FDA · Medical device recall

Radiesse…

May 19, 2011

Date posted by the FDA
May 19, 2011
Status
Terminated
Filed by
Merz Aesthetics, Inc.
Product
Radiesse, Injectable Implant - 1.3cc Syringe, Sterile, Sterilized Using Steam, Use Before 2013-01, Ref 8046M7, Lot 1024417, Manufactured by: BioForm Medical Inc, 4133 Courtney Rd #10 Franksville, WI 53126. Subdermal implantation for the correction of moderate to severe facial wrinkles and folds, such as masolabial folds and it is also intended for restoration and/or correction of the signs of facial fat loss (lipoatrophy) in people with human immunodeficiency virus.
Reason
Merz Aesthetics, Inc. is recalling 1 lot (1034417) of Radiesse 1.3 cc Implant. The lot was supposed to be scrapped because it had failed to meet its Loss On Drying (LOD) specification. Unfortunately it had been marked as approved, released and shipped.
Distribution pattern
International only: Taiwan.
Quantity
695 units
Codes and lots
Lot 1024417
Recall started
April 6, 2011
Device type
Implant, Dermal, For Aesthetic Use
Root cause (FDA)
Employee error
What the company told customers
The firm, Merz Aesthetics, Inc., contacted its consignee/customer via e-mail on April 6, 2011. The e-mail states the product, problem and action to be taken. The customer was instructed to follow up with the concerned and inform the status time to time to the Recall Coordinator. The product was confirmed to be quarantined and has been requested to be sent back. If you need any additional information regarding this issue or if you have any questions, contact the International Regulatory Affairs Specialist at 262-835-3300 ext. 3081.
Recall number
Z-2221-2011

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2221-2011. The agency record governs; check it and the company before acting.