FDA · Medical device recall
AD-TECH Drill Sleeve Guides…
- Date
- Sept. 25, 2019
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Ad-Tech Medical Instrument Corporation
- Product
- AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N
- Reason
- The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
- Distribution pattern
- NY, PA, WI, AZ, IL, CA, NH, GA, MA, MN, UT, MI, WA, FL, TX, OH, OR, PA, MD, and India, Korea, Taiwan, Canada, Hong Kong, Denmark, Russia, Singapore, Japan, Brazil. UK, Spain, Czech Republic, France
- Quantity
- 315 units
- Codes and lots
- All serial numbers distributed between January 2014 and May 2019
- Recall started
- Aug. 27, 2019
- Device type
- Drills, Burrs, Trephines & Accessories (Manual)
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- The firm initiated the recall by letter on 08/27/2019. The firm requested the units be returned.
- Recall number
- Z-2544-2019