Menu
FDA · Medical device recall

VITROS Chemistry Products FS Diluent Pack 3…

July 10, 2024 Class II

Date
July 10, 2024
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS Chemistry Products FS Diluent Pack 3 comes with 3 packs per sales unit.
Reason
VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides.
Distribution pattern
Worldwide - US Nationwide distribution in the states of FL, IL, NY, PA and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Spain, Singapore, Sweden, The Netherlands, United Kingdom.
Quantity
1000
Codes and lots
Lot Code: Model No 6801754; UDI-DI 10758750007004; Lot Number 01-1645; Expiration Date: 05-Jul-2024
Recall started
May 21, 2024
Device type
General Purpose Reagent
Root cause (FDA)
Under Investigation by firm
What the company told customers
An URGENT PRODUCT CORRECTION NOTIFICATION dated 5/31/24 was sent to customers. REQUIRED ACTIONS " Discontinue using, render unusable, and discard your remaining inventory of VITROS FS Diluent Pack 3, Lot 01-1645, including packs currently loaded on your VITROS System(s). " Complete and return the enclosed Confirmation of Receipt form no later than June 10, 2024. Upon receipt of your completed Confirmation of Receipt form, QuidelOrtho will credit or replace your discarded inventory. " Save this notification with your User Documentation or post this notification by each VITROS 5,1 FS/4600/5600/XT 7600 System until the issue has been resolved. " Please forward this notification if the affected product was distributed outside of your facility. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. DISTRIBUTOR REQUIRED ACTIONS " Discard your remaining inventory of VITROS FS Diluent Pack 3, Lot 01-1645. QuidelOrtho will credit your account. " Send the enclosed customer letter and customer Confirmation of Receipt form to all customers who were shipped VITROS FS Diluent Pack 3, Lot 01-1645, from your facility. " Complete and return the enclosed distributor Confirmation of Receipt form no later than June 10, 2024. " Please forward this notification if the affected product was distributed outside of your facility. Contact Information We apologize for the inconvenience this may cause your facility. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
Recall number
Z-2266-2024

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2266-2024. The agency record governs; check it and the company before acting.

Report a mistake on this page