FDA · Medical device recall
Disposable Surgical Cable…
- Date posted by the FDA
- Aug. 13, 2010
- Status
- Terminated
- Filed by
- Remington Medical Inc.
- Product
- Disposable Surgical Cable, Product Code: FL-601-97 (100), Single use, Rx Only, Remington Medical, Inc., 6830 Meadowridge Court, Alpharetta, GA 30005 Usage: Temporary pacing of open hearts: used with External Pacing Generators (EPGs).
- Reason
- The Safe Connect component may separate from the cable, causing the cable to become unplugged from the EPG.
- Distribution pattern
- AZ, CA, CO, CT, DE, FL, GA, IL, IN, LA, MD, MI, MO, MS, NC, NM, NY, OH, OK, OR, PA, SC, TX, TN, TX, VA, WA, WI and Canada.
- Quantity
- 1488 units
- Codes and lots
- Lot number: 100701
- Recall started
- May 19, 2010
- Device type
- Cable, Transducer And Electrode, Patient, (Including Connector)
- Root cause (FDA)
- Process design
- What the company told customers
- Consignees were notified on/about 05/19/2010. They were instructed to examine their inventory and quarantine product subject to the recall. If they distributed the product, they were advised to identify their customers and notify them of the recall (via the recall letter). A Return Authorization was provided. More information can be obtained by calling 770.888.8520.
- Recall number
- Z-2223-2010