FDA · Medical device recall

Giraffe Shuttle…

June 3, 2020 Class II

Date
June 3, 2020
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808.
Reason
During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
Distribution pattern
US Nationwide distribution in the states of AZ, CA, CO, DE, FL, GA, ID, IL, IN, LA, MD, MI, MS, NC, ND, NM, NV, NY, OH, Ok, OR, PA, SC, TN, TX, VA, WA, WI and WV.
Quantity
204 units.
Codes and lots
Serial numbers: GSHW60086, GSHX60001, GSHX60014, GSHY70002-GSHY70195, GSHZ70101-GSHZ70108, MAAP50034, MAAQ50170, MAAR50021, MAAT50013, MAAT60045, MAAU61029, MAAU61040 and MAAV6059.
Recall started
April 2, 2020
Device type
Incubator, Neonatal
Root cause (FDA)
Device Design
What the company told customers
GE Healthcare notified customers on 04/02/2020 via "Urgent Medical Device Correction" letter. The recall letter identified the affected product and serial numbers. The firm instructed the users to discontinue use of the product if at any time they smell a gas odor. GE Healthcare will correct all affected products at a later date when the correction is available at no cost.
Recall number
Z-2111-2020

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2111-2020. The agency record governs; check it and the company before acting.