FDA · Medical device recall
Prism Laser 7642 Laser System
- Date posted by the FDA
- Aug. 29, 2011
- Status
- Terminated
- Filed by
- American Stress Technologies, Inc.
- Product
- Prism Laser 7642 Laser System
- Reason
- It was discovered The Prism 7642 was not certified at the time of delivery. The product lacks the required labeling and the product was sold and delivered without the submission of the Initial Product Report. The product lacks User In formation
- Distribution pattern
- Nationwide Distribution
- Quantity
- 1
- Codes and lots
- S/N 7642/2009
- Recall started
- April 27, 2011
- Device type
- Analytical Measuring And Detection Laser Product
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- American Stress Technologies sent a customer notification letter dated April 1, 2011, informing them of the action they have taken to correct this issue. (1) ASTI conducted laser safety tests and checks on May 3,2011 to verify the Prism was compliant with FDA regulations. This was confirmed in a June 17, 2011, electronic mail message. (2) ASTI placed the required labels described in the initial product report on the Prism on May 3, 2011. (3) ASTI delivered updated user manuals to UNO on May 3, 2011, that describe potential laser hazards and the required labeling in the same way as described in the initial product report. (4) ASTI completed the above corrective actions at no cost to UNO as confirmed in a June 20, 2011, electronic mail message. If you have any further questions please call (724) 410-1030
- Recall number
- Z-2600-2011