HarmonyAIR A-Series Surgical Lighting System and 1 more product
- Reason
- The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.
- Status
- Ongoing
- Distribution
- Worldwide - US Nationwide distribution.
- Enforcement report date
- June 24, 2026
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
- Filed by
- Steris Corporation
- Reason
- The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.
- Distribution pattern
- Worldwide - US Nationwide distribution.
What the record says about this device
A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.
“The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.”HarmonyAIR A-Series Surgical Lighting System, FDA enforcement report, 2026-06-24. The caption is the record's own words.
A generic device, not the product. The mark follows the record's own words.
If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.
FDA device name: Light, Surgical, Ceiling Mounted
2 products in this recall
HarmonyAIR A-Series Surgical Lighting System
- Product
- HarmonyAIR A-Series Surgical Lighting System
- Quantity
- 2101 units
- Codes and lots
- UDIs: 00724995185947, 00724995186418, 00724995187156, 00724995188009, 00724995188498, 00724995189198, 00724995185954, 00724995186425, 00724995187170, 00724995188016 00724995188504, 00724995189211, 00724995185961, 00724995186432, 00724995187187 00724995188047, 00724995188535, 00724995189228, 00724995185978, 00724995186548 00724995187217, 00724995188054, 00724995188542, 00724995189280,00724995185985, 00724995186555, 00724995187248, 00724995188078, 00724995188580, 00724995189372, 00724995186067, 00724995186562, 00724995187262, 00724995188122, 00724995188597, 00724995189419, 00724995186074, 00724995186579, 00724995187279, 00724995188221, 00724995188603, 00724995189990, 00724995186081, 00724995186593, 00724995187286, 00724995188238, 00724995188610, 00724995190002, 00724995186098, 00724995186616, 00724995187347, 00724995188245, 00724995188627, 00724995190033, 00724995186111, 00724995186623, 00724995187378, 00724995188252, 00724995188634, 00724995190040, 00724995186128, 00724995186661, 00724995187385, 00724995188269, 00724995188641, 00724995190354, 00724995186135, 00724995186678, 00724995187415, 00724995188276, 00724995188658, 00724995190361, 00724995186203, 00724995186685, 00724995187453, 00724995188283, 00724995188665, 00724995199364, 00724995186302, 00724995186692, 00724995187484, 00724995188290, 00724995188672, 00724995199913, 00724995186319, 00724995186722, 00724995187538, 00724995188306, 00724995188689, 00724995199944, 00724995186326, 00724995186739, 00724995187569, 00724995188313, 00724995188702, 00724995200480, 00724995186333, 00724995186746, 00724995187576, 00724995188320, 00724995188726, 00724995200886, 00724995186340, 00724995186753, 00724995187583, 00724995188337, 00724995188849, 00724995200909,00724995186357, 00724995186760, 00724995187606, 00724995188344, 00724995188948, 00724995200930, 00724995186364, 00724995186784, 00724995187620, 00724995188375, 00724995189068, 00724995200947, 00724995186371, 00724995186807, 00724995187712, 00724995188412, 00724995189075, 00724995200961, 00724995186388, 00724995186906, 00724995187729, 00724995188467, 00724995189105, 00724995201074, 00724995186395, 00724995187125, 00724995187989, 00724995188474, 00724995189136, 00724995217372, 00724995186401, 00724995187149, 00724995187996, 00724995188481, 00724995189181, 00724995222970
- Recall started
- May 15, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Light, Surgical, Ceiling Mounted
- Root cause (FDA)
- Material/Component Contamination
- What the company told customers
- An URGENT MEDICAL DEVICE RECALL notification letter dated 5/1/26 was sent to customers. STERIS Action All affected Customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their lighting system(s) to perform the correction as soon as possible. For the paint issue, each affected lighting system will undergo a full visual inspection for any signs of paint delamination; as needed a service kit will be installed to prevent paint chipping from occurring or if deemed necessary replaced. For the button issue, each affected lighting system will undergo inspection for any signs of potential separation from the light head and handle. Buttons failing this inspection will be replaced. User Action Users can continue to use their lighting system(s) prior to completion of the correction. Users are reminded to ensure no excess fluids remain on light head during or after cleaning, per the Operator Manuals. Should any issues be observed prior to the correction being completed, users should submit a complaint by contacting STERIS Customer Service at 1-800-548-4873. Users are also reminded to report any adverse events via FDA s MedWatch Database should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact your local Account Manager.
- Recall number
- Z-2449-2026
ALYON Surgical Lighting System
- Product
- ALYON Surgical Lighting System
- Quantity
- 549 units
- Codes and lots
- UDIs: 00724995223182 00724995223137, 00724995223151, 00724995223175, 00724995222895, 00724995222994 00724995223014, 00724995223021, 00724995223083, 00724995223106, 00724995222758 00724995223120, 00724995223113, 00724995222840, 00724995223038, 00724995223076 00724995222987, 00724995223007, 00724995222703, 00724995222727, 00724995223090 00724995225711, 00724995225681, 00724995225698, 00724995225728, 00724995222826 00724995222802, 00724995222857, 00724995222697, 00724995222956, and 00724995223229
- Recall started
- May 15, 2026
- Voluntary or mandated
- Voluntary: Firm initiated
- Device type
- Light, Surgical, Ceiling Mounted
- Root cause (FDA)
- Material/Component Contamination
- What the company told customers
- An URGENT MEDICAL DEVICE RECALL notification letter dated 5/1/26 was sent to customers. STERIS Action All affected Customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their lighting system(s) to perform the correction as soon as possible. For the paint issue, each affected lighting system will undergo a full visual inspection for any signs of paint delamination; as needed a service kit will be installed to prevent paint chipping from occurring or if deemed necessary replaced. For the button issue, each affected lighting system will undergo inspection for any signs of potential separation from the light head and handle. Buttons failing this inspection will be replaced. User Action Users can continue to use their lighting system(s) prior to completion of the correction. Users are reminded to ensure no excess fluids remain on light head during or after cleaning, per the Operator Manuals. Should any issues be observed prior to the correction being completed, users should submit a complaint by contacting STERIS Customer Service at 1-800-548-4873. Users are also reminded to report any adverse events via FDA s MedWatch Database should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact your local Account Manager.
- Recall number
- Z-2450-2026
This recall among the others
- FDA's enforcement report came 40 days after the firm started the recall. For half of the 860 FDA medical device recall events reported in 2026 so far with a start date, it took 49 days or fewer (by year).
- Steris: 7 other recalls or notices on this site in the three years before this one (from June 25, 2023).
Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.
What should I do? How serious is a recall? Classes and warnings explained · How to read this record
Source: FDA · Medical device enforcement reports, recall numbers Z-2449-2026, Z-2450-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.
Cite this record
FDA recall record Z-2449-2026, June 24, 2026. "HarmonyAIR A-Series Surgical Lighting System and 1 more product." PlainRecords, https://plainrecords.com/recall/device/z-2449-2026/ (as of the day it was built).At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections