FDA · Medical device recall
Maxi Sky 1000 (Arjo) and V10 (BHM) Bariatric Ceiling Rail…
- Date
- Sept. 22, 2021
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Arjohuntleigh Magog, Inc.
- Product
- Maxi Sky 1000 (Arjo) and V10 (BHM) Bariatric Ceiling Rail - Mounted Lift with measuring function
- Reason
- The spreader bar can disconnect from the scale bottom attachment.
- Distribution pattern
- Distribution in US - AZ, CA, CO, IL, LA, MA, MI, MO, OH, PA, WA
- Quantity
- 62 units
- Codes and lots
- All units of Model 700-00536: Class III 454 kg scale and Model 700-00526: 454 kg scale
- Recall started
- March 4, 2021
- Device type
- Lift, Patient, Non-Ac-Powered
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Customers identified as owners of the affected scales will be notified of the identified product risk and the Field Safety Corrective Action through the Field Safety Notice. Affected customers need to ensure that all caregivers and users of the affected devices are made aware of the Field Safety Notice. There will be two options for customers: - Contact Arjo to order parts and instruction required to proceed with the disconnection of the affected scale by on-site biomed personnel to continue to use the ceiling lift safely, or - Contact Arjo to ask to an Arjo service technician to visit a customer facility to perform the scale disconnection to continue to use the ceiling lift safely. The Field Safety Notice will be accompanied by Customer Response Form. Following its return to the local Arjo office and the availability of the permanent solution, a customer will be contacted to complete the action.
- Recall number