulin to an external leak. If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death.”Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6720. 6. BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod)., FDA enforcement report, 2026-07-08. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Alternate Controller Enabled Insulin Infusion Pump

3 products in this recall

Omnipod 5 Pods…

Product
Omnipod 5 Pods. Includes the below Model/REF Numbers: 1. POD-OMNI-I1-6720. 2. POD-OMNI-I1-6220. 3. OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). 4. OMNI-I1-6220. 5. OMNI-I1-6720. 6. BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod).
Quantity
4,936,056 units
Codes and lots
1. Model/REF Number: POD-OMNI-I1-6720. Pack UDI-DI: 20385083000579. Single Pod UDI-DI: 10385083000527. Lot Numbers: PH1U01032521, PH1U01102521, PH1U12102421, PH1U12172421, PH1U11212421, PH1U11252421, PH1U11222421, PH1U12092421, PH1U09112421, PH1U11262421, PH1U04032521, PH1U12022421, PH1U12182421, PH1U04302524, PH1U05022521, PH1U01222521, PR1U11202533, PH1U09122421, PH1U10102421, PH1U12122421, PH1U01302522, PH1U02222521, PH1U05052521, PH1U10042421, PH1U12102412, PH1U01312521, PH1U03312521, PH1U05012521, PH1U09122422, PH1U09162421, PH1U09192421, PH1U09212421, PH1U09262411, PH1U09262421, PH1U10112422, PH1U10122421, PH1U10172421, PH1U12132422, PH1U01132521, PH1U01172521, PH1U02132511, PH1U03202521, PH1U03262521, PH1U04042521, PR1U11192531, PH1U11192511, PR1U01162621, PH1U09172421, PH1U09182421, PH1U09202421, PH1U10142421, PH1U10152421, PH1U10162421, PH1U01232522, PH1U01312511, PH1U02032521, PH1U02042521, PH1U02212521, PH1U03282521, PH1U04092521, PH1U04102521, PH1U04112521, PH1U10172522, PH1U10182521, PH1U10202521, PH1U10212521, PH1U10222521, PH1U10232521, PH1U10242521, PH1U10252521, PH1U11152521, PH1U11172511, PH1U11172521, PH1U11182521, PR1U11182531, PR1U11182541, PR1U11192541, PH1U11192521, PR1U11192532, PR1U11192542, PR1U11202521, PR1U11202532, PR1U11202541, PR1U11202542, PR1U11212521, PR1U11222531, PR1U11222541, PR1U11222521, PR1U11232541, PR1U11242521, PR1U11242531, PR1U11252541, PR1U11252512, PR1U11252521, PR1U11262521, PR1U12012521, PR1U12022521, PR1U12032521, PR1U12102522, PR1U12112521, PR1U12122521, PR1U12132511, PR1U12132521, PR1U12152521, PR1U12162521, PR1U12172521, PR1U01132621, PR1U01152621, PR1U01172621, PR1U01202621, PR1U01212621, PR1U01222621, PR1U01232621, PR1U01272621, PR1U01302641, PR1U01312631, PR1U02022641, PR1U02032631, PR1U02032641, PR1U02042641, PR1U02052631, PR1U02052641, PR1U02062631, PR1U02112641, PR1U02132611, PR1U02132641, PR1U11252531, PR1U11262541, PH1U06282422. 2. Model/REF Number: POD-OMNI-I1-6220. Pack UDI-DI: 20385083000425. Single Pod UDI-DI: 10385083000411. Lot Numbers: PH1U10182422, PH1U02202521, PH1U04222521, PH1U04232511, PH1U04232521, PH1U11142521, PH1U12042521, PH1U12052521, PH1U01092621. 3. Model/REF Number: OMNI-I1-6729 (10-Pack), OMNI-I1-6720 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: PH1U01222522. 4. Model/REF Number: OMNI-I1-6220. Pack UDI-DI: 2038508300043. Single Pod UDI-DI: 10385083000411. Lot Numbers: PH1U04182521, PH1U04212521, PH1U12192411. 5. Model/REF Number: OMNI-I1-6720. Pack UDI-DI: 20385083000548. Single Pod UDI-DI: 10385083000527. Lot Numbers: PH1U01022521, PH1U12202421, PH1U12192421, PH1U10212421, PH1U10232421. 6. Model/REF Number: BLE-H1-529 (10-Pack), BLE-H1-520 (Single Pod). Pack UDI-DI: 20385083000135. Single Pod UDI-DI: 10385083000114. Lot Numbers: PH1U01072521, PH1U01242521, PH1U03082521, PH1U04172521, PH1U01272513.
Recall started
May 20, 2026
Voluntary or mandated
Voluntary: Firm initiated
Device type
Alternate Controller Enabled Insulin Infusion Pump
Root cause (FDA)
Process design
What the company told customers
Insulet initially issued a press release on 05/26/2026 along with an "Urgent: Medical Device Correction" on the same day. Consignees were notified of the issue and hazard and instructed to not used Pods from affected lots and to check the lot number on the Pod tray lid, box or Pod itself to determine if affected Pods are on hand. If impacted Pods are on hand, consignees were instructed to discontinue use of any impacted Pod immediately, immediately change the Pod if currently using an affected lot, and acknowledge the email confirming receipt. Confirming receipt of the email is needed even if no affected Pods are on hand. Customers will receive a return kit with clear instructions on how to send back any unused affected Pods.
Recall number
Z-2603-2026

