U.S. safety records, as the agencies wrote them. Independent.
FDA · Medical device recall

The Drill Stop Kit is the version of the Mini Sterilizable Tray that is used to store…

Recall started · FDA enforcement report Aug. 6, 2025 Class II

Reason
The Drill stop kit contains incorrect components.
Status
Ongoing
Distribution
Worldwide - US Nationwide distribution in the states of AL, FL, LA, MD, NJ, NV, NY, OH, OK, TN, TX, UT, VA and the country of Japan.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Aug. 6, 2025
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Implant Direct Sybron Manufacturing LLC
Product
The Drill Stop Kit is the version of the Mini Sterilizable Tray that is used to store, sterilize, verify Drill Stop length, and facilitate the mounting/removal of the Drill Stops. The Drill Stop Kit includes the Release Bracket, which is used for mounting and removing Drill Stops, as well as verifying the length of the Drill Stops via a ruler and depth markings. The Drill Stop Kit also allows for optional storage of some DLC drills. Drill Stops: The Drill Stop is a reusable (up to 30 osteotomies) hollow cylinder with retention prongs that clips onto the hub of the DLC drill, creating a stop function at the desired predetermined drill depth (6mm, 8mm, 10mm, 11.5mm, 13mm and 16mm). The Drill Stops come in two formats (Short – 6-13mm depth, and Long – 6-16mm depth), and in several diameters for different diameter drills.
Reason
The Drill stop kit contains incorrect components.
Distribution pattern
Worldwide - US Nationwide distribution in the states of AL, FL, LA, MD, NJ, NV, NY, OH, OK, TN, TX, UT, VA and the country of Japan.
Quantity
45 Kits
Codes and lots
Lot # 15006511/UDI: (10)841307125310
Recall started
May 1, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Sterilization Wrap Containers, Trays, Cassettes & Other Accessories
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 2, 2025, Implant Direct issued a "Urgent: Medical Device Recall" Notification" letter to affected consignees via UPS or Email. Implant Direct ask consignees to take the following actions: 1. Please check your inventory immediately for any affected units and quarantine to ensure the product is not used. 2. PLEASE SHARE THIS NOTICE WITH THOSE WHO NEED TO BE AWARE WITHIN YOUR ORGANIZATION. 3. IMPLANT DIRECT KINDLY REQUESTS YOUR COOPERATION IN COMPLETING AND EMAILING BACK THE ENCLOSED ACKNOWLEDGEMENT FORM IN ORDER TO CONFIRM YOUR RECEIPT OF THIS URGENT MEDICAL DEVICE RECALL LETTER, REGARDLESS OF WHETHER YOU HAVE ANY PRODUCT IN YOUR INVENTORY.
Recall number
Z-2172-2025

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device or a product, the reason the record gives, the record's wordsA simplified drawing for a device record; the mark follows the reason the record gives, not a photograph of the product. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: The Drill Stop Kit is the version…2 The record describes the wrong item in a package3 What the record says4 What the record listsFOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepincorrect components

“The Drill stop kit contains incorrect components.”The Drill Stop Kit is the version of the Mini Sterilizable Tray that is used to store, sterilize, verify Drill Stop length, and facilitate the mounting/removal of the Drill Stops. The Drill Stop Kit includes the Release Bracket, which is used for mounting and removing Drill Stops, as well as verifying the length of the Drill Stops via a ruler and depth markings. The Drill Stop Kit also allows for optional storage of some DLC drills. Drill Stops: The Drill Stop is a reusable (up to 30 osteotomies) hollow cylinder with retention prongs that clips onto the hub of the DLC drill, creating a stop function at the desired predetermined drill depth (6mm, 8mm, 10mm, 11.5mm, 13mm and 16mm). The Drill Stops come in two formats (Short – 6-13mm depth, and Long – 6-16mm depth), and in several diameters for different diameter drills., FDA enforcement report, 2025-08-06. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Distributed in (FDA's record): Alabama Florida Louisiana Maryland Nevada New Jersey New York Ohio Oklahoma Tennessee Texas Utah Virginia and nationwide

This recall among the others

  • FDA's enforcement report came 97 days after the firm started the recall. For half of the 972 FDA medical device recall events reported in 2025 with a start date, it took 44 days or fewer (by year).
  • Implant Direct Sybron Manufacturing: 1 other recall or notice on this site in the three years before this one (from Aug. 7, 2022).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-2172-2025. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-2172-2025, Aug. 6, 2025. "The Drill Stop Kit is the version of the Mini Sterilizable Tray that is used to store." PlainRecords, https://plainrecords.com/recall/device/z-2172-2025/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance