FDA · Medical device recall

Cuffable Blood Pressure Cuffs…

June 22, 2011

Date posted by the FDA
June 22, 2011
Status
Terminated
Filed by
GE Healthcare, LLC
Product
Cuffable Blood Pressure Cuffs; Manufactured for Vital Signs, Inc., a GE Healthcare Company 20 Campus Road, Totowa, NJ 07512 Made in Mexico. The cuff is used to determine a subject's blood pressure.
Reason
BP cuffs may not properly inflate due to a leak
Distribution pattern
Worldwide Distribution -- USA (nationwide) including Puerto Rico and countries of: Argentina, Australia, Belgium, Canada, Chile, Ecuador, Finland, Germany, Hong Kong, Israel, Italy, Latvia, Lebanon, Nicaragua, Panama, Peru, Saudi Arabia, Slovenia, UK, Uruguay and Venezuela
Quantity
2,684,996 units
Codes and lots
510(k) K911213 Medical Device Listing Number D110124 for year 2009- Lot #2689- Lot #3659 For year 2010 - Lot #0010-Lot #2530
Recall started
April 26, 2011
Device type
Blood Pressure Cuff
Root cause (FDA)
Device Design
What the company told customers
GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 26, 2011 to all affected customers. The letter identified the product, the problem and the action to be taken. Customers were instructed to discontinue use of the product, isolate all affected products, return only unused products and distribute this letter to any other healthcare institutions that are potentially affected by this issue. The letter asks that each customer complete and return the enclosed confirmation form via fax to 800-535-7923, in order to confirm receipt of the notification letter. For any questions, contact your local Vital Signs Account Manager or the International Customer Service at 1-800-932-0760.
Recall number
Z-2581-2011