FDA · Medical device recall

Infant Warmer System (IWS)

June 21, 2017 Class II

Date
June 21, 2017
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
Infant Warmer System (IWS)
Reason
GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.
Distribution pattern
Worldwide Distribution - USA (nationwide) Distribution.
Quantity
34986
Codes and lots
Medical device listing number: D103053, D103054, D103056, D103057, D183529
Recall started
May 5, 2017
Device type
Warmer, Infant Radiant
Root cause (FDA)
Under Investigation by firm
What the company told customers
GE Healthcare sent an Urgent Medical Device Correction letter dated May 5, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Product Correction : Attached to this letter, we provide instructions as part of a Service Manual Addendum on how to correct the issue. Please add this new Addendum to the Service Manual of your device(s) and train the affected users accordingly. Please acknowledge that you have received this letter and that you understand that an action needs to be taken on your part to correct this issue by filling out and returning the attached Customer Response form. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately per the contact information above.
Recall number
Z-2323-2017