FDA · Medical device recall

AMX Navigate Mobile X-Ray System

Oct. 6, 2021 Class II

Date
Oct. 6, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
AMX Navigate Mobile X-Ray System
Reason
An unexpected column motion while attempting to park the column on the AMX Navigate mobile X-ray system.
Distribution pattern
Domestic Distribution Only
Quantity
3 units
Codes and lots
AMXNV2000001WK, AMXNV2000002WK, DF2401600003WK
Recall started
June 24, 2021
Device type
System, X-Ray, Mobile
Root cause (FDA)
Process control
What the company told customers
On June 24, 2021 the firm contacted the facilities that received the device via email. An AMX Navigate system was provided to 3 healthcare facilities for evaluation under an agreement with GE Healthcare. While at the healthcare facilities, these systems remained under the ownership of GE Healthcare. After one of the healthcare facilities alerted GE Healthcare to a potential issue with its system, GE Healthcare removed the system from the 3 healthcare facilities in order to perform a detailed engineering analysis as part of its root cause investigation. These 3 systems have not been returned to the healthcare facilities and remain in GE Healthcare s possession.
Recall number
Z-2590-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2590-2021. The agency record governs; check it and the company before acting.