FDA · Medical device recall
Centricity PACS-IW with Universal Viewer Product Usage: The… and 1 more product
- Date
- Sept. 2, 2015
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- GE Healthcare
- Reason
- Images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be missing when a system parameter MapRoute is set to a value greater than 1.
- Distribution pattern
- Worldwide Distribution - U.S. nationwide in the states of: AK, CA, CO, DE, FL, HI, ID, NY, PA, SD, TN and TX; France, Germany, Greece, India, Israel, Malaysia, Mexico, Northern Ireland (UK), Panama, Poland, Russia, South Africa, Saudi Arabia, Switzerland, Turkey, UAE and Uruguay.
2 products in this recall
Centricity PACS-IW with Universal Viewer Product Usage: The Centricity PACS-IW with Universal Viewer is a…
- Product
- Centricity PACS-IW with Universal Viewer Product Usage: The Centricity PACS-IW with Universal Viewer is a picture archiving and communications system, a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. Its hardware components may include workstations, digitizers, communications devices, computers, video monitors, magnetic, optical disk, or other digital data storage devices, and hardcopy devices. The software components may provide functions for performing operations related to image manipulation, enhancement, compression or quantification. Used in Radiology.
- Quantity
- 82 systems
- Codes and lots
- Model Number: Centricity PACS-IW with Universal Viewer Version 5.0 through 5.0 SP7 Codes: 2090255-001 UV 6.0 Web Client Media Kit. 05 Jun 15 to present
- Recall started
- Aug. 11, 2015
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design (manufacturing process)
- What the company told customers
- The firm sent out an Urgent Medical Device Correction letter dated 8/11/2015 to all affected customers. Letter states that No User activity required. GE Healthcare has completed a remote inspection of your system and found that this system parameter is currently set to the default setting of 1. The firm will correct all affected systems by a providing software update. The letter also states that if the customer has any questions they should call GE Healthcare Service at 1-800-437-1171 or their local service representative.
- Recall number
- Z-2469-2015
Centricity Universal Viewer Product Usage: The Centricity Universal Viewer Version 6.0 software is a picture…
- Product
- Centricity Universal Viewer Product Usage: The Centricity Universal Viewer Version 6.0 software is a picture archiving and communications system, a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. Its hardware components may include workstations, digitizers, communications devices, computers, video monitors, magnetic, optical disk, or other digital data storage devices, and hardcopy devices. The software components may provide functions for performing operations related to image manipulation, enhancement, compression or quantification. Used in Radiology.
- Quantity
- 26 systems
- Codes and lots
- Model Number: Centricity Universal Viewer Version 6.0 software Codes: 2090255-001 UV 6.0 Web Client Media Kit. 05 Jun 15 to present
- Recall started
- Aug. 11, 2015
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design (manufacturing process)
- What the company told customers
- The firm sent out an Urgent Medical Device Correction letter dated 8/11/2015 to all affected customers. Letter states that No User activity required. GE Healthcare has completed a remote inspection of your system and found that this system parameter is currently set to the default setting of 1. The firm will correct all affected systems by a providing software update. The letter also states that if the customer has any questions they should call GE Healthcare Service at 1-800-437-1171 or their local service representative.
- Recall number
- Z-2470-2015
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-2469-2015, Z-2470-2015. The agency record governs; check it and the company before acting.