FDA · Medical device recall
AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677)…
- Date
- Aug. 1, 2018
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Ad-Tech Medical Instrument Corporation
- Product
- AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from vendor 00-0001062., AD0021502, AD0021628. Product Usage: The drill sleeve guides are accessories to the Ad-Tech Cranial Drill Bits. The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement. The DSG-6.3-090-2.4N drill sleeve guides are intended to be used with the 2.4mm Cranial Drill Bit diameter.
- Reason
- The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
- Distribution pattern
- US Nationwide distribution in the states of CA, FL, IL, MI, NH, NY, OH, OR, PA, SC, UT, WV. Worldwide distribution to Canada, Denmark, Germany, India, Israel, Japan, Russia, South Korea, Taiwan, United Kingdom.
- Quantity
- 65 Total
- Codes and lots
- Catalog Number DSG-6.3-090-2.4N. Batch Number (Lot Number): 107104 (710220), 114390 (715001), 111423 (716180), 105176 (613121), 113775 (719190), 115847 (712211), 113776 (719190), 115000 (714201), 114724 (711101), 116389 (712121), 117146 (810110).
- Recall started
- May 2, 2018
- Device type
- Drills, Burrs, Trephines & Accessories (Simple, Powered)
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Customers were notified on 05/02/2018 via a MEDICAL DEVICE RECALL letter. Instructions included to immediately examine your inventory and quarantine any product subject to recall and replacement. Customers/distributors who received affected devices will be contacted to return their unopened/unused and/or reused stock of the affected product.
- Recall number
- Z-2539-2018
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-2539-2018. The agency record governs; check it and the company before acting.