FDA · Medical device recall
VITROS Chemistry Products FS Diluent Pack 3…
- Date
- July 10, 2024
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Chemistry Products FS Diluent Pack 3, Model/Catalog Number: 6801754. Product Description: VITROS Chemistry Products FS Diluent Pack 3 comes with 3 packs per sales unit.
- Reason
- VITROS Chemistry Products FS Diluent Pack 3, Lot 01-1645 may cause results to be negatively biased when used to dilute patient samples tested with VITROS Chemistry Products CRP Slides.
- Distribution pattern
- Worldwide - US Nationwide distribution in the states of FL, IL, NY, PA and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Spain, Singapore, Sweden, The Netherlands, United Kingdom.
- Quantity
- 1000
- Codes and lots
- Lot Code: Model No 6801754; UDI-DI 10758750007004; Lot Number 01-1645; Expiration Date: 05-Jul-2024
- Recall started
- May 21, 2024
- Device type
- General Purpose Reagent
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- An URGENT PRODUCT CORRECTION NOTIFICATION dated 5/31/24 was sent to customers. REQUIRED ACTIONS " Discontinue using, render unusable, and discard your remaining inventory of VITROS FS Diluent Pack 3, Lot 01-1645, including packs currently loaded on your VITROS System(s). " Complete and return the enclosed Confirmation of Receipt form no later than June 10, 2024. Upon receipt of your completed Confirmation of Receipt form, QuidelOrtho will credit or replace your discarded inventory. " Save this notification with your User Documentation or post this notification by each VITROS 5,1 FS/4600/5600/XT 7600 System until the issue has been resolved. " Please forward this notification if the affected product was distributed outside of your facility. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. DISTRIBUTOR REQUIRED ACTIONS " Discard your remaining inventory of VITROS FS Diluent Pack 3, Lot 01-1645. QuidelOrtho will credit your account. " Send the enclosed customer letter and customer Confirmation of Receipt form to all customers who were shipped VITROS FS Diluent Pack 3, Lot 01-1645, from your facility. " Complete and return the enclosed distributor Confirmation of Receipt form no later than June 10, 2024. " Please forward this notification if the affected product was distributed outside of your facility. Contact Information We apologize for the inconvenience this may cause your facility. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
- Recall number
- Z-2266-2024
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-2266-2024. The agency record governs; check it and the company before acting.