Omnipod DASH Pod…

Product
Omnipod DASH Pod. Includes the below Model/REF Numbers: 1. POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). 2. POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). 3. BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). 4. BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod).
Quantity
1,790,784 units
Codes and lots
1. Model/REF Number: POD-BLE-I1-525 (5-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000051. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11052421, PD1U11012421, PD1U11022421, PD1U02152521, PD1U02192521, PD1U04242522, PD1U10282522, PD1U10292521, PD1U10302521, PD1U10312521, PD1U11032521, PD1U11042521, PD1U01082621. 2. Model/REF Number: POD-BLE-I1-529 (10-Pack), POD-BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11192421, PD1U09102421, PD1U09092421, PD1U09062421, PD1U10242421, PD1U10252421, PD1U10282421, PD1U10292421, PD1U10312421, PD1U11042421. 3. Model/REF Number: BLE-I1-529 (10-Pack), BLE-I1-520 (Single Pod). Pack UDI-DI: 20385082000020. Single Pod UDI-DI: 10385082000023. Lot Numbers: PD1U11152421, PD1U11182421, PD1U11202421, PD1U09042421, PD1U11132421, PD1U11122421, PD1U11142421, PD1U11152411, PD1U11162421, PD1U04292521, PD1U09032421, PD1U11112421, PD1U04262521. 4. Model/REF Number: BLE-C1-529 (10-Pack), BLE-C1-520 (Single Pod). Pack UDI-DI: 20385082000112. Single Pod UDI-DI: 10385082000115. Lot Numbers: PD1U03042521, PD1U09052421, PD1U11072421, PD1U11082421, PD1U03062521, PD1U03032521, PD1U03072521, PD1U02282525, PD1U02282526, PD1U04252521, PD1U11062521, PD1U11072521, PD1U11102521, PD1U11112511, PD1U11112522, PD1U11122521.
Recall started
May 20, 2026
Voluntary or mandated
Voluntary: Firm initiated
Device type
Pump, Infusion, Insulin
Root cause (FDA)
Process design
What the company told customers
Insulet initially issued a press release on 05/26/2026 along with an "Urgent: Medical Device Correction" on the same day. Consignees were notified of the issue and hazard and instructed to not used Pods from affected lots and to check the lot number on the Pod tray lid, box or Pod itself to determine if affected Pods are on hand. If impacted Pods are on hand, consignees were instructed to discontinue use of any impacted Pod immediately, immediately change the Pod if currently using an affected lot, and acknowledge the email confirming receipt. Confirming receipt of the email is needed even if no affected Pods are on hand. Customers will receive a return kit with clear instructions on how to send back any unused affected Pods.
Recall number
Z-2604-2026

Omnipod (Eros) Pod…

Product
Omnipod (Eros) Pod. Includes the below Model/REF Numbers: 1. POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). 2. ZXR425 (10-Pack), ZXR420 (Single Pod). 3. ZXP425 (10 Pack), ZXP420 (Single Pod).
Quantity
289,046 units
Codes and lots
1. Model/REF Number: POD-ZXP425 (10 Pack), POD- ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72514. 2. Model/REF Number: ZXR425 (10-Pack), ZXR420 (Single Pod). Pack UDI-DI: N/A. Single Pod UDI-DI: N/A. Lot Numbers: L72516, L72521. 3. Model/REF Number: ZXP425 (10 Pack), ZXP420 (Single Pod). Pack UDI-DI: 20385081120033. Single Pod UDI-DI: 10385081120036. Lot Numbers: L72515, L72509, L72513, L72511, L72510, L71480, L72514.
Recall started
May 20, 2026
Voluntary or mandated
Voluntary: Firm initiated
Device type
Pump, Infusion, Insulin
Root cause (FDA)
Process design
What the company told customers
Insulet initially issued a press release on 05/26/2026 along with an "Urgent: Medical Device Correction" on the same day. Consignees were notified of the issue and hazard and instructed to not used Pods from affected lots and to check the lot number on the Pod tray lid, box or Pod itself to determine if affected Pods are on hand. If impacted Pods are on hand, consignees were instructed to discontinue use of any impacted Pod immediately, immediately change the Pod if currently using an affected lot, and acknowledge the email confirming receipt. Confirming receipt of the email is needed even if no affected Pods are on hand. Customers will receive a return kit with clear instructions on how to send back any unused affected Pods.
Recall number
Z-2605-2026

This recall among the others

  • FDA's enforcement report came 49 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).
  • Insulet: 5 other recalls or notices on this site in the three years before this one (from July 9, 2023).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement reports, recall numbers Z-2603-2026, Z-2604-2026, Z-2605-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-2603-2026, July 8, 2026. "Omnipod 5 Pods… and 2 more products." PlainRecords, https://plainrecords.com/recall/device/z-2603-2026/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance