Prism Laser 7642 Laser System

Aug. 29, 2011

Date posted by the FDA
Aug. 29, 2011
Status
Terminated
Filed by
American Stress Technologies, Inc.
Product
Prism Laser 7642 Laser System
Reason
It was discovered The Prism 7642 was not certified at the time of delivery. The product lacks the required labeling and the product was sold and delivered without the submission of the Initial Product Report. The product lacks User In formation
Distribution pattern
Nationwide Distribution
Quantity
1
Codes and lots
S/N 7642/2009
Recall started
April 27, 2011
Device type
Analytical Measuring And Detection Laser Product
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
American Stress Technologies sent a customer notification letter dated April 1, 2011, informing them of the action they have taken to correct this issue. (1) ASTI conducted laser safety tests and checks on May 3,2011 to verify the Prism was compliant with FDA regulations. This was confirmed in a June 17, 2011, electronic mail message. (2) ASTI placed the required labels described in the initial product report on the Prism on May 3, 2011. (3) ASTI delivered updated user manuals to UNO on May 3, 2011, that describe potential laser hazards and the required labeling in the same way as described in the initial product report. (4) ASTI completed the above corrective actions at no cost to UNO as confirmed in a June 20, 2011, electronic mail message. If you have any further questions please call (724) 410-1030
Recall number
Z-2600-2011