FDA · Medical device recall
VITROS 3600 Immunodiagnostic System Software Versions 3.3.3 and below… and 4 more products
- Date
- July 8, 2020
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Ortho Clinical Diagnostics
- Reason
- Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances
- Distribution pattern
- US Nationwide Distribution Foreign: Country AR AU BR CA CL CN CO DO FR ID IN JP KR MX MY NP PE PH SG TH TT VE VN
5 products in this recall
VITROS 3600 Immunodiagnostic System Software Versions 3.3.3 and below Product Code: 6802783…
- Product
- VITROS 3600 Immunodiagnostic System Software Versions 3.3.3 and below Product Code: 6802783, Unique Identifier: 10758750002979
- Quantity
- 779 (U.S.= 225;OUS= 554)
- Codes and lots
- Affected Serial Number Range: J36000109-J36001317
- Recall started
- May 20, 2020
- Device type
- Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Ortho Clinical Diagnostics issued on 20 May 2020, Customer letter (Ref. CL2020-126) sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all VITROS 3600, 5600 and XT7600 . Letter states reason for recall, health risk, and action to take: In the meantime, manually adding the values in both the serum screen AND the plasma screen (even though the plasma screen appears to be updated already) will prevent the issue by correctly marking the edits to both body fluids with M1. To manually enter these values for Plasma in Options & Configurations: navigate to Plasma, select the assay, select Review/Edit Configuration, edit Qualitative fields to match Serum. If it appears they currently match, retype at least one of the changes, then touch Save. Once the Qualitative Range values have been updated, load an ADD using Retain Configuration. Verify that both Serum and Plasma values match the intended modifications. This indicates that the modifications were marked with a M1 code and will continue to be retained when that option is used during future ADD loads. " A review of previously reported patient results is recommended if you identify an assay modification in which Serum and Plasma do not match. Review Plasma sample results for these assays versus your intended thresholds. Note: if your VITROS System is connected via e-connectivity, the Ortho Care Technical Solutions Center can assist you with obtaining your patient results from the past 90 days. " If you believe this software anomaly has affected sample results in your laboratory, please contact Ortho Care Technical Solutions Center. " Complete the Confirmation of Receipt form Foreign affiliates were informed by email approximately on 20 May 2020 of the issue and instructed to notify their consignees of the issue and required actions until the software mitigation is available. The anomaly will be resolved with software version
- Recall number
- Z-2480-2020
VITROS 3600 Immunodiagnostic System…
- Product
- VITROS 3600 Immunodiagnostic System - Refurbished Product Code: 6802914 Unique Identifier: 10758750007103
- Quantity
- 27 ( U.S.= 0; OUS= 27)
- Codes and lots
- Affected Serial Number : 36000192 36000396 36000430 36000431 36000446 36000468 36000504 36000517 36000577 36000580 36000585 36000631 36000647 36000366 36000267 36000386 36000404 36000420 36000562 36000614 36000575 36000589 36000480 36000582 36000687 36000492 36001042
- Recall started
- May 20, 2020
- Device type
- Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Ortho Clinical Diagnostics issued on 20 May 2020, Customer letter (Ref. CL2020-126) sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all VITROS 3600, 5600 and XT7600 . Letter states reason for recall, health risk, and action to take: In the meantime, manually adding the values in both the serum screen AND the plasma screen (even though the plasma screen appears to be updated already) will prevent the issue by correctly marking the edits to both body fluids with M1. To manually enter these values for Plasma in Options & Configurations: navigate to Plasma, select the assay, select Review/Edit Configuration, edit Qualitative fields to match Serum. If it appears they currently match, retype at least one of the changes, then touch Save. Once the Qualitative Range values have been updated, load an ADD using Retain Configuration. Verify that both Serum and Plasma values match the intended modifications. This indicates that the modifications were marked with a M1 code and will continue to be retained when that option is used during future ADD loads. " A review of previously reported patient results is recommended if you identify an assay modification in which Serum and Plasma do not match. Review Plasma sample results for these assays versus your intended thresholds. Note: if your VITROS System is connected via e-connectivity, the Ortho Care Technical Solutions Center can assist you with obtaining your patient results from the past 90 days. " If you believe this software anomaly has affected sample results in your laboratory, please contact Ortho Care Technical Solutions Center. " Complete the Confirmation of Receipt form Foreign affiliates were informed by email approximately on 20 May 2020 of the issue and instructed to notify their consignees of the issue and required actions until the software mitigation is available. The anomaly will be resolved with software version
- Recall number
- Z-2481-2020
VITROS 5600 Integrated System Software Versions 3.3.3 and below Product Code: 6802413 Unique Identifier…
- Product
- VITROS 5600 Integrated System Software Versions 3.3.3 and below Product Code: 6802413 Unique Identifier: 10758750002740
- Quantity
- 2617 units (U.S.=1272; OUS=1344)
- Codes and lots
- Affected Serial Number Range : J56000114- J56001879
- Recall started
- May 20, 2020
- Device type
- Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Ortho Clinical Diagnostics issued on 20 May 2020, Customer letter (Ref. CL2020-126) sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all VITROS 3600, 5600 and XT7600 . Letter states reason for recall, health risk, and action to take: In the meantime, manually adding the values in both the serum screen AND the plasma screen (even though the plasma screen appears to be updated already) will prevent the issue by correctly marking the edits to both body fluids with M1. To manually enter these values for Plasma in Options & Configurations: navigate to Plasma, select the assay, select Review/Edit Configuration, edit Qualitative fields to match Serum. If it appears they currently match, retype at least one of the changes, then touch Save. Once the Qualitative Range values have been updated, load an ADD using Retain Configuration. Verify that both Serum and Plasma values match the intended modifications. This indicates that the modifications were marked with a M1 code and will continue to be retained when that option is used during future ADD loads. " A review of previously reported patient results is recommended if you identify an assay modification in which Serum and Plasma do not match. Review Plasma sample results for these assays versus your intended thresholds. Note: if your VITROS System is connected via e-connectivity, the Ortho Care Technical Solutions Center can assist you with obtaining your patient results from the past 90 days. " If you believe this software anomaly has affected sample results in your laboratory, please contact Ortho Care Technical Solutions Center. " Complete the Confirmation of Receipt form Foreign affiliates were informed by email approximately on 20 May 2020 of the issue and instructed to notify their consignees of the issue and required actions until the software mitigation is available. The anomaly will be resolved with software version
- Recall number
- Z-2482-2020
VITROS 5600 Integrated System…
- Product
- VITROS 5600 Integrated System - Refurbished Software Versions 3.3.3 and below Product Code: 6802915 Unique Identifier: 10758750007110
- Quantity
- 130 units (U.S.=57; OUS=73)
- Codes and lots
- Affected Serial Number: 56001535 56001152 56001641 56001437 56000786 56000255 56001235 56001280 56001724 56001062 56000308 56001347 56001184 56000315 56003277 56001352 56000458 56001570 56000825 56000261 56001182 56000559 56000818 56000747 56001139 56000317 56000513 56000755 56000883 56001334 56001445 56001141 56000756 56000824 56000718 56001021 56000339 56000741 56000727 56001131 56001558 56001530 56001089 56000843 56001264
- Recall started
- May 20, 2020
- Device type
- Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Ortho Clinical Diagnostics issued on 20 May 2020, Customer letter (Ref. CL2020-126) sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all VITROS 3600, 5600 and XT7600 . Letter states reason for recall, health risk, and action to take: In the meantime, manually adding the values in both the serum screen AND the plasma screen (even though the plasma screen appears to be updated already) will prevent the issue by correctly marking the edits to both body fluids with M1. To manually enter these values for Plasma in Options & Configurations: navigate to Plasma, select the assay, select Review/Edit Configuration, edit Qualitative fields to match Serum. If it appears they currently match, retype at least one of the changes, then touch Save. Once the Qualitative Range values have been updated, load an ADD using Retain Configuration. Verify that both Serum and Plasma values match the intended modifications. This indicates that the modifications were marked with a M1 code and will continue to be retained when that option is used during future ADD loads. " A review of previously reported patient results is recommended if you identify an assay modification in which Serum and Plasma do not match. Review Plasma sample results for these assays versus your intended thresholds. Note: if your VITROS System is connected via e-connectivity, the Ortho Care Technical Solutions Center can assist you with obtaining your patient results from the past 90 days. " If you believe this software anomaly has affected sample results in your laboratory, please contact Ortho Care Technical Solutions Center. " Complete the Confirmation of Receipt form Foreign affiliates were informed by email approximately on 20 May 2020 of the issue and instructed to notify their consignees of the issue and required actions until the software mitigation is available. The anomaly will be resolved with software version
- Recall number
- Z-2483-2020
VITROS XT 7600 Integrated System Software Versions 3.5.1 and below Product Code:6844461 Unique Identifier…
- Product
- VITROS XT 7600 Integrated System Software Versions 3.5.1 and below Product Code:6844461 Unique Identifier: 10758750031610
- Quantity
- Total= 353 ( U.S.= 167; OUS 186)
- Codes and lots
- Affected Serial Number Range: J76000111-J76000525
- Recall started
- May 20, 2020
- Device type
- Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Ortho Clinical Diagnostics issued on 20 May 2020, Customer letter (Ref. CL2020-126) sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all VITROS 3600, 5600 and XT7600 . Letter states reason for recall, health risk, and action to take: In the meantime, manually adding the values in both the serum screen AND the plasma screen (even though the plasma screen appears to be updated already) will prevent the issue by correctly marking the edits to both body fluids with M1. To manually enter these values for Plasma in Options & Configurations: navigate to Plasma, select the assay, select Review/Edit Configuration, edit Qualitative fields to match Serum. If it appears they currently match, retype at least one of the changes, then touch Save. Once the Qualitative Range values have been updated, load an ADD using Retain Configuration. Verify that both Serum and Plasma values match the intended modifications. This indicates that the modifications were marked with a M1 code and will continue to be retained when that option is used during future ADD loads. " A review of previously reported patient results is recommended if you identify an assay modification in which Serum and Plasma do not match. Review Plasma sample results for these assays versus your intended thresholds. Note: if your VITROS System is connected via e-connectivity, the Ortho Care Technical Solutions Center can assist you with obtaining your patient results from the past 90 days. " If you believe this software anomaly has affected sample results in your laboratory, please contact Ortho Care Technical Solutions Center. " Complete the Confirmation of Receipt form Foreign affiliates were informed by email approximately on 20 May 2020 of the issue and instructed to notify their consignees of the issue and required actions until the software mitigation is available. The anomaly will be resolved with software version
- Recall number
- Z-2484-2020
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-2480-2020, Z-2481-2020, Z-2482-2020, Z-2483-2020, Z-2484-2020. The agency record governs; check it and the company before acting.