FDA · Medical device recall

AD-TECH Drill Sleeve Guides…

Sept. 25, 2019 Class II

Date
Sept. 25, 2019
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Ad-Tech Medical Instrument Corporation
Product
AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N
Reason
The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Distribution pattern
NY, PA, WI, AZ, IL, CA, NH, GA, MA, MN, UT, MI, WA, FL, TX, OH, OR, PA, MD, and India, Korea, Taiwan, Canada, Hong Kong, Denmark, Russia, Singapore, Japan, Brazil. UK, Spain, Czech Republic, France
Quantity
315 units
Codes and lots
All serial numbers distributed between January 2014 and May 2019
Recall started
Aug. 27, 2019
Device type
Drills, Burrs, Trephines & Accessories (Manual)
Root cause (FDA)
Under Investigation by firm
What the company told customers
The firm initiated the recall by letter on 08/27/2019. The firm requested the units be returned.
Recall number
Z-2544-2019

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2544-2019. The agency record governs; check it and the company before acting.