FDA · Medical device recall
AMX Navigate Mobile X-Ray System
- Date
- Oct. 6, 2021
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- GE Healthcare, LLC
- Product
- AMX Navigate Mobile X-Ray System
- Reason
- An unexpected column motion while attempting to park the column on the AMX Navigate mobile X-ray system.
- Distribution pattern
- Domestic Distribution Only
- Quantity
- 3 units
- Codes and lots
- AMXNV2000001WK, AMXNV2000002WK, DF2401600003WK
- Recall started
- June 24, 2021
- Device type
- System, X-Ray, Mobile
- Root cause (FDA)
- Process control
- What the company told customers
- On June 24, 2021 the firm contacted the facilities that received the device via email. An AMX Navigate system was provided to 3 healthcare facilities for evaluation under an agreement with GE Healthcare. While at the healthcare facilities, these systems remained under the ownership of GE Healthcare. After one of the healthcare facilities alerted GE Healthcare to a potential issue with its system, GE Healthcare removed the system from the 3 healthcare facilities in order to perform a detailed engineering analysis as part of its root cause investigation. These 3 systems have not been returned to the healthcare facilities and remain in GE Healthcare s possession.
- Recall number
- Z-2590-2021