FDA · Medical device recall

Maxi Sky 1000 (Arjo) and V10 (BHM) Bariatric Ceiling Rail…

Sept. 22, 2021 Class II

Date
Sept. 22, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Arjohuntleigh Magog, Inc.
Product
Maxi Sky 1000 (Arjo) and V10 (BHM) Bariatric Ceiling Rail - Mounted Lift with measuring function
Reason
The spreader bar can disconnect from the scale bottom attachment.
Distribution pattern
Distribution in US - AZ, CA, CO, IL, LA, MA, MI, MO, OH, PA, WA
Quantity
62 units
Codes and lots
All units of Model 700-00536: Class III 454 kg scale and Model 700-00526: 454 kg scale
Recall started
March 4, 2021
Device type
Lift, Patient, Non-Ac-Powered
Root cause (FDA)
Under Investigation by firm
What the company told customers
Customers identified as owners of the affected scales will be notified of the identified product risk and the Field Safety Corrective Action through the Field Safety Notice. Affected customers need to ensure that all caregivers and users of the affected devices are made aware of the Field Safety Notice. There will be two options for customers: - Contact Arjo to order parts and instruction required to proceed with the disconnection of the affected scale by on-site biomed personnel to continue to use the ceiling lift safely, or - Contact Arjo to ask to an Arjo service technician to visit a customer facility to perform the scale disconnection to continue to use the ceiling lift safely. The Field Safety Notice will be accompanied by Customer Response Form. Following its return to the local Arjo office and the availability of the permanent solution, a customer will be contacted to complete the action.
Recall number
Z-2479-2021