FDA · Medical device recall
Power supply provided with Afinion 2…
- Date
- June 10, 2020
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- Abbott Diagnostics Technologies AS
- Product
- Power supply provided with Afinion 2, Catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, 1116679, 1116680, 1116681, 1116682, 1116684, 1116770, 1116771, 1116772, 1116777, 1116778, 1116970, 1116971, 1116985, 1116986, IVD. The firm name on the label is Abbott Diagnostics Technologies AS, Oslo, Norway. Power supply provided with Alere Afinion AS100 Analyzer, Catalog numbers 1115175, 1115390, 1116049, 1116050, 1116053, 1116054, 1116456, and 1116980. The firm name on the label is Alere Technologies AS, Oslo, Norway.
- Reason
- The analyzer system may have an electrostatic discharge and sparking of the power supply provided with the analyzers resulting in the power supply becoming non-functional.
- Distribution pattern
- Distribution was made to NJ. There was no direct government/military distribution. Foreign distribution was made to Canada, Argentina, Australia, Bangladesh, Brazil, Cambodia, Chile, China, Colombia, Denmark, Finland, France, Georgia, Germany, Guatemala, Hong Kong, Iceland, India, Italy, Ivory Coast, Japan, Korea, Madagascar, Malaysia, Mexico, Mongolia, Myanmar, The Netherlands, Norway, Oman, Pakistan, Peru, Philippines, Poland, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Timur, Trinidad & Tobago, United Kingdom, and Zimbabwe.
- Quantity
- 42,237 instruments
- Codes and lots
- All serial numbers are affected.
- Recall started
- March 23, 2020
- Device type
- Hemoglobin A1c Test System
- Root cause (FDA)
- Device Design
- What the company told customers
- The recalling firm/manufacturer issued medical device correction letters dated 3/23/2020 on 3/23/2020 to their direct distributors requesting end user customer information. The recalling firm/manufacturer issued medical device correction letters to the end users dated 3/31/2020 via a third-party notification firm. The notification informed the end user of the issue and requested to be notified of the number of power supplies needed for replacement. The affected power supplies were to be discarded by the end user.
- Recall number
- Z-2192-2020
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-2192-2020. The agency record governs; check it and the company before acting.