U.S. safety records, as the agencies wrote them. Independent.
FDA · Medical device recall

Hillrom Q-Link 13 component… and 1 more product

Recall started · FDA enforcement report July 16, 2025 Class I

Reason
The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but…
Status
Ongoing
Distribution
Worldwide

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
July 16, 2025
Risk class
Class I (FDA's definition, 21 CFR 7.3(m): a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Baxter Healthcare Corporation
Reason
The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.
Distribution pattern
Worldwide

How this hazard works

A simplified drawing with a generic icon, following the words of the agency's notice. It shows how the hazard is described, not an event.

Four steps: a generic product icon, what the notice describes, what it says, the remedyA simplified drawing of a kind of product with a mark for what the notice describes, then the phrases of the notice's own text. A generic test dummy acts out the risk the notice names, critical injury; it is not an event.1 The product: Hillrom Q-Link 13 component, Product Code…2 The record describes a patient who may fall3 What the record says4 The record's own words are belowpatient fallcritical injury

“The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.”Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts: 1) Uno 102 EE Mobile lift - Product Code 2010004; 2) Viking L Mobile lift - Product Code 2040044; 3) Viking XL Mobile lift - Product Code 2040043; 4) Viking M Mobile lift - Product Code 2040045A; 5) Viking S Mobile Lift - Product Code 2040006; 6) Viking XS Mobile Lift - Product Code 2040007; 7) LikoLight Mobile lift - Product Code 2030001. ., FDA enforcement report, 2025-07-16. The caption is the record's own words.

A generic icon of the kind of product, not the product. The mark follows what the notice says; the phrases on the right are the notice's own words. The test dummy stands for the risk the notice names (“critical injury”); it is not a record of something that happened.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Lift, Patient, Non-Ac-Powered

2 products in this recall

Hillrom Q-Link 13 component…

Product
Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts: 1) Uno 102 EE Mobile lift - Product Code 2010004; 2) Viking L Mobile lift - Product Code 2040044; 3) Viking XL Mobile lift - Product Code 2040043; 4) Viking M Mobile lift - Product Code 2040045A; 5) Viking S Mobile Lift - Product Code 2040006; 6) Viking XS Mobile Lift - Product Code 2040007; 7) LikoLight Mobile lift - Product Code 2030001. .
Quantity
9899 units
Codes and lots
All Q-link 13 manufactured between 8/27/2013 to 2/27/2025: 1) Product Code 2010004, UDI/DI 00887761981775; 2) Product Code 2040044, UDI/DI 00887761981867; 3) Product Code 2040043, UDI/DI 00887761981850; 4) Product Code 2040045A, UDI/DI 00887761943193; 5) Product Code 2040006, UDI/DI N/A; 6) Product Code 2040007, UDI/DI N/A; 7) Product Code 2030001, UDI/DI N/A. .
Recall started
May 30, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Lift, Patient, Non-Ac-Powered
Root cause (FDA)
Device Design
What the company told customers
Baxter issued an URGENT PRODUCT RECALL notice to is consignees on 05/30/2025 via USPS first class mail. The notice described the problem with the device, hazard, and requested the following: "Actions to be taken by customers 1. Please locate and stop using all Q-link 13 components on any of the products listed in the affected product table. 2. Contact Baxter Technical Support to discuss alternative options, if necessary, at 800-445-3720, option 2, between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or by email at HRC_Technical_Support@Baxter.com. 3. Please post this letter in areas where affected mobile lifts are stored and used. 4. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 6. Please forward a copy of this communication to all departments within your institution that use the affected mobile lifts. 7. If you are a dealer, wholesaler or distributor/reseller that distributed any affected product to other facilities, please forward this communication to your customers and check the associated box on the customer portal. Further Information and Support For general questions regarding this communication, please contact Baxter Technical Support at 800-445-3720 Option 2 between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or by email at HRC_Technical_Sup
Recall number
Z-2031-2025

Hillrom LikoScale component…

Product
Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.
Quantity
9839 units
Codes and lots
All LikoScale Adapter Kits manufactured between 8/27/2013 to 2/27/2025
Recall started
May 30, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Lift, Patient, Non-Ac-Powered
Root cause (FDA)
Device Design
What the company told customers
Baxter issued an URGENT PRODUCT RECALL notice to is consignees on 05/30/2025 via USPS first class mail. The notice described the problem with the device, hazard, and requested the following: "Actions to be taken by customers 1. Please locate and stop using all Q-link 13 components on any of the products listed in the affected product table. 2. Contact Baxter Technical Support to discuss alternative options, if necessary, at 800-445-3720, option 2, between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or by email at HRC_Technical_Support@Baxter.com. 3. Please post this letter in areas where affected mobile lifts are stored and used. 4. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 6. Please forward a copy of this communication to all departments within your institution that use the affected mobile lifts. 7. If you are a dealer, wholesaler or distributor/reseller that distributed any affected product to other facilities, please forward this communication to your customers and check the associated box on the customer portal. Further Information and Support For general questions regarding this communication, please contact Baxter Technical Support at 800-445-3720 Option 2 between the hours of 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or by email at HRC_Technical_Sup
Recall number
Z-2032-2025

This recall among the others

  • FDA's enforcement report came 47 days after the firm started the recall. For half of the 972 FDA medical device recall events reported in 2025 with a start date, it took 44 days or fewer (by year).
  • Baxter Healthcare: 83 other recalls or notices on this site in the three years before this one (from July 17, 2022).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement reports, recall numbers Z-2031-2025, Z-2032-2025. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-2031-2025, July 16, 2025. "Hillrom Q-Link 13 component… and 1 more product." PlainRecords, https://plainrecords.com/recall/device/z-2031-2025/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